Contents

  1. pp. 3-4
    Document contents and section listing
  2. Warning pp. 5-9
    Safety notices, symbols, and technical support contacts
  3. Overview pp. 10-13
    System overview, components, and plate/tube identification
  4. Procedure pp. 14-16
    Mounting sampler to flow cytometer and calibration steps
  5. Overview pp. 17-34
    Software interface, settings, and SIP movement offset
  6. Procedure pp. 35-47
    Plate selection, active sample setup, and acquisition workflow
  7. Procedure p. 48
    Cleaning procedures for NovoSampler Pro
  8. pp. 49-55
    Error codes, symptoms, and corrective actions
  9. pp. 56-57
    Supplementary reference information and tables
  10. p. 50 table, 15 rows
  11. p. 54 table, 7 rows
  12. p. 52 table, 10 rows
  13. p. 53 table, 13 rows
  14. p. 28 table, 6 rows
  15. p. 57 table, 14 rows
  16. p. 55 table, 4 rows
  17. p. 51 table, 9 rows
  18. p. 27 table, 6 rows
  19. p. 3 table, 40 rows

Troubleshooting

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x0001Collision of SIPDirty SIP or SIP wash apparatusClean the SIP or SIP wash apparatus following the procedures described in Section 1.1 Preventative Maintenance in NovoCyte® Flow Cytometer Maintenance Guide.
0x0002Running out of NovoFlowNovoFlow solution is not sufficient to continue to run any samplesRefill the NovoFlow container following the procedures described in Section 2.1.2 Add Instrument Reagents in NovoCyte® Flow Cytometer Operator’s Guide.
0x0002NovoCyte Fluidics Station is not working properlyReplace the NovoCyte Fluidics Station.
0x0003Running out of NovoRinseNovoRinse solution is not sufficient to continue to run any samplesRefill the NovoRinse container following the procedures described in Section 2.1.2 Add Instrument Reagents in NovoCyte® Flow Cytometer Operator’s Guide.
0x0003NovoCyte Fluidics Station is not working properlyReplace the NovoCyte Fluidics Station.
0x0004Running out of NovoCleanNovoClean solution is not sufficient to continue to run any samplesRefill the NovoClean container following the procedures described in Section 2.1.2 Add Instrument Reagents in NovoCyte® Flow Cytometer Operator’s Guide.
0x0004NovoCyte Fluidics Station is not working properlyReplace the NovoCyte Fluidics Station.
0x0005Waste container is fullWaste container is too full to continue to run any samplesEmpty the waste container following the procedures described in Section 2.1.3 Empty Waste in NovoCyte® Flow Cytometer Operator’s Guide.
0x0005NovoCyte Fluidics Station is not working properlyReplace the NovoCyte Fluidics Station.
0x0006System voltage is out of rangeSystem Error►►Restart the NovoCyte instrument. ►►Contact ACEA technical support if the error persists.
0x0007System electric current is out of rangeSystem Error►►Restart the NovoCyte instrument. ►►Contact ACEA technical support if the error persists.
0x0008Firmware configuration errorSystem Error►►Restart the NovoCyte instrument. ►►Contact ACEA technical support if the error persists.
0x0009
0x000A
0x000B…… laser self test errorSpecified laser is not functioning properlyThe instrument will automatically reset the laser and run a laser self-test. It takes approximately 5 to 10 minutes. ►►►

