The unit is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
Source
“Operating temperature range -20...100 °C”
| Parameter | Value |
|---|---|
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S508491 |
| Parameter | Value |
|---|---|
| Operating temperature range | -20...100 °C |
| temperature set point / display | digital |
| Internal temperature sensor | Pt100 |
| Resolution of display | 0,1 K |
| Interface digital | USB (Device), RS232 |
| Interface | |
| Temperature stability at -10°C | 0,2 K |
| Alarm message | optic, acoustic |
| Safety classification | III / FL |
| Heating power at 240V | 2,1 kW |
| Heating power at 230V | 2 kW |
| Heating power at 220V | 1,8 kW |
| Cooling power at 15°C | 1,5 kW |
| Cooling power at 0°C | 1 kW |
| Cooling power at -10°C | 0,7 kW |
| Cooling power at -20°C | 0,3 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Refrigerant quantity | 0,35 kg |
| CO2 equivalent | 0,5 t |
| Circulation pump | B |
| at 0,5 bar | 21 l/min |
| at 1,0 bar | 17 l/min |
| at 1,5 bar | 11 l/min |
| at 2,0 bar | 6 l/min |
| max. delivery | 25 l/min |
| max. delivery pressure | 2,5 bar |
| Pump connection | G3/4 male |
| min. filling capacity | 3,8 l |
| expansion tank | 1,7 l |
| Overall dimensions WxDxH ** | 420x487x579 mm |
| Net weight | 65 kg |
| sound pressure level +/- 4 dB(A) | 63 dB(A) |
| Power supply requirement | 220-240V 1~ 50Hz |
| max. current | 15 A |
| Fuse | 16 A |
| Pressure equipment category | Art. 4.3 PED |
** Overall dimensions WxDxH
The unit is rated for an operating temperature range of -20 to 100 °C, so the process must stay within these limits.
Defines the application window and excludes processes outside this range.
“Operating temperature range -20...100 °C”
Cooling capacity drops from 1.5 kW at 15 °C to 0.3 kW at -20 °C, so low-temperature duty needs careful heat-load sizing.
Prevents undersizing the chiller for cold set points.
“at 15°C 1,5 kW at 0°C 1 kW at -10°C 0,7 kW at -20°C 0,3 kW”
The chiller is air-cooled and must be installed with sufficient space for air flow and service access.
Air-cooled units need clearance; ignoring this causes overheating and reduced capacity.
“Please respect space requirements. See operating conditions at www.huber-online.com”
Ambient temperature must remain between 5 °C and 40 °C, and capacity may fall if ambient rises.
Site conditions directly affect cooling performance and reliability.
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
The power supply must be 220–240 V single-phase 50 Hz with a 16 A fuse and 15 A maximum current.
Determines electrical supply, breaker sizing and cabling for installation.
“Power supply requirement 220-240V 1~ 50Hz max. current 15 A Fuse 16 A”
Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions.
Prevents installation on unstable mains that could damage the compressor or pump.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Single-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.
Avoids missing or wrong plug on delivery and delays at commissioning.
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
The refrigerant is R-449A with GHS hazard H280, so leak and ventilation precautions apply.
Affects site safety assessment and refrigerant handling requirements.
“R-449A (A1, H280)”
The unit is EMC Class A Group 1, which is intended for industrial environments and may cause radio interference in residential areas.
Determines whether the installation location is electrically compatible.
“Classification (disturbance) to EN55011: Class A, Group 1”
The pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so system pressure drop must be checked.
Ensures the pump can overcome the external circuit pressure drop.
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
The pump connection is G3/4 male, so mating hoses and fittings must match this thread.
Prevents mismatch of process connections during installation.
“Pump connection G3/4 male”
Minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, which sets the heat-transfer fluid charge.
Needed to prepare the correct fluid volume and expansion allowance.
“min. filling capacity 3,8 l expansion tank 1,7 l”
The unit has an autostart function for power failure, so it may restart automatically after mains recovery.
Affects risk assessment and process safety after power interruptions.
“Autostart function for power failure”
A drain valve is optional, so draining for service or fluid change requires ordering it separately.
Impacts maintenance planning and fluid replacement procedure.
“drain valve #6839”