Agilent 7100 Capillary Electrophoresis System User Manual
Contents
Overviewpp. 13-18
System overview, learning products, and documentation scope
Specpp. 19-66
Power cords, bench space, dimensions, weight, and performance specifications in multiple languages
Procedurepp. 67-90
Initial setup, initialization mode selection, and manual configuration
Procedurepp. 91-108
Graphical user interface and direct control of instrument components
Procedurepp. 109-156
CE operation, high voltage system, preconditioning, detector parameters, and fraction collection
Procedurepp. 157-168
Operating the instrument in high pressure CE+p mode
Procedurepp. 169-180
CEC operation, preconditioning, and control via CE diagram
Procedurepp. 181-210
Diagnostics, firmware, diode array detector, and fuse replacement
Procedurepp. 211-248
Maintenance overview, EMF, pre-puncher cleaning, lamp replacement, and replenishment system care
Orderingpp. 249-272
Overview of maintenance parts and ordering information
Warningpp. 273-282
Safety symbols, WEEE directive, and legal notices
pp. 283-286
Alphabetical index and document part number
p. 23table, 3 rows
p. 24table, 4 rows
p. 87table, 6 rows
p. 22table, 9 rows
p. 83table, 5 rows
p. 21table, 12 rows
p. 78table, 2 rows
p. 82table, 3 rows
p. 79table, 4 rows
p. 70table, 27 rows
Table 3 Performance specifications for the Agilent 7100 Capillary Electrophoresis System
Type
Specification
Comments
Raw data channels
Detector signals and spectrum, voltage, current, leak current, power, cassette temperature, pressure, lamp voltage and detector temperature
Electrophoresis power
• Voltage range
• Current
• Power
• Operation under constant voltage, current or power
• Programmable polarity switch
setable 0 to ± 30 kV supply
setable 0–300 µA
setable 0–6 W
CE specific software
• Mobility report, time corrected areas, pI calibration and bio polymer size calibration
Table 6 Performance specifications for the Agilent 7100 Capillary Electrophoresis System, Capillary Cassette
Type
Specification
Comments
Capillary cassette
High-speed forced-air cooler with Peltier element
Temperature range
10 °C below ambient, up to 60 °C (min. 10 °C cassette temperature)
Minimum total capillary length
31 cm
Capillary compatibility
365 µm o.d.
Table 11 Telnet commands
Value
Description
?
Displays syntax and descriptions of commands
/
Displays current LAN settings
ip <x.x.x.x>
Sets new IP address
sm <x.x.x.x>
Sets new subnet mask
gw <x.x.x.x>
Sets new default gateway
exit
Exits shell and saves all changes
Table 3 Performance specifications for the Agilent 7100 Capillary Electrophoresis System
Type
Specification
Comments
Safety features
Extensive diagnostics, error detection and display, current leak detection; low current limit
Liquid leak sensor; safety sensors on door and cover disabling high voltage; vial sensor;
Control and data evaluation
Agilent control software (e.g. ChemStation, OpenLAB CDS ChemStation Edition, MassHunter)
Communication
LAN, controller-area network (CAN), RS-232C, APG remote: ready, start, stop and shut-down signals. USB (future use). Analog out and in.
0 to1 V
GLP features
Early maintenance feedback (EMF), electronic records of maintenance and errors
Diagnostic functions
RFID tag for lamp, early maintenance feedback, supported by Agilent Lab Advisor software with integrated diagnosis test suite
Pressure system
• Programmable with -100 to +100 mbar on inlet
• Flushing with 1 bar or with high pressure 2–12 bar
• Vial pressurization with high pressure 2–12 bar on inlet and/or outlet
Injection modes
• Self correcting injection system with injection from inlet or outlet
• Programmable range
• Pressure
• Electrokinetic
up to 10,000 seconds
–100 to +100 mbar
–30 to +30 kV
Replenishment
• Satellite station for refilling of inlet or outlet buffer vials with fresh buffer for automatic, continuous operation.