Troubleshooting

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x0106NovoSampler (Pro) is disconnected when powered upThe NovoSampler (Pro) is disconnected when powered up►Shut down NovoCyte Flow Cytometer. ►Reconnect the cable of the NovoSampler (Pro) to NovoCyte flow cytometer. ►Turn on NovoCyte and follow the prompts to calibrate the NovoSampler (Pro). ►Contact ACEA technical support if the error persists.
0x0109NovoCyte Fluidics Station is not connectedCable between the NovoCyte Fluidics Station and the NovoCyte instrument is not properly connected►Turn off the instrument. ►Reconnect the cable between the NovoCyte Fluidics Station and the NovoCyte instrument. ►Power up the instrument. ►Contact ACEA technical support if the error persists.
0x010A 0x010B 0x010C 0x010D… liquid level sensor failureSpecified liquid level sensor in the NovoCyte Fluidics Station or NovoCyte Fluidics Cart is not working properly►Reconnect the NovoCyte Fluidics Station or NovoCyte Fluidics Cart cable. ►Restart the NovoCyte instrument. ►Replace the NovoCyte Fluidics Station or NovoCyte Fluidics Cart. ►Contact ACEA technical support if the error persists.
0x010FSheath filter is clogged. Please replace the sheath filter and run the Priming procedureSheath in-line filter is cloggedReplace the sheath in-line filter following the procedures described in Section 1.2 Replacing Fluidic System Consumables in NovoCyte® Flow Cytometer Maintenance Guide.
0x0110Recovering collision errorPlate stops at an incorrect positionNo action is needed. Instrument will automatically recover the error.
0x0111Orbital shaker homing position reset failureOrbital shaker of NovoSampler Pro is not working properlyNo action is needed. The NovoSampler Pro can be used normally.
0x7000NovoCyte Fluidics Cart firmware errorNovoCyte Fluidics Cart firmware is not working properly►Restart the instrument ►Replace the NovoCyte Fluidics Cart ►Contact ACEA technical support if the error persists.

Troubleshooting

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x0014Pressure is out of limitWaste container is not correctly connected to the instrumentCheck the quick coupling connectors to ensure the waste container is correctly connected to the instrument.
0x0014Pressure is out of limitSheath in-line filter is cloggedReplace the sheath in-line filter following the procedures described in Section 1.2 Replacing Fluidic System Consumables in NovoCyte® Flow Cytometer Maintenance Guide.
0x0014Pressure is out of limitFlow cell is cloggedConduct Unclog from NovoExpress.
0x0015NovoSampler firmware errorNovoSampler or NovoSampler Pro firmware is not working properlyRe-install or upgrade the NovoSampler or NovoSampler Pro firmware.
0x0016 0x0017 0x0018…… laser does not emitSpecified laser is not connected properly or laser is not working properly►Reconnect the laser cable properly. ►Restart the instrument. ►Contact ACEA technical support if the error persists.
0x0019 0x001A 0x001B…… laser communication errorInterference or bad connection or bad laser►Reconnect the laser properly. ►Restart the instrument. ►Contact ACEA technical support if the error persists.
0x001CSample injection probe reset failedBad connection or optocoupler damaged►Click the OK button in the prompted dialog box or wait 10 seconds for automatically error handling. ►Restart the instrument. ►Contact ACEA technical support if the error persists.
0x001DSampling Pump reset failedBad connection or optocoupler damaged►Restart the instrument. ►Contact ACEA technical support if the error persists.
0x001EThe SIP module needs to be upgraded to support NovoSampler ProSIP module is not compatible with connected NovoSampler ProContact ACEA technical support for service.
0x0020System initialization is pausedLiquid level in the reagent containers is not within normal range when NovoCyte is powered up.►Make sure that the instrument reagent containers are placed correctly and the liquid level in the containers is within the normal range. ►Click OK on the prompted dialog to continue system initialization.

Troubleshooting

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x0021Sheath fluid pump reset failedBad connection or optocoupler damaged► Restart the instrument. ► Contact ACEA technical support if the error persists.
0x0022Resetting NovoSampler Pro to zero position failedNovoSampler Pro is not working properly.Click Reset. Restart the NovoSampler Pro.
0x0023Resetting NovoSampler Pro to home position failedNovoSampler Pro is not working properly.Click Reset. Restart the NovoSampler Pro.
0x0100Instrument cover openedInstrument cover is open or not tightly closedClose the instrument cover.
0x0101NovoFlow running lowNovoFlow solution is below the pre-set volume limitRefill the NovoFlow container following the procedures described in Section 2.1.2 Add Instrument Reagents in NovoCyte® Flow Cytometer Operator’s Guide.
0x0101NovoFlow running lowNovoCyte Fluidics Station not working properlyReplace NovoCyte Fluidics Station.
0x0102NovoRinse running lowNovoRinse solution is below the pre-set volume limitRefill the NovoRinse container following the procedures described in Section 2.1.2 Add Instrument Reagents in this guide.
0x0102NovoRinse running lowNovoCyte Fluidics Station not working properlyReplace NovoCyte Fluidics Station.
0x0103NovoClean running lowNovoClean solution is below the pre-set volume limitRefill the NovoClean container following the procedures described in Section 2.1.2 Add Instrument Reagents in NovoCyte® Flow Cytometer Operator’s Guide.
0x0103NovoClean running lowNovoCyte Fluidic Station is not working properlyReplace NovoCyte Fluidics Station.
0x0104Waste container is close to fullWaste is above the pre-set volume limitEmpty the waste container following the procedures described in Section 2.1.3 Empty Waste in NovoCyte® Flow Cytometer Operator’s Guide.
0x0104Waste container is close to fullNovoCyte Fluidics Station not working properlyReplace NovoCyte Fluidics Station.
0x0105Cover of NovoSampler (Pro) openedCover of NovoSampler (Pro) is openedClose the cover.

Plate Manager Panel

IconDescription
Mix: Sets the mixing conditions. Mix can be set for None (no mixing), 1 Cycle, 2 Cycles, 3 Cycles for every defined number of wells or every defined number of minutes. Default mixing conditions (i.e. Mix 1 Cycle Every 1 Well) will be automatically applied and cannot be modified by user when Standard or HT mode is selected.
Rinse: Sets the rinse cycle to clean the sample injection probe. Rinse can be set for None (no rinse), 1 Cycle, 2 Cycles, 3 Cycles for every defined number of wells. Default rinse conditions (i.e. Rinse 1 Cycle Every 1 Well when Standard mode is selected, or None when HT mode is selected) will be automatically applied and cannot be modified by user when Standard or HT mode is selected. There is always one rinse cycle at the end of each plate run even the Rinse setting is set to None.
Mixing Parameters: Sets the mixing parameters for the orbital shaker. When NovoSampler Pro works in Custom mode, these mixing parameters for each plate type are defined in the Manage Plate Type window from the Instrument → NovoSampler Pro menu, and can be modified in this area for the current experiment. When NovoSampler Pro works in Standard or HT mode, default mixing parameters will be automatically applied and cannot be modified by user. Please refer to Table 2.1 in this guide for more details.
Speed: The mix speed of the orbital shaker.
Acceleration: The acceleration time for the orbital shaker to reach the specified Mix Speed.
Duration: The mixing time for the orbital shaker to mix the sample in one cycle.

Technical Specifications

ParameterSpecification
NovoSampler Pro Dimension (W × D × H)14.6 × 10.2 × 11.0 in (37 × 26 × 28 cm)
NovoSampler Pro Weight19.8 lb (9 kg)
NovoCyte Flow Cytometer with NovoSampler Pro installed: Work space requirements (W × D × H)186 × 80 × 80 cm
Operating temperature15 °C ~ 32 °C
Operating Humidity< 80% and non-condensing
Power requirements100 ~ 220 VAC, 47 ~ 63 Hz, ≤ 20 W
CompatibilityCompatible with standard 96-well plates (flat-bottom, U-bottom, V-bottom), standard 24-well, 48-well plates, and NovoSampler 24 tube rack for 12 × 75 mm tubes
Carryover rateless than 0.1% (with one rinse cycle per well)
Mix modeorbital shaker
Rinseautomatically rinse the sample injection probe and fluidic tubing.
Minimum sample volume90 µL for 96-well plates
Dead volume50 µL for 96-well plates
Operating SystemMicrosoft Windows®
SoftwareNovoExpress

Table 5-2 NovoExpress Prompted Error Messages and Message IDs

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x7001NovoFlow liquid level sensor is not connectedNovoFlow liquid level sensor is not connected or does not work properly►Reconnect the NovoFlow liquid level sensor ►Contact ACEA technical support if the error remains after reconnection. The sensor may be damaged and need to be replaced.
0x7002Waste liquid level sensor is not connectedWaste liquid level sensor is not connected or does not work properly►Reconnect the NovoFlow liquid level sensor ►Contact ACEA technical support if the error remains after reconnection. The sensor may be damaged and need to be replaced.
0x7100NovoCyte Fluidics Cart communication lostNovoCyte Fluidics Cart is not properly connected to the instrument►Ensure the DVI cable between NovoCyte Fluidics Cart and the instrument is properly connected ►Restart the instrument ►Contact ACEA technical support if the error persists.
Communication error (code: xx, xx). Please restart NovoCyte and NovoExpress.USB cable between the NovoCyte instrument and workstation is not connected properly►Reconnect the USB cable between the NovoCyte instrument and the workstation. ►Restart the instrument and NovoExpress software. ►Contact ACEA technical support if the error persists.

Troubleshooting

Message IDsSoftware MessagesPossible CausesRecommended Solutions
0x000C 0x000D 0x000E….laser is not connectedSpecified laser is not detected►Restart the NovoCyte instrument. ►Contact ACEA technical support if the error persists.
0x000FNovoSampler communication lostThe cable between the NovoSampler (Pro) and NovoCyte Flow Cytometer is not securely connectedReconnect the cable between the NovoSampler (Pro) and NovoCyte Flow Cytometer.
0x000FNovoSampler communication lostNovoSampler (Pro) is not communicating with NovoCyte►Restart the NovoCyte instrument. ►Contact ACEA technical support if the error persists.
0x0010NovoSampler (Pro) has not been calibratedNovoSampler (Pro) is newly installed or re-connected.Follow the prompted instructions to calibrate the NovoSampler (Pro).
0x0011NovoSampler (Pro) calibration failed……NovoSampler (Pro) is not installed properlyRe-install and calibrate the NovoSampler (Pro) following the procedures described in Section 1.2 Installation in NovoSampler® Operator’s Guide or NovoSampler® Pro Operator’s Guide.
0x0011NovoSampler (Pro) calibration failed……The NovoSampler (Pro) cover was opened during calibration.Close the NovoSampler (Pro) cover and redo the calibration.
0x0012The movement of plate is out of rangeThe movement of the orbital shaker is blocked.►Check the path of the orbital shaker to make sure there are no objects blocking the movement. Clear the block if there is any. ►Click Instrument → NovoSampler Pro → Calibrate in NovoExpress Software to re-calibrate the NovoSampler (Pro).
0x0012The movement of plate is out of rangeNovoSampler (Pro) is not installed properlyRe-install and calibrate the NovoSampler (Pro) following the procedures described in Section 1.2 Installation in NovoSampler® Operator’s Guide or NovoSampler® Pro Operator’s Guide.
0x0013Cover of NovoSampler (Pro) is opened during moving plateThe cover of NovoSampler (Pro) is open during moving plateClose the NovoSampler (Pro) cover. NovoSampler (Pro) will automatically reset and be ready for operation.

Plate Manager Panel

IconDescription
Duplicate Sample(s) on Selected Well(s): Creates new samples on empty wells, duplicating the template of the selected samples of the selected wells. If part of the selected wells have samples defined and the rest part have no samples defined, or the selected samples of the selected wells are in different specimen, this tool will be disabled. If the specimen contains more than one sample, and all the samples are selected, clicking this icon will create duplicated samples under a newly created specimen.
New Plate: Creates new plate and specimen, and creates new samples in the new specimen, duplicating the template of the selected samples of the selected wells. The new samples have the same well ID with the selected wells.
Show Run Order: When selected, a blue outline shows on the icon . The sample acquisition sequence will be shown in the plate view.
Mode: User is allowed to select the working mode (Standard, High- Throughput and Custom) of the NovoSampler Pro in this area. Default mixing parameters (i.e. Mixing Speed, Acceleration, Duration, Mix and Rinse pattern), Stop condition and Flow Rate setting will be automatically applied when Standard or HT mode is selected. User can only customize these settings when Custom mode is selected. Selecting the working mode of NovoSampler Pro is only available before the sample is created.
This area is designed to input plate information. The left drop-down box is the Plate ID. The Plate ID is a positive integer and can be modified by highlighting the area and entering a new number. The middle drop-down box is the Plate Type. The drop-down box lists all the available plate type defined in the Manage Plate Types window from the Instrument → NovoSampler Pro menu. Pull down the list and select the corresponding plate type for the experiment. The right box is the Plate Name. Input plate name in this box to help identify the plate. The plate type cannot be modified once the sample acquisition has begun.
This area is designed to input the specimen and sample information. The left drop-down box is the Specimen ID and Specimen Name, separated by a colon. The right box is the Sample Name. When creating new samples, the samples will be created in a specimen with specimen name specified in the Specimen Name box. If the Specimen Name box is empty or does not exist, a new specimen will be created. The default new specimen name is Specimen1, Specimen2,… The default sample name is Sample1, Sample2,…

Table of Content

РазделНазваниеСтраница
Prologue5
I.Revision History5
II.Contact Information5
III.Trademark6
IV.Intended Use6
V.Using this Guide6
VI.Conventions6
VII.Safety and Limitations7
VIII.Technical Support9
1.Introduction10
1.1Hardware Overview10
1.2Installation13
2.Software Overview17
2.1Software User Interface17
2.2Menu Bar18
2.3Cytometer Control Panel22
2.4Experiment Manager Panel23
2.5Plate Manager Panel24
2.6Work List31
2.7Statistical Tables33
2.8Options34
3.Sample Acquisition35
3.1Preparation35
3.2Experiment Setup36
3.3Instrument QC Test37
3.3.1Prepare QC Particles37
3.3.2Run the QC Test37
3.4Single Sample Acquisition38
3.5Edit Work List40
3.5.1Creating Compensation Control Samples40
3.5.2Creating Experiment Samples40
3.5.3Plate View43
3.6Automatic Sample Acquisition43
3.7Analysis of Data46
3.8Shutdown47
4.Cleaning NovoSampler Pro48
5.Troubleshooting49
Appendix56
Appendix AAccessories and Consumables56
Appendix BTechnical Specifications57

Engineering Facts

14 facts · every claim linked to a page
Selection

Only the NovoSampler 24 tube rack is compatible with the NovoSampler Pro; other rack types can cause malfunction and damage.

Prevents selecting an incompatible rack that would damage the instrument.

Source
“Only NovoSampler 24 tube rack could be used with the NovoSampler Pro.”
Installation

The NovoSampler Pro must be flush against the NovoCyte on the left and back sides; gaps cause calibration failure or probe collision.

Ensures correct alignment and avoids probe collision during installation.

Source
“Make sure there is no gap between the NovoSampler Pro and NovoCyte on both left and back sides.”
Installation

Tighten the mounting screws only finger-tight; excessive force damages the instrument.

Prevents thread or housing damage during mounting.

Source
“Use appropriate force (i.e., finger tight) when tightening the screws.”
Installation

Calibration must be performed successfully before any sample collection is allowed.

Defines the mandatory commissioning step before first use.

Source
“Calibration can be postponed, but the software will not allow for samples to be collected until a calibration is performed successfully.”
Installation

Remove all plates from the orbital shaker before calibration to avoid collision and instrument damage.

Prevents probe collision during the calibration routine.

Source
“Before calibrating the NovoSampler Pro remove any plates in the orbital shaker.”
Operation

The volume stop condition must be set when using the NovoSampler Pro.

Required acquisition setting for correct sampler operation.

Source
“The volume stop condition must be set when using NovoSampler Pro.”
Operation

Run Single Well does not perform mixing automatically; the user must click Mix Sample if mixing is needed.

Affects sample suspension and data quality in single-well mode.

Source
“When Run Single Well is selected, the NovoSampler Pro does not perform the mixing function before sample acquisition.”
Operation

The plate type cannot be changed once sample acquisition has started.

Requires correct plate type selection before starting a run.

Source
“The plate type cannot be modified once the sample acquisition has begun.”
Operation

Enabling the plate-type reminder is recommended to avoid probe collision from a wrong plate selection.

Reduces risk of probe collision and instrument damage.

Source
“It is highly recommended to enable this setting to remind user to check and make sure the correct plate type is selected for the experiment to avoid collision of the sample injection probe.”
Safety

The cover must remain closed during sample acquisition; the lid sensor is a backup, not a substitute for confirming the instrument has stopped.

Protects the operator from moving parts during operation.

Source
“ensure that the cover to the NovoSampler Pro is closed during sample acquisition.”
Safety

Do not reach into the NovoSampler Pro while it is running; the sample injection probe is a moving part that can cause injury.

Prevents operator injury from the moving probe.

Source
“The sample injection probe is a moving part and may cause injury.”
Maintenance

The NovoSampler Pro should only be uninstalled when powered off; uninstalling while powered up may damage it.

Prevents hardware damage during removal or service.

Source
“NovoSampler Pro should only be uninstalled when powered off.”
Performance

The automatic rinse function yields a carryover rate of less than 0.1%.

Quantifies cross-contamination performance for assay design.

Source
“the NovoSampler has a carryover rate of less than 0.1%.”
Performance

Prepared QC particles must be used within four hours after preparation.

Sets the time limit for valid QC testing.

Source
“The prepared QC Particles sample should be used within four hours after preparation.”