Induztr
Resumen técnico Visor de documentos Modelos relacionados

Ministat 125w Immersion Circulator

Contenido

  1. Overview p. 1
    Product name and cover of the Ministat 125w datasheet
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 11 rows

Technical data according to DIN 12876

ParameterValue
Width bath opening WxD / bath depth178 x 80 / 120 mm
Overall dimensions WxDxH **225x370x429 mm
Net weight26 kg
Power supply requirement110-120V 1~ 60Hz
max. current11,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S544570

** Please respect space requirements. See operating conditions at www.huber-online.com

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires cooling water with inlet temperature between 5°C and 20°C and a differential pressure of at least 2 bar, so the site water supply must be verified before installation.

Cooling water outside this range will reduce cooling power or cause condensation, affecting selection and site preparation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion/electrical risk in the cooling circuit.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Engineers must size the cooling water supply for worst-case water temperature to avoid insufficient cooling.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or damage the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Ensures the power supply quality at the installation site meets the unit's operating limits.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

Avoids delays and rework if the correct plug type is not ordered.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, so installation and ventilation must account for flammability.

A3 classification requires attention to ignition sources and ventilation in the installation area.

Source
“R-290 (A3, H220)”
Selection

The bath has a 2.7 l volume but only 1.3 l capacity when a displacement rack is used, which limits usable volume for samples.

Selection must account for reduced usable bath volume when accessories are installed.

Source
“Bath capacity with displacement rack”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, which constrains fluid choice at low temperatures.

Using a fluid with higher viscosity can overload the pump and impair circulation.

Source
“max. permissible kin. viscosity”
Installation

The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.

IP20 offers no water protection, and ambient limits affect location and enclosure requirements.

Source
“max. ambient temperature”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The pressure pump delivers up to 22 l/min at 0.7 bar and the suction pump up to 16 l/min at 0.4 bar, defining external circuit capability.

External loop pressure drop and flow requirements must be checked against these pump limits.

Source
“max. delivery pressure”