Introduces variable speed dry pump technology and target applications
Overviewp. 2
Lists analytical instrumentation applications such as MALDI/TOF and LC/MS/MS
Specp. 4
Peak pumping speed and performance data tables
p. 4table, 10 rows
Triscroll inverter – Technical Specification
TS300inv
TS600inv
Peak pumping speed
250 l/m, 15 m3/hr (8.8 cfm)
500 l/m, 30m3/h (17.7 cfm)
Ultimate pressure
1.3 x 10-2 mbar (10 mTorr)
9.3 x 10-3 mbar (7 mTorr)
Inlet connection
NW25
NW40
Exhaust connection
1⁄4" Female NPT with swivel
3⁄8" Female NPT with swivel
NW16 (adapter provided)
NW25 (adapter provided)
Gas ballast
1⁄4" Female NPT
Weight – pump only
26 kg (57 lb.)
31 kg (68 lbs)
Power requirements
Single phase 200-240 VAC, 50-60 Hz, 5 amps (max)
Operating speed
40-62 Hz (factory setting) : 1800 RPM
Noise level per ISO 11201, variable w/frequency
55-68 dBA
Engineering Facts
13 facts · every claim linked to a page
Selection
The pump speed can be finely tuned to match the optimum working pressure at the mass spectrometer inlet, simplifying retrofit with existing rotary vane pumps.
Allows engineers to match existing system pumping characteristics and avoid redesign during retrofit.
Source
“The pumping speed can be finely adjusted to deliver the optimum working pressure at the mass spectrometer inlet”
Performance
Rotational speed is adjustable to the nearest 1 Hz, giving precise control of pumping speed.
Precise speed control is essential for tuning inlet pressure in high-gas-flow API mass spectrometers.
Source
“The TriScroll Inverter pumps adjust rotational speed to the nearest 1 Hz for precise control of pumping speed”
Operation
The pump provides constant pumping speed worldwide regardless of line frequency.
Ensures consistent performance across 50 Hz and 60 Hz power grids without reconfiguration.
Source
“Constant pumping speed worldwide regardless of line frequency”
Installation
Remote start and stop is possible via a contact closure using a standard D-shell connector.
Enables integration with system control electronics without additional interface hardware.
Source
“Remote start and stop with a contact closure made possible by a standard D-shell connector”
Operation
Pump parameters can be monitored via a serial interface.
Supports remote diagnostics and system-level monitoring in automated laboratory setups.
Source
“Monitoring of pump parameters via serial interface”
Performance
Operating at lower rotational speeds significantly reduces noise in the work area with little loss in base pressure performance.
Allows quieter operation without sacrificing vacuum quality, important for laboratory environments.
Source
“Significant decrease in noise levels in the work area resulting from pump’s operation at lower rotational speeds with little loss in base pressure performance of the pump.”
Maintenance
Long-life tip seals routinely last more than a year before replacement.
Reduces maintenance frequency and cost of ownership compared to oil-sealed rotary vane pumps.
Source
“Long-life tip seals routinely last more than a year before replacement”
Cumplimiento
Dry TriScroll pumps eliminate oil disposal costs and simplify regulatory and environmental compliance.
Avoids oil handling, disposal, and associated environmental compliance burdens.
Source
“TriScroll Pumps eliminate the frequent maintenance requirements of oil-sealed rotary vane pumps, simplifying regulatory and environmental compliance, and eliminating oil disposal costs”
Installation
The inverter's low power consumption eliminates tripping of circuit breakers caused by high starting current.
Prevents nuisance breaker trips during pump startup, simplifying electrical installation.
Source
“Eliminate tripping of circuit breakers caused by high starting current”
Selection
The TS300inv provides 250 l/m peak pumping speed while the TS600inv provides 500 l/m.
Peak pumping speed is a primary selection criterion for matching the mass spectrometer gas load.
Source
“250 l/m, 15 m3/hr (8.8 cfm)”
Selection
The TS300inv achieves 1.3 x 10-2 mbar ultimate pressure and the TS600inv achieves 9.3 x 10-3 mbar.
Ultimate pressure determines the achievable base vacuum and suitability for the backing application.
Source
“1.3 x 10-2 mbar (10 mTorr)”
Installation
Both pumps require single-phase 200-240 VAC, 50-60 Hz, 5 amps maximum power.
Defines the electrical supply and circuit capacity needed for installation.