Introduction to FTIR microscopic imaging of large samples
Procedurep. 2
Considerations for managing large data file sizes
Procedurep. 3
Processing considerations and comparison table for data handling
Procedurep. 4
Total spectral integrated intensities analysis of FTIR data
p. 5
Comparison of resulting images from the analysis
p. 6
Summary of findings and conclusions
p. 3table, 8 rows
Table 1. Measurement parameters. All mosaics were collected with the same scan parameters: 16 co-additions for the scan, 32 co-additions for background, 4 cm-1 spectral resolution.
Objective
4x
15x
Integration time (µs)
12
8
Attenuation setting
Open
Open
Single FPA tile FOV (µm)
1230x1230
350x350
Nominal pixel size (µm)
19.2x19.2
5.5x5.5
Mosaic config.
7x6
27x19
Total mosaic size (mm)
8.6x7.4
8.4x6.7
Disk usage (GB)
4.3
42.4
Collection time (hh:mm)
1:03
10:22
Engineering Facts
13 facts · every claim linked to a page
Selection
The 4x objective increases the single-tile field of view to 2.4x2.4 mm with a 128x128 FPA, which is about 11 times the area of the 15x configuration.
Determines whether a single-tile survey is feasible or a large mosaic is required for the sample size.
Source
“Using Agilent’s unique 4x IR objective to increase the FOV to provide up to 2.4x2.4 mm in a single tile using a 128x128 FPA.”
Selection
A 128x128 FPA can be installed to keep the 5.5x5.5 µm pixel size with the standard 15x objective while doubling the field of view to 700x700 µm.
Guides detector upgrade choice when spatial resolution must be preserved but a larger area per tile is needed.
Source
“Larger FPAs of 128×128 pixels can be installed to image a FOV of 700×700 µm while maintaining the pixel size at 5.5×5.5 µm with the standard 15x objective.”
Operation
The UDR 4 setting limits computed spectra to 3975 cm-1, so a low-pass optical filter with a cutoff near 3975 cm-1 must be used to prevent aliasing.
Specifies the required optical filter when under-sampling is used, preventing spectral artifacts.
Source
“To avoid aliasing from spectral contribution beyond 3975 cm-1, a low pass optical fi lter was employed with a cut-off at approximately 3975 cm-1.”
Performance
The 15x objective transmits down to about 850 cm-1 while the 4x objective cuts off near 1000 cm-1 because its CaF2 lenses absorb in the low-wavenumber range.
Determines whether the 4x objective can be used when low-wavenumber bands are analytically important.
Source
“The cutoff is at approximately 850 cm-1 for the 15x objective and 1000 cm-1 for the 4x objective.”
Selection
The 4x objective is suited for rapid survey scans, but the 15x objective with 5.5 µm pixels is required to resolve fine details such as blood vessels.
Sets the resolution requirement that dictates objective choice for morphology-level detail.
Source
“The 15x objective with 5.5 µm per pixel was required to resolve details such as blood vessels.”
Selection
The 15x objective should be preferred over the 4x when important chemical information lies below 1100 cm-1, such as phosphate, glycogen, cholesterol, or hydroxyapatite bands.
Prevents loss of key low-wavenumber bands when choosing the wide-field objective.
Source
“it is advisable to use the 15x rather than the 4x objective if the low wavenumber range below 1100 cm-1 contains important chemical information.”
Operation
The 4x objective requires a matched 4x condenser, so a pair of 4x objectives is needed for transmission imaging, while transflection on reflective slides needs only one.
Affects the bill of materials and optical configuration for transmission versus transflection setups.
Source
“the set of two 4x IR objectives—one as condenser and one as objective—is an excellent complement to the standard set of 15x objectives for FTIR imaging.”
Operation
On reflective slides such as MirrIR, a single objective is sufficient in transflection geometry, avoiding the need for a matched condenser.
Simplifies the optical setup and reduces cost when reflective substrates are used.
Source
“For samples on reflective slides (e.g. MirrIR from KevleyTechnologies), one objective is sufficient in transflection geometry.”
Operation
The attenuator and integration times were optimized to the highest integration time before non-linear detector behavior, detected as a non-zero response below the FPA spectral cutoff.
Defines the procedure to maximize signal without saturating or operating the FPA non-linearly.
Source
“The attenuator and integration times were optimized to achieve the highest possible integration time before the onset of non-linear behavior”
Operation
Stage step size and offset must be set for optimum stitching, and the step size equals the field of view of a single FPA tile.
Ensures seamless mosaics without gaps or overlap artifacts during large-area acquisition.
Source
“Stage parameters for step size and offset were set for optimum stitching of images during mosaic acquisition.”
Performance
Binning averages neighboring pixels to improve signal-to-noise ratio at the expense of spatial resolution.
Provides a trade-off option when SNR is marginal but full spatial resolution is not required.
Source
“a procedure called “binning” that averages a defi ned number of neighboring pixels to one pixel at the expense of the spatial resolution, but with the advantage of increased signal-to-noise ratio (SNR).”
Operation
Increasing UDR from 4 to 8 with a matching low-pass filter limits the range to 1975 cm-1, halving data size and speeding processing.
Offers a way to cut storage and processing load when the high-wavenumber region is not needed.
Source
“the UDR 4 setting used in this study could be increased to UDR 8, allowing a spectral range up to 1975cm-1 (with an appropriate low pass optical fi lter to prevent aliasing).”
Operation
Storing each FTIR image as a separate file via the multifile option allows mosaics of virtually unlimited size.
Enables large-area acquisitions without hitting single-file size limits.
Source
“The advantage of this option is that mosaics of virtually unlimited size can be collected.”