Product introduction and main features of KISS 205B
Specp. 2
Technical data according to DIN 12876 standard
p. 1table, 37 rows
Technical data according to DIN 12876 — KISS 205B (Order-No.: 2004.0009.98)
Characteristic
Value
Operating temperature range
45...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
Pressure / Suction pump
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connection
M16x1 male
Bath volume
4,8 l
Filling capacity
2,5 l
Width bath opening WxD
105x90 mm
Bath depth
150 mm
Height of bath opening
190 mm
Overall dimensions WxDxH **
178x337x355 mm
Net weight
9 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S393791
** Overall dimensions WxDxH
Engineering Facts
12 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method: 45–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator.
Determines whether the unit can reach the required process temperatures and which cooling option must be specified.
Source
“Operating temperature range
45...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and an external fuse between 10 A and 16 A.
Needed to size the electrical supply and branch circuit protection before installation.
Source
“Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A”
Installation
An external branch circuit protection according to UL 489 is required.
Mandatory for compliant electrical installation in UL jurisdictions.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage tolerance is ±10% only if the frequency tolerance does not run in the opposite direction; for example, -10% voltage with +3% frequency is not allowed.
Prevents specifying the unit on a supply where combined voltage and frequency deviations exceed the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
The pump connection is M16x1 male, and the included accessories are hose connector NW12, blank plug, sleeve nuts, cover for bath bridge, cooling coil, and pump adaptor.
Determines the mating hose/adaptor needed for external circuit plumbing.
Source
“Pump connection
M16x1 male”
Installation
The cooling coil is for (tap) water with a 3/8" connection.
Needed to plan the water cooling supply line and fittings.
Source
“Cooling coil for (tap) water (3/8").”
Operation
The plastic pump is not resistant to acetone and polyglycols, so water mixed with glysantine or ethylene glycol is recommended for freeze protection.
Prevents selecting an incompatible heat transfer fluid that would damage the pump.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
The optional Pt100 sensor connection can only display (not control) process temperature and is available only factory fitted at additional cost.
Must be ordered with the unit if external temperature monitoring is needed; it cannot be retrofitted in the field.
Source
“Option: Pt100 sensor connection #10688 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge).”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms the safety concept required for unattended operation and flammable liquids classification.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The autostart function restarts the unit after a power failure.
Affects risk assessment for unattended processes and restart behavior after power loss.
Source
“Autostart function for power failure”
Installation
The degree of protection is IP20, and the ambient temperature must be between 5 °C and 40 °C.
Defines the permissible installation environment and protection against ingress.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
Space requirements must be respected per the operating conditions.
Prevents installing the unit in a location without adequate clearance for operation and ventilation.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”