Maintenance procedures for ESI, AJS ESI, APCI, and MMI sources
Qualitypp. 137-145
Regulatory compliance, EMC, WEEE, and end-of-life information
p. 146
Agilent website contact reference
p. 31table, 4 rows
p. 104table, 4 rows
Table 2 Tuning mixes
Source
Tuning mix
Part Number
ESI
ESI-L Low Concentration Tuning Mix
G1969-85000
AJS ESI
ESI-L Low Concentration Tuning Mix
G1969-85000
MMI
MMI-L Low Concentration Tuning Mix
G1969-85020
APCI
APCI-L Low Concentration Tuning Mix
G1969-85010
Table 1 Correct needle position
Nebulizer
Distance
Microflow Nebulizer Assembly
0.003 inch
ES Tested Nebulizer
0.000 (flush)
Tested ESI Nebulizer Assembly (older style)
0.003 inch
Tested APCI Nebulizer Assembly
0.000 (flush)
Engineering Facts
14 facts · every claim linked to a page
Safety
The instrument contains magnets that can interfere with pacemakers and other active implanted medical devices, so users with AIMDs must follow manufacturer guidance.
Site safety planning and personnel restrictions must account for strong magnets near the door and front cover.
Source
“This product contains magnets that may affect the functioning of pacemakers, implanted heart defibrillators, or other active implanted medical devices”
Selection
The AJS ESI source uses super-heated nitrogen thermal gradient focusing and can improve sensitivity about five-fold for many small-molecule compounds.
Source selection should weigh the sensitivity gain of AJS ESI against ESI for small-molecule work.
Source
“improves sensitivity 5x or more for many small-molecule compounds that undergo electrospray ionization.”
Selection
The Multimode source can run ESI only, APCI only, or mixed ESI/APCI and supports LC flow rates up to 2 mL/min.
Choosing MMI enables broad compound coverage and compatibility with higher-flow LC methods.
Source
“Simultaneous ESI and APCI capability and high LC flow rate compatibility (up to 2 mL/min)”
Operation
A checktune verifies peak width and mass axis before acquisition, while an autotune is only needed after preventive maintenance or if checktune fails.
This defines the routine tuning workflow and avoids unnecessary autotune downtime.
Source
“An autotune only needs to run after preventative maintenance or if there is a problem with checktune.”
Selection
The tuning mix must match the installed source, with distinct part numbers for ESI, AJS ESI, MMI, and APCI.
Ordering the wrong tuning mix for a source leads to miscalibration and mass assignment errors.
Source
“Use Table 2 to select the correct tuning mix for the source installed.”
Operation
Requesting tune control locks the LC/TQ so that no sample run or worklist can be started during tuning.
Operators must release tune control before scheduling analytical runs.
Source
“When tune has control of the LC/TQ instrument, a single sample run or a worklist cannot be started.”
Safety
The system must always be vented before pressing the power button to prevent instrument damage.
Shutdown procedures must include a full vent step to protect the turbo and vacuum components.
Source
“To prevent instrument damage, always vent the system before pressing the power button.”
Safety
The spray shield and spray chamber components are likely to be hot and must not be touched.
Maintenance and shutdown planning must include cool-down and burn protection.
Source
“Do not touch the spray shield and related spray chamber components. They are likely to be hot.”
Operation
When not acquiring data, the LC stream should be switched away from the MS ion source to waste.
This protects the source from unnecessary sample and solvent exposure during equilibration.
Source
“When not acquiring data, switch the direction of the LC stream away from the MS ion source and to waste.”
Maintenance
The calibrant bottle should only be snug, because over-tightening can prevent pressurization and cause evaporation of the tuning mix.
Correct bottle installation avoids leaks and preserves tuning mix concentration.
Source
“The bottle only needs to be snug. Do not over-tighten the bottle.”
Maintenance
Tuning mixes are not interchangeable, and failing to change the mix when the source or inlet changes can cause miscalibration and mass assignment errors.
Source changeover procedures must include replacing the tuning mix to maintain calibration.
Source
“The tuning mixes are not interchangeable.”
Maintenance
Rough pump oil should be checked quarterly and replaced if it is dark or contains suspended particles.
Scheduled oil inspection prevents pump wear and vacuum instability.
Source
“If the pump oil is dark or full of suspended particles, replace it.”
Installation
Nitrogen gas filters have a maximum operating pressure of 250 PSIG, maximum temperature of 100 °C, and maximum flow of 8 liters per minute.
Gas supply and plumbing must be rated within these limits to avoid filter failure.
Source
“Maximum operating pressure is 250 PSIG.”
Installation
Stainless steel tubing requires purifiers with stainless steel fittings, since brass fittings increase the risk of cross-threading.
Correct fitting material selection prevents leaks and damaged connections during installation.
Source
“Using brass fittings with stainless steel tubing increases the risk of cross-threading.”