• Selectable buffer levelling.
Vials
• 100 µL (polypropylene or glas) with resealing snap caps
• 1 mL (polypropylene) with resealing snap caps
• 2 mL (glas) with resealing snap caps
Table 10 Link configuration switches
SW3
SW4
SW5
Link configuration
OFF
Speed and duplex mode determined by auto-negotiation
ON
OFF
OFF
Manually set to 10 Mbps, half-duplex
ON
OFF
ON
Manually set to 10 Mbps, full-duplex
ON
ON
OFF
Manually set to 100 Mbps, half-duplex
ON
ON
ON
Manually set to 100 Mbps, full-duplex
Table 2 Physical specifications
Type
Specification
Comments
Weight
35 kg (77.2 lbs)
Dimensions (width × depth × height)
350 x 510 x 590 mm (13.8 x 20.1 x 23.2 inches)
Line voltage
100–240 VAC ± 10%
Wide-ranging capability
Line frequency
50 or 60 Hz ± 5%
Power consumption
350 VA / 300 W / 1024 BTU/h
Maximum
Ambient operating temperature
5–40 °C (41–104 °F)
Ambient non-operating temperature
-40 to 70°C (-40 to 158 °F)
Humidity
below 80% at 31 °C (87.8 °F)
Non-condensing
Operating altitude
Up to 2000 m (6500 ft)
External cooling
max. 0.5 bar (7.2 psi), max. 50 °C (122 °F)
Waterbath
External pressure
2-12 bar (29-174 psi)
Oil-free air or nitrogen
Safety standards: IEC, CSA, UL
Housing
Installation category II, Pollution degree 2 All materials are recyclable For indoor use only
Table 7
Factory default settings
Initialization ('init') mode
Using default, switches 7 and 8 set to ON.
Link configuration
Speed and duplex mode determined by auto-negotiation
Table 9
Using Default parameters
IP address:
192.168.254.11
Subnet mask:
255.255.255.0
Default gateway
Not specified
In the "Using Default" mode, the parameters stored in the memory of the 7100 Capillary Electrophoresis instrument are not cleared automatically. If not changed by the user, they are still available when you switch back to the mode "Using Stored".
Table 8 Initialization mode switches
SW6
SW7
SW8
Init mode
OFF
OFF
OFF
Bootp
OFF
OFF
ON
Bootp & Store
OFF
ON
OFF
Using Stored
OFF
ON
ON
Using Default
Table 6 Accessories
Description
Part number
Quantity
Accessory kit
G7100-68705
1
– HPCE startup capillary kit
1600-68700
1
– Detector interface, green, 50 µm
G7100-60210
1
– Detector interface, red, extended 50 µm
G7100-60230
1
– Coupler
0100-1633
2
– O-ring 25.12-mm-id 1.78-mm-XSECT-DIA
0905-1163
1
– O-ring 3-mm
0905-1196*
5
– Container
1540-0307
1
– Fuse T6.3AH, 250 VAC
2110-1018
2
– Fuse T6.3AL, 250 VAC
2110-0623
2
– Fuse T3.15AL, 250 VAC
2110-1417
4
– Air filter
3150-0619
1
– Tubing assembly
5063-6527
1
– Screwdriver, slot shaft
8710-2607
1
– Screwdriver, Pozidriv shaft
8710-0899
1
– 12 mm socket wrench
8710-2076
1
– Lift tubing assy
G1600-67003
1
– Vial crimp/snap polypropylene 1 mL 100/PK
5182-0567
1
– Snap caps, polyurethane 100/PK
5181-1512
1
– Cable, remote
5061-3378
1
– Bottle, 500 mL, electrolyte, pressure plus
9301-6531
2
– Bottle, 100 mL, electrolyte, pressure plus
9301-6532
1
– CE System User’s Guide
G7100-90000
1
– Glass filter, solvent inlet
5041-2168
1
– Electrode alignment assy
G7100-60107
1
– LAN cable (crossover)
5023-0203
1
CE Installation Qualification Kit
5063-6514
1
*Part number 0905-1196 is not orderable. Please re-order part number 5062-8544 (5/PK)
Engineering Facts
14 facts · every claim linked to a page
Installation
The instrument requires additional clearance of 2.5 cm on each side and about 8 cm at the rear for air circulation and electrical connections.
