Huber LCS 80 Technical Data

Contenido

  1. Overview p. 1
    Product introduction and model identification for LCS 80
  2. Spec p. 2
    Technical data according to DIN 12876 with performance parameters
  3. p. 1 table, 40 rows

Huber LCS 80 — Technical data according to DIN 12876

ParameterValue
Order-No.VDH32100163
Operating temperature range-45...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperaturkonstanz bei -40°C2 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C20 kW
bei -15°C20 kW
at -20°C18 kW
at -25°C16 kW
at -30°C12 kW
at -40°C8,5 kW
Kälteleistung bei Umgebungstemperatur 35°C
at 15°C20 kW
bei -15°C20 kW
at -20°C18 kW
at -25°C16 kW
at -30°C12 kW
at -40°C8,5 kW
Kälteleistung bei Umgebungstemperatur 40°C
at -40°C8 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity16 kg
CO2 equivalent22,35 t
Circulation pump:
max. delivery53 l/min
max. delivery pressure4,7 bar
Pump connectionG1 male
min. filling capacity40 l
Bath volume advance warning75 l
max. filling capacity105 l
Overall dimensions WxDxH **2015x1100x2000 mm
Net weight880 kg
sound pressure level +/- 4 dB(A)77 dB(A)
Power supply (3 Phase)460V 3~ 60Hz
max. current (3 Phase)50 A
Fuse (3 phase)3x60A

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's operating temperature range is -45 to 15 °C, so it must be matched to the process temperature window before selection.

Defines whether the unit can cover the required process temperatures at all.

Source
“Operating temperature range -45...15 °C”
Performance

Cooling capacity is 20 kW at 15 °C but drops to 8.5 kW at -40 °C, so sizing must use the capacity at the actual operating point, not the nominal figure.

Prevents undersizing when the process runs at low temperatures.

Source
“at -40°C 8,5 kW”
Performance

Cooling capacity is quoted at 20 °C, 35 °C and 40 °C ambient, and above 35 °C ambient the capacity is reduced, so the installation site's ambient temperature directly limits available cooling.

Ambient conditions at the installation site must be checked against the capacity tables.

Source
“Reduction of cooling power above +35°C environmental temperature”
Installation

The unit requires a 460 V 3-phase 60 Hz supply with 50 A max current and 3x60 A fusing, so the site electrical infrastructure must be sized accordingly.

Determines the electrical supply, cable and protection device rating needed on site.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site connector.

Affects electrical installation planning and the scope of supply.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is suitable for outdoor mounting down to -20 °C ambient and up to +50 °C, which widens where it can be placed but requires weather protection planning.

Determines whether an indoor plant room is mandatory or outdoor siting is acceptable.

Source
“Suitable for outside mounting (min. environmental temperature -20°C)”
Installation

The operating panel is supplied as a remote control on a 20 m cable, so the control location must be planned within that cable length.

Constrains the layout between the chiller and the operator station.

Source
“Delivered with remote control: operating panel connected via 20m cable”
Installation

The unit is 2015x1100x2000 mm and weighs 880 kg, so transport routes, floor loading and space requirements must be verified.

Needed for rigging, floor loading and plant room layout.

Source
“Overall dimensions WxDxH ** 2015x1100x2000 mm”
Installation

The pump connection is G1 male with 53 l/min max delivery and 4.7 bar max delivery pressure, so the external circuit and hoses must match this thread and pressure rating.

Ensures the process loop connects correctly and is not over-pressurised.

Source
“Pump connection G1 male”
Operation

The bath volume ranges from 40 l minimum to 105 l maximum with a warning at 75 l, so the fill volume must be kept within these limits.

Prevents operation with too little or too much heat transfer fluid.

Source
“min. filling capacity 40 l”
Operation

The recommended heat transfer fluid is Temper -55, so the fluid must be selected to match the low-temperature duty.

Wrong fluid can freeze or degrade at the -45 °C lower limit.

Source
“Recommended thermofluid: Temper -55 (Heat Transfer Fluid)”
Safety

The refrigerant is R-449A (A1, H280) with 16 kg charge and 22.35 t CO2 equivalent, so refrigerant handling and leak reporting obligations apply.

Triggers F-gas/CO2-equivalent record-keeping and service qualifications.

Source
“Refrigerant quantity 16 kg”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and -5% voltage with +2% frequency is not allowed, so supply quality must be verified.

Poor site power quality can void compliance and damage the compressor.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.

Affects siting relative to sensitive electronics and residential zones.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”