Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unichiller 015w-H OLÉ Technical Data

Contenido

  1. Overview p. 1
    Product name and model identification on cover
  2. Spec p. 2
    Technical data table according to DIN 12876
  3. p. 2 table, 6 rows

Technical data according to DIN 12876

ParameterValue
max. current15,5 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S597049

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit is water-cooled, so a cooling water supply with at least 1 bar differential pressure and 5–20 °C intake is required for selection.

Determines whether the site's cooling water infrastructure is suitable before purchase.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes inside the unit must be insulated if the cooling water temperature can fall below the condensation point.

Prevents condensation and corrosion/electrical risk during installation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility decisions for the cooling loop.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation

Cooling power drops and cooling water flow rate may increase as the cooling water temperature rises.

Affects achievable performance and required cooling water flow at the actual site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump specified.

Prevents pump damage and leakage when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Avoids freezing and damage when operating near or below 0 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Ensures the electrical supply stays within the unit's permitted operating envelope.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento

The unit is EMC classified as Class A, Group 1 to EN55011, which has implications for the installation environment.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

Avoids ordering the wrong plug configuration and delays at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection

The optional Pt100 sensor connection is factory-fitted only and displays (not controls) process temperature.

Must be decided at order time; cannot be retrofitted in the field.

Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”
Operation

The OLÉ controller includes an autostart function for power failure.

Affects restart behavior and safety planning after a power interruption.

Source
“Autostart function for power failure”
Safety

The unit has adjustable overtemperature protection according to DIN 12876.

Provides a configurable safety limit to protect the process and equipment.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

Space requirements around the unit must be respected per the operating conditions.

Ensures adequate clearance for operation and maintenance at the installation site.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”