Technical Reference Summary

Covered Models & Document Scope

Espectrofotómetro UV-VIS UV-1900i; la sección del documento abarca la cuantificación de ácidos nucleicos (medición de trazas de ADN bicatenario mediante celdas TrayCell y Nano Stick).

Technical Data

Parameter Value Condition
Sample volume 10 mm standard cell = 2.5 to 4.0 mL; TrayCell = 0.7 to 10 μL; Nano Stick = 2 μL min. p.3 ↗ UV-1900i
Wavelength range 190 to 1,100 nm p.3 ↗ UV-1900i
Spectral bandwidth 1 nm p.3 ↗ UV-1900i
Light source 20 W halogen lamp and deuterium lamp p.3 ↗ UV-1900i
Monochromator LO-RAY-LIGH grade blazed holographic grating in Czerny-Turner mounting p.3 ↗ UV-1900i
Detector Silicon photodiode p.3 ↗ UV-1900i
Sample compartment internal dimensions W 110 × D 250 × H 115 mm p.3 ↗ UV-1900i
Distance between light beams 100 mm p.3 ↗ UV-1900i
Dimensions W 450 × D 501 × H 244 mm p.3 ↗ UV-1900i
Weight 16.6 kg p.3 ↗ UV-1900i
Output device USB memory (optional); Extended memory (optional) p.3 ↗ UV-1900i
Display 24-bit color touch screen p.3 ↗ UV-1900i
Scan speed up to 29,000 nm/min p.3 ↗ UV-1900i

Operating Limits

  • Minimum sample volume with Nano Stick: 2 μL min. p.3 ↗
  • Minimum sample volume with TrayCell: 0.7 μL p.3 ↗