Product introduction and general description of HB120 heating bath
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 27 rows
Technical data according to DIN 12876
Parameter
Value
Order-No.
HB120
Operating temperature range
60...250 °C
Resolution of display
0,01 K
Temperature stability at 70°C
0,1 K
temperature set point / display
5,7" colour Touchscreen
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
II / FL
Heating power at 460V
12 kW
Heating power at 400V
12 kW
Heating power at 380V
10,5 kW
Circulation pump:
MK pump
max. delivery
100 l/min
max. delivery pressure
3,5 bar
Pump connection
M30x1,5 male
Overall dimensions WxDxH
479x539x695 mm
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
Net weight
88 kg
Power supply (3 Phase)
380-460V 3~ 50/60Hz
max. current (3 Phase)
20 A
Fuse (3 phase)
3x25A
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S575934
Engineering Facts
12 facts · every claim linked to a page
Selection
The unit is designed exclusively for external open operation and has no expansion tank, so the application must provide filling and volume compensation.
Determines whether the unit can be used with a closed loop or requires an external open bath/reservoir with volume compensation.
Source
“The machine is developed exclusively for external open operation, and does not possess an expansion tank.”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so the installer must provide a suitable plug and connection.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The supply voltage tolerance is ±10% only if the frequency tolerance does not act in the opposite direction, so supply quality must be checked before installation.
Prevents selecting or installing the unit on a supply where combined voltage and frequency deviations exceed the permitted envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.
Affects electrical safety design, earth connection and residual-current device selection.
Source
“Attention: leakage current > 3,5mA”
Installation
The pump connection is an M30x1.5 male thread, so mating hoses or adapters must match this thread.
Ensures correct selection of hoses, unions and adapters for the external circuit.
Source
“Pump connection M30x1,5 male”
Operation
The minimum achievable temperature of 60 °C depends on ambient temperature being below 60 °C and on the heat loss of the application.
Prevents overestimating cooling capability when the unit has no active refrigeration.
Source
“The ability to reach the min. temperature 60°C, depends on how far the ambient temperature is below 60°C, and the heat loss of the application.”
Selection
The application heat loss must be between 0.1 kW and 2.8 kW depending on the model to reach the 60 °C minimum temperature.
Allows checking whether the connected application load is compatible with reaching the minimum temperature.
Source
“The heat loss must be between 0.1kW and 2.8kW, depending on the HB model, to reach the minimum temperature of 60°C.”
Installation
The unit requires a three-phase 380–460 V 50/60 Hz supply with a maximum current of 20 A and 3x25 A fusing.
Defines the electrical supply, cable and protection requirements for installation.
Source
“Power supply (3 Phase) 380-460V 3~ 50/60Hz”
Operation
The ambient temperature must remain between 5 °C and 40 °C for correct operation.
Sets the allowable site conditions and affects placement and room ventilation.
Source
“min. ambient temperature 5 °C”
Conformité
The equipment is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for use in residential environments.
Informs EMC compliance planning and possible restrictions in residential or light-industrial locations.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit has adjustable overtemperature protection according to DIN 12876, which must be set for the application.
Requires commissioning personnel to configure the safety limit correctly for the thermal fluid and application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance
The circulation pump delivers up to 100 l/min and 3.5 bar, which must be sufficient for the external circuit pressure drop.
Used to verify hydraulic compatibility of the external application with the pump capability.