Contents

  1. Overview pp. 3-4
    Purpose, audience, and conventions of the operating manual
  2. pp. 5-6
    Table of contents listing chapters and topics
  3. Procedure pp. 7-10
    Conditioning Intuvo capillary columns and PCI autotune steps
  4. Procedure pp. 11-12
    Connecting and disconnecting EI HES model wiring to CI
  5. Overview pp. 13-14
    Version identification and model overview table
  6. Overview pp. 15-18
    System introduction and vacuum gauge description
  7. Warning pp. 19-24
    Electrostatic discharge risks and hydrogen safety warnings
  8. Warning pp. 25-28
    General safety precautions and warning statements
  9. Quality pp. 29-32
    Safety, regulatory, and electromagnetic compatibility certifications
  10. Procedure pp. 33-34
    Steps for replacing the primary fuses
  11. Procedure pp. 35-54
    Column conditioning and capillary column installation in GC/MSD
  12. Procedure pp. 55-76
    Intuvo 9000 column installation, replacement, and gasket procedures
  13. Procedure pp. 77-118
    EI operation, firmware, web user interface, autotune, and venting
  14. Procedure pp. 119-136
    CI autotune, reagent gas control, and CI operation setup
  15. Procedure pp. 137-150
    Tools, spare parts, ion source cleaning, and vacuum maintenance
  16. Procedure pp. 151-168
    Disassembling and assembling EI HES, SS, Inert, and HydroInert sources
  17. Procedure pp. 165-168
    Installing EI HES and connecting/disconnecting wiring
  18. Procedure pp. 197-200
    Replacing electron multiplier horn and closing analyzer chamber
  19. Procedure pp. 201-232
    Removing/installing radiators, switching sources, cleaning CI source, filament service
  20. p. 125 table, 6 rows
  21. p. 36 table, 8 rows
  22. p. 56 table, 7 rows
  23. p. 91 table, 7 rows
  24. p. 170 table, 8 rows
  25. p. 171 table, 11 rows
  26. p. 186 table, 5 rows
  27. p. 187 table, 14 rows
  28. p. 17 table, 10 rows
  29. p. 152 table, 3 rows

Table 16 Default Tune Control Limits, used by CI autotune only

Reagent gasIon polarityAbundance targetPeakwidth targetMaximum repellerMaximum emission current, µAMax electron energy, eV
MethanePositive1x1060.64240240
MethaneNegative1x1060.6450240
IsobutanePositiveN/AN/AN/AN/AN/A
IsobutaneNegative1x1060.6450240
AmmoniaPositiveN/AN/AN/AN/AN/A
AmmoniaNegative1x1060.6450240

Table 5 Gas flows

FeatureGas flowsGas flows
High vacuum pumpDiffusionTurbo
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
CI capabilityNoYes
CI reagent flows, mL/minNA1 to 2
JetClean option H2 flowNA0.4 mL/min

*Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. †Expect degradation of spectral performance and sensitivity.

Table 6 Gas flows

FeatureGas flowsGas flows
DiffusionTurbo
High vacuum pump
Optimal He column flow mL/min11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.53 mm (30 m)0.53 mm (30 m)
JetClean option H2 flowNA0.4 mL/min

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable) plus JetClean H2 flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity.

Table 9 MS Parameters menu

ActionDescription
High Vacuum PressureOnly with Micro-Ion vacuum gauge installed.
Turbo Pump SpeedDisplays the turbo pump speed.
Foreline PressureDisplays the foreline pressure.
MSD Fault StatusReports a summary fault status code (number) in ‘dec’ (decimal) and ‘hex’ (hexadecimal) format covering all possible fault combinations.
Ion Source Temp, oCDisplays and sets the ion source temperature.
Mass Filter Temp, oCDisplays and sets the mass filter temperature.
CI ReagentDisplays CI reagent gas and flow rate (if installed).

MS parameters cannot be set from the LCP while an online OpenLab CDS session is connected to the MSD.

Table 21 Parts list for the EI SS or Inert source (Figure 59)

Item numberItem descriptionPart number (SS)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374

Table 21 Parts list for the EI SS or Inert source (Figure 59) (continued)

Item numberItem descriptionPart number (SS)Part number (Inert)
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 25 Parts list for the EI standard or EI Inert source (Figure 64)

Item numberItem descriptionPart number (SSL)Part number (Inert)
1Gold plated set screwG1999-20022G1999-20022
2Gold plated screwG3870-20021G3870-20021
3Interface socketG1099-20136G1099-20136
4Source bodyG1099-20130G2589-20043
5Drawout cylinderG1072-20008G1072-20008

Table 25 Parts list for the EI standard or EI Inert source (Figure 64) (continued)

