Unichiller P012 OLÉ Technical Data Sheet

Sommaire

  1. Overview p. 1
    Product name and cover page for Unichiller P012 OLÉ
  2. Spec p. 2
    Technical data according to DIN 12876 with table

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller uses flammable refrigerant R-290 (A3, H220), so the installation must respect flammable-refrigerant safety requirements.

A3 refrigerant classification affects room ventilation, ignition-source restrictions and local approval for installation.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Selection

The plastic circulation pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant pump ordered.

Media compatibility determines whether the standard unit can be used or a special pump is required.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.

Correct heat-transfer fluid prevents freezing and damage when operating down to -20 °C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The unit requires a minimum ambient temperature of 5 °C and a maximum of 40 °C.

Site ambient conditions must be checked before installation to keep the chiller within its operating envelope.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises.

Sizing must account for actual room temperature, otherwise available cooling capacity will be lower than the datasheet values.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and mismatched combinations are not allowed.

Supply quality must be verified; an out-of-tolerance combination can void compliance with EN60034-1.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to avoid rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug.

The installer must provide and fit the appropriate plug or connection for three-phase units.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has IP20 protection, so it must be installed in a dry location protected from water ingress.

IP20 limits where the chiller can be placed and requires protection against splashing water and dust.

Source
“Degree of Protection IP20”
Conformité

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require EMC measures and is not intended for residential environments without assessment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The optional Pt100 sensor connection only displays process temperature and cannot be used for control.

Users needing process-temperature control must not rely on this option and should select a different control configuration.

Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Adequate clearance is needed for air-cooled operation and service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”