The unit requires cooling water with a differential pressure of at least 3 bar and no more than 6 bar, so the facility water supply must be verified before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the water-cooled refrigeration circuit.
Source
“min. cooling water differential pressure
3 bar
max. cooling water pressure
6 bar”
Installation
The chiller is designed for cooling water intake between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the dew point.
Cooling water outside this range reduces cooling capacity or causes condensation, affecting reliability and requiring pipe insulation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 1.2 kW at 15°C, 0.9 kW at 0°C, 0.7 kW at -10°C and 0.35 kW at -20°C, so the required cooling load must be checked against the actual working temperature.
Cooling capacity drops significantly at lower temperatures, so selecting the unit requires matching the load at the intended set point.
Source
“Cooling power at ambient temperature +20°C
at 15°C
1,2 kW
at 0°C
0,9 kW
at -10°C
0,7 kW
at -20°C
0,35 kW”
Safety
The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects handling, ventilation and ignition-source control during installation and service.
A3 flammability classification imposes specific safety requirements on the installation location and maintenance procedures.
Source
“Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump option ordered.
Using incompatible media will damage the pump and cause leakage or failure; alternative materials must be specified at purchase.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Installation
The electrical supply must be 208–240 V single or two-phase at 50/60 Hz with a maximum current of 12 A and a 16 A fuse.
The site power connection and protection must be sized to these values to avoid nuisance tripping or inadequate supply.
Source
“Power supply requirement
208-240V 1~/2~ 50/60Hz
max. current
12 A
Fuse
16 A”
Conformité
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Unacceptable supply combinations can cause overheating or malfunction, so the site supply quality must be checked against EN60034-1.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit has IP20 protection and must be installed in an environment between 5°C and 40°C ambient temperature.
IP20 offers no water protection and the ambient limits define the allowable installation location and ventilation.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used to avoid corrosion.
Water chemistry must be compatible with these materials to prevent corrosion and fouling of the water-cooled condenser.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
The pump and cooling water connections are M16x1 male and G1/2 male respectively, which determines the hose and fitting selection.
Correct adapters and hoses must be procured before installation to avoid delays and leaks.
Source
“Pump connection
M16x1 male
Cooling water connection
G1/2 male”
Operation
Whisper mode reduces sound pressure to 51 dB(A) ±4 dB(A) but also reduces pump delivery to 14 l/min and 0.2 bar.
If low noise is required, the reduced pump performance in whisper mode must be acceptable for the process.