Item numberItem descriptionPart number (SSL)Part number (Inert)
6Drawout plate05971-20134G2589-20100
74-turn filamentG7005-60061G7005-60061
8Spring washer3050-13743050-1374
8Flat washer3050-09823050-0982
9Lens insulatorG3170-20530G3170-20530
10Entrance lensG3170-20126G3170-20126
11Ion focus lens05971-2014305971-20143
12Repeller insulatorG1099-20133G1099-20133
13RepellerG3870-60172G3870-60173
14Flat washer3050-06273050-0627
15Belleville spring washer3050-13013050-1301
16Repeller nut0535-00710535-0071
17Source heater block assemblyG3870-60180G3870-60179
18Repeller block insertG3870-20135G3870-20135

Table 3 5977C Series MSD features

FeatureDiffusionTurbo
High vacuum pumpDiffusionTurbo
Optimal He column flow mL/min 11 to 2
Maximum recommended gas flow mL/min*1.54
Maximum gas flow, mL/min†26.5
Max column id0.25 mm (30 m)0.53 mm (30 m)
CI capability‡NoYes
Inert ion sources availableYesYes
GC compatibility9000 and 88909000 and 8890
Foreline pumps availablePfeiffer DuoPfeiffer Duo, Pfeiffer MVP-070-3x, Agilent IDP-3 24V
DIP** capability (3rd party)YesYes

* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable). Based on helium gas use. For other gases the maximum flow will vary. † Expect degradation of spectral performance and sensitivity. ‡ Turbo pump models are field upgradeable to CI. ** Direct insertion probe.

Table 19 Parts list for EI HES (Figure 51)

Item numberItem descriptionPart number (XTR)
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Extractor lens (5)*, with 3 mm openingG7004-20061

Engineering Facts

14 facts · every claim linked to a page
Selection

The diffusion-pump model cannot be upgraded to CI, so CI requirements force selection of a turbo-pump model.

Determines whether the chosen MSD model can later support chemical ionization.

Source
“Turbo pump models are field upgradeable to CI.”
Selection

Diffusion-pump models accept only 0.25 mm columns and lower gas flows than turbo models, constraining column and flow choices.

Column diameter and carrier flow must match the pump type selected.

Source
“Max column id 0.25 mm (30 m)”
Selection

The Intuvo 9000 GC cannot be used with CI on this MSD, so CI users must pair with an 8890 GC.

GC platform choice is constrained when CI operation is required.

Source
“If you are using the Agilent Intuvo 9000 Gas Chromatograph with your MSD, Chemical Ionization (CI) is not currently supported.”
Installation

Columns must be conditioned before connection to the MSD interface to avoid source contamination and sensitivity loss.

Prevents column bleed deposits in the ion source that degrade sensitivity.

Source
“Conditioning a column before it is connected to the GC/MSD interface is essential.”
Installation

Carrier gas purity must be at least 99.9995% to protect the column and MSD during installation and operation.

Gas purity affects column bleed, source contamination, and data quality.

Source
“Always use carrier gas that is at least 99.9995% pure.”
Installation

New ferrules can loosen after thermal cycling, so tightness must be rechecked or self-tightening nuts used.

Prevents leaks at the interface after initial heat cycles.

Source
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Installation

Intuvo guard and jumper chips are single-use consumables that deform on installation and cannot be reused or cleaned.

Spares must be stocked and misinstalled chips replaced, not reinstalled.

Source
“The column Guard Chip and the Jumper Chip are both single-use, consumable parts.”
Operation

After pumpdown reaches 'Okay to run', a two-hour thermal equilibration wait is needed for reproducible data.

Acquiring before thermal equilibrium yields non-reproducible results.

Source
“After the message Okay to run appears, wait 2 hours for the MS to reach thermal equilibrium.”
Operation

CI reagent gas flow is set as a percentage of maximum, recommended 20% for PCI and 40% for NCI.

Sets correct reagent gas flow for PCI versus NCI methods.

Source
“The recommended flow is 20% for a PCI source and 40% for an NCI source.”
Operation

After switching from EI to CI or venting, the MSD must be baked out at least two hours before tuning.

Prevents contamination and failed tunes after mode changes.

Source
“the MSD must be baked out for at least 2 hours before tuning.”
Safety

Hydrogen must be shut off at its source before MSD power is removed, or it can accumulate and explode.

Critical safety interlock when using hydrogen carrier or JetClean gas.

Source
“the hydrogen gas flow must be off before turning off the MSD power.”
Safety

If a power failure occurs while unattended, hydrogen must be shut off and the manifold opened to atmosphere with a 10-minute wait.

Defines the safe restart sequence after an unattended power failure with hydrogen.

Source
“Wait at least 10 minutes to allow any hydrogen to dissipate.”
Maintenance

The quadrupole mass filter needs no periodic maintenance and should generally never be disturbed; cleaning is service-only.

Avoids unnecessary invasive maintenance that could damage the analyzer.

Source
“The mass filter (quadrupole) requires no periodic maintenance.”
Maintenance

The HED ceramic insulator must never be touched during analyzer maintenance.

Prevents damage to sensitive detector components during service.

Source
“The HED ceramic insulator must never be touched.”