Abbreviations and overview of the 7200 QTOF system
Overviewpp. 12-14
System components, vacuum gauge, and hardware layout
Warningpp. 15-17
Important safety warnings, explosion and fire precautions
Qualitypp. 18-21
Safety symbols, EMC compliance, and intended use
Procedurepp. 23-31
Column selection, conditioning, and installation procedures
Procedurepp. 33-56
EI operation, pumping down, venting, and temperature control
Procedurepp. 57-74
CI setup, reagent gas flow, and CI autotune procedures
Procedurepp. 75-88
RIS installation, removal, and ion source exchange
Procedurepp. 89-113
Tools, ion source cleaning, filament and vacuum maintenance
p. 65table, 13 rows
p. 24table, 9 rows
p. 43table, 2 rows
p. 67table, 4 rows
p. 90table, 10 rows
p. 10table, 21 rows
p. 11table, 8 rows
p. 40table, 7 rows
p. 61table, 2 rows
p. 62table, 8 rows
Table 7 Typical CI application settings
Parameter
Methane Positive
Methane Negative
Isobutane Positive
Isobutane Negative
Ammonia Positive
Ammonia Negative
EI
Ion polarity
Positive
Negative
Positive
Negative
Positive
Negative
N/A
Emission
150 μA
50 μA
150 μA
50 μA
150 μA
50 μA
35 μA
Electron energy
150 eV
150 eV
150 eV
150 eV
150 eV
150 eV
70 eV
Filament
1
1
1
1
1
1
1 or 2
Repeller
3 V
3 V
3 V
3 V
3 V
3 V
30 V
Ion focus
130 V
130 V
130 V
130 V
130 V
130 V
90 V
Entrance lens offset
20 V
20 V
20 V
20 V
20 V
20 V
25 V
Shutoff valve
Open
Open
Open
Open
Open
Open
Closed
Suggested flow
20%
40%
20%
40%
20%
40%
N/A
Source temp
250 °C
150 °C
250 °C
150 °C
250 °C
150 °C
230 °C
Quad temp
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
150 °C
Interface temp
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
280 °C
Autotune
Yes
Yes
No
Yes
No
Yes
Yes
N/A Not available
Table 2
Feature
7200 Series
High-vacuum pump 1
Split flow turbo
High-vacuum pump 2
Split flow turbo
High-vacuum pump 3
Turbo
Optimal carrier gas flow, mL/min
1.0 to 1.5
Reagent gas flow, mL/min
1 to 2
Collision cell gas flow
1.5
Maximum recommended gas flow, mL/min*
2.0
Maximum gas flow, mL/min†
2.4
Maximum column id
0.32 mm (30 m length)
* Total gas flow into the MS = column flow + collision cell gas flow + reagent gas flow (if applicable) + Agilent Quick Swap flow (if applicable)
† Expect degradation of spectral performance and sensitivity.
Table 4 Recommended temperature settings
Zone
EI operation
MS Source
230 °C
MS Quad
150 °C
Table 8 Default Tune Control Limits, used by CI autotune only
Methane
Methane
Ammonia
Ammonia
Ion polarity
Positive
Negative
Positive
Negative
Abundance target
1x106
1x106
N/A
1x106
Peakwidth target
0.7
0.7
N/A
0.7
Maximum repeller
4
4
N/A
4
Reagent gas spans Methane (Positive, Negative) and Ammonia (Positive, Negative); the first column has no header in the printed table.
Table 3 Influence of carrier and collision cell gas flows on vacuum
Column Flow (mL/min)
CC Gas On N2 = 1.5 mL/min Rough Vac
CC Gas On N2 = 1.5 mL/min Quad Vac
CC Gas On N2 = 1.5 mL/min TOF Vac
CC Gas Off Rough Vac
CC Gas Off Quad Vac
CC Gas Off TOF