Bench space planning must account for these clearances to ensure proper cooling and cable access.
Source
“It needs an additional 2.5 cm (1.0 inch) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.”
Installation
The instrument must be operated in an upright position.
Mounting or placing the instrument in any other orientation is not permitted.
Source
“The instrument must be operated in a upright position.”
Safety
The module remains partially energized when switched off as long as the power cord is plugged in.
Service or maintenance work requires unplugging the power cord to avoid shock hazard.
Source
“The module is partially energized when switched off, as long as the power cord is plugged in.”
Safety
The power plug must remain easily accessible and the cable removable at any time for emergency disconnection.
Installation layout must not obstruct the power plug for emergency disconnection.
Source
“Make sure that it is always possible to access the power plug.”
Installation
For CEC or CE+p mode an external supply of oil-free air or nitrogen with a maximum pressure of 15 bar can be used.
Sites planning CEC or CE+p operation must provide a suitable external gas supply and adapter.
Source
“an external pressure supply of oil-free air or nitrogen with a maximum pressure of 15 bar (218 psi) can be used.”
Operation
The cassette temperature setpoint must be within the specified range, with the lowest applicable temperature being 10 °C below ambient and the highest 60 °C.
Method temperature settings must respect ambient conditions and the 60 °C upper limit.
Source
“The lowest applicable temperature is 10 °C below ambient temperature. The highest temperature is 60 °C.”
Operation
Operating the cassette below the dew point can cause condensation, arcing and damage to the cassette or instrument.
Users must keep cassette temperature above dew point or reduce ambient humidity to avoid damage.
Source
“if the temperature of the cassette is below the dew point condensation might take place.”
Operation
Outlet vials must be filled with enough liquid to touch the capillaries to prevent droplets that can cause pressure and affect injection amount.
Proper outlet vial filling is required for reproducible injections and to avoid contamination.
Source
“Outlet vials (for example, waste vials) should also be filled with enough liquid to touch the capillaries.”
Operation
For replenishment, a capped vial filled with buffer or water should be placed in position 49 to prevent buffer precipitation and needle blockage.
Following this practice avoids blockage of the replenishment needle and maintains reproducibility.
Source
“it is good practice to have a capped vial filled with buffer or water in position 49”
Operation
The tray can only be moved by the Agilent ChemStation when the tray door is closed for safety reasons.
Operators must close the tray door before commanding tray movement to avoid safety issues.
Source
“For safety reasons, the tray can only be moved by the Agilent ChemStation when the tray door is closed.”
Installation
The default ramp up rate of the 7100 Capillary Electrophoresis is 5 kV/s.
Method developers should know the default voltage ramp rate when designing separation methods.
Source
“The default ramp up rate of the 7100 Capillary electrophoresis is 5 kV/s.”
Operation
When using fraction collection, all timetable entries using voltage, outlet vial or pressure settings are deleted from the timetable during the run.
Users must set voltage and ramp time in the appropriate screens because timetable entries are overridden.
Source
“When using fraction collection all timetable entries which use voltage, outlet vial or pressure settings will be deleted from the timetable.”
Installation
The 7100 CE is shipped with switches 7 and 8 set to ON, giving a fixed default IP address of 192.168.254.11.
Network configuration requires knowing the factory default IP and switch settings.
Source
“The 7100 CE is shipped with switches 7 and 8 set to ON. This means that the instrument is set to a fixed default IP: 192.168.254.11”
Installation
To perform any LAN configuration, switches SW1 and SW2 must be set to OFF.
LAN setup cannot proceed unless these switches are correctly configured.
Source
“To perform any LAN configuration, SW1 and SW2 must be set to OFF.”