User management and run log (with errors) with the software
Housing
All materials are recyclable
Table 4: System Editor Screen 1 Options
Screen Section
Parameter
Description
Section Parameter
Description
System Name
Enter a name for your system.
Laboratory
Enter a laboratory name.
Restrict Execution To Specified Server
Click to display the name of the workstation or enter the name of the workstation connected to the system. Click No Restriction to allow the system to be run from any workstation. NOTE: The system must be physically connected.
Change Management
If applicable, select the box under Change Management in order to restrict the system editing rights to the current user. NOTE: This option prevents system configuration modification by anyone other than the system owner or a user with VkModifyOthersSystems privilege.
Table 3: 400-DS Workstation Screen Options
Option
Description
System Configuration
Use this option to configure the system. See System Configuration on page 29.
Diagnostics
Use this option to check the diagnostics of a dissolution apparatus or accessory within the system. See Manual Control/Diagnostics on page 47.
Method Editor
Use this option to create a method, change or delete method parameters, run a report of the method setup, view the method audit trail, and verify the integrity of the method. See Method Editor on page 52.
Run Method
Use this option to run a method. See Running a Method on page 66.
Test Reports
Use this option to run a report of the completed method. See Test Reports on page 70.
Security Change User
To change the user, click Change User. The Logon screen displays. Enter the appropriate user identification and password and click Logon.
Lock Application
To lock the 400-DS Workstation screen, click Lock Application. The 400-DS Workstation Locked screen displays. Click the lock or the link to unlock the screen. The Logon screen displays. Enter the appropriate user identification and password as applicable and click Logon.
Audit Trail
Click Audit Trail. The Security Audit Trail screen displays. Click Show Report to display the report. Change the start date, end date, and/or user identification as desired and click Retrieve Records to change the parameters for the information displayed.
Permissions
Use this option to view permissions assigned to the current user and the descriptions of the corresponding privileges. Click Permissions. The User Group Membership screen displays.
Status
Click one of the pie-shaped graphics under status to view the status of the corresponding dissolution system.
Options in the left pane may be grayed out depending on user permissions.
System Configuration
Screen Section
Parameter
Description
Vial Size
Click the drop-down arrow to indicate the appropriate vial size.
Syringe Size
Click the drop-down arrow to indicate the appropriate syringe size.
Restrict Execution To Specified Server
Manufacturer
Enter the manufacturer of the calibrated temperature probe.
Restrict Execution To Specified Server
Model Name
Enter the model name of the calibrated temperature probe.
Restrict Execution To Specified Server
Serial Number
Enter the serial number of the calibrated temperature probe.
Set Point Temperatures (ºC)
Enter the 3 set points for the cell temperatures. Enter the temperature for each cell at each set point as determined from the diagnostics screen (see Temperature Calibration on page 41).
Volume Calibration Coefficient
Offset
The current offset displays in this area.
Volume Calibration Coefficient
Slope
The current slope displays in this area.
Table 7: Parameters Tab Options
Option
Description
Sample Information
Product Name
Enter the name of the product.
Notes
Enter any relevant notes regarding the sample.
User Defined Labels
Label 1
Label 2
Label 3
The fields in this area have default values of LOT, BATCH, and GROUP. These fields are customizable; enter the information that best serves the needs of the method parameters.
Error Tolerance (±)
Temperature
Enter the desired tolerance. If the tube temperature goes over or under the set temperature (see Verifying the Cell Temperature on page 50) by the amount of this tolerance, an error is recorded as part of the results.
Speed
Enter the desired tolerance. If the RPM goes over or under the set speed (see Verifying Reciprocation on page 51) by the amount of this tolerance, an error is recorded as part of the results.
Profile Interval
Enter the timepoint in hh:mm format at which the temperature and speed settings are recorded.
Change Management
If applicable, select the box under Change Management in order to restrict the method editing rights to the current user or any user with VkModifyOthersMethod.
NOTE: Profile measurements are optional. Values are always recorded at sample timepoints independent of this setting.
Table 8: Automated App7 Sequence Options
Option
Description
Sample Row
Use the up and down arrows to indicate the sample tray row number for which the parameters are being set. After clicking Add to List, use the up and down arrows to indicate the next row. NOTE: The sample row will increment automatically from the previous timepoint.
Media
Use the drop-down arrows to indicate Media A, Media B, Media C, Media D, Media E, or Rinse media.
Sampled Volume (mL)
Use the up and down arrows to indicate appropriate volume based on your vial size.
Dissolution Volume (mL)
Use the up and down arrows to indicate appropriate volume based on the system cell size.
Cell Temperature
Enter the desired tube temperature in degrees Celsius for the first timepoint.
Hold Time (mm:ss)
Enter the desired duration for the sample holder to remain stationary prior to dipping in mm:ss format.
Dip Speed (DPM)
Use the up and down arrows to indicate the desired dips per minute (DPM) for each applicable row of the 400-DS.
Dip Interval (hhh:mm:ss)
Enter the desired duration for dipping in hhh:mm:ss format.
NOTE: The sample row will increment automatically from the previous timepoint.
Table 9: Automated App7 Options
Screen Section
Parameter
Description
Media Description
Media A
Media B
Media C
Media D
Media E
Rinse Media
Enter descriptions as appropriate for the different medias.
Syringe Properties
Aspirate Flow Rate (mL/s)
Use the up and down arrows to indicate the rate at which the media is pulled into the syringe.
Dispense Flow Rate (mL/s)
Use the up and down arrows to indicate the rate at which the media is dispensed from the syringe.
Aspirate Dwell Time (s)
Use the up and down arrows to indicate the duration of time the plunger holds at the top of the stroke before dispensing. NOTE: The time specified is for full syringe volume. For smaller syringe volumes, it is calculated proportionately.
Prime Volume (mL)
Use the up and down arrows to indicate the volume to be pulled before taking a sample.
Syringe Clean Cycles
Use the up and down arrows to indicate the number of times to rinse the syringe between samples.
Sample Cell Clean Cycles
Use the up and down arrows to indicate the number of times to rinse the sample cell between samples.
Needles Clean Cycles
Use the up and down arrows to indicate the number of times to rinse the needles between samples.
Auto Clean After Method Completion
Auto Clean Cycles
Specify the number of cleaning cycles to be performed automatically at the conclusion of each test. Each cycle will flush the specified solution through the entire system - including syringes, sample cells and needles - and then dispense to waste.
Auto Clean Media Source
Indicate the source of the solution to be pumped through the system during for the cleaning cycle. Options include Media A, B, C, D, E and Rinse Port.
Table 10: Method Start Options
Screen Section
Parameter
Description
Sample Information
Product Name
Enter the product name.
Notes
Enter any appropriate notation.
Label 1
Enter the appropriate information based on the user-defined labels (see Creating a Method on page 53).
Label 2
Label 3
Active Channels
Enter the number of active heating channels. NOTE: Heating channels are counted from left to right.
Injection Control
Enable Injections
N/A
Temperature
Delayed Start
Vessel Temperature Start
N/A
Bath Temperature Start
N/A
Time Delayed Start
Use this option to program a delayed start. Enter the desired date and time to start the method.
Engineering Facts
14 facts · every claim linked to a page
Selection
The instrument is intended for small-volume dissolution testing of medical devices such as drug-eluting stents, so it should be selected only for that application.
Confirms the intended application before purchase and prevents selecting it for unsuitable dosage forms.
Source
“The Agilent 400-DS Dissolution Apparatus (G7975A) is the first compendial small-volume dissolution apparatus for testing medical devices such as drug-eluting stents (DES).”
Performance
The system can test up to 13 samples simultaneously, or 12 samples plus a control or standard.
Determines throughput and whether the lab's sample batch size fits the instrument capacity.
Source
“Saves time by simultaneously testing up to 13 samples, or 12 samples and a control or standard, while providing users direct visibility of each dissolution cell.”
Installation
The instrument requires additional clearance of 10 cm on each side, about 20 cm at the rear, and at least 40 cm above the unit.
Bench planning must account for these clearances for air circulation and access to connections.
Source
“It needs an additional 10 cm (4 inches) of space on either side and approximately 20 cm (8 inches) in the rear for air circulation and easy access to liquid and electric connections. Ensure a minimum clearance of 40 cm (16 inches) above the unit.”
Installation
The instrument must be operated in a fume cupboard or a well-ventilated room on a stable, level, non-flammable bench.
Ventilation and bench suitability are site prerequisites for safe installation.
Source
“Operate the instrument in a fume cupboard or in a well ventilated room.”
Safety
The power connector must remain easily reachable and unpluggable for emergency disconnection.
Installation layout must preserve access to the power plug as the primary disconnect.
Source
“Make sure the power connector of the instrument can be easily reached and unplugged.”
Installation
Only the Agilent-supplied power cord may be used, and it must not be used for other equipment.
Prevents electric shock or equipment damage from an incorrect or misused power cord.
Source
“Use only the power cord supplied by Agilent with this instrument.”
Safety
The instrument must not be operated from a power outlet without a protective ground connection.
Grounding is mandatory to avoid electric shock or short circuit.
Source
“Do not operate the instrument from a power outlet that does not have a protective ground connection.”
Operation
The 400-DS may be used with up to 100% concentrations of methanol and ethanol; other organic solvents have not been tested or validated.
Media and solvent selection must be limited to validated solvents to avoid unsafe or unvalidated operation.
Source
“The 400-DS (G7975A) may be used with up to 100% concentrations of Methanol and Ethanol. Other organic solvents have not been tested/validated for use.”
Installation
For a single-system setup, the loop back switch to the right of the right serial port must be set to the left (ON).
Correct switch position is required for proper serial communication in a single-system configuration.
Source
“Ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON).”
Installation
In a multiple-system setup, the final 400-DS in the chain must have its loop back switch set to the left (ON), while non-final units must be set to the right (OFF).
Daisy-chain termination depends on correct switch settings for each unit in the series.
Source
“If this is the final Agilent 400-DS in the series, ensure the loop back switch (located to the right of the right serial port) is switched to the left (ON).”
Operation
Clicking Fill more than once without purging will overflow the cells.
Prevents media overflow and potential spillage during manual media control.
Source
“Do not click Fill more than one time without purging in between. The media will overflow the cells.”
Operation
When fewer than 13 active channels are used, sample vials must be present in positions 1 and 13 of each sample tray row.
Proper sensing, alignment, and balance of the sample tray depend on vials in these positions.
Source
“Ensure sample vials are present in positions 1 and 13 of each sample tray row if less than 13 active channels are used.”
Maintenance
The waste container must have enough free volume and its filling level must be checked regularly.
Waste handling capacity and regular checks are needed to avoid spills during operation.
Source
“The residual free volume in the appropriate waste container must be large enough to collect the waste liquid.”
Compliance
The software is 21 CFR Part 11 compliant and a single PC can independently control up to four 400-DS systems.
Supports regulated-environment selection and central control of multiple units from one workstation.
Source
“21 CFR Part 11 compliant software and a single PC can independently control up to four 400-DS systems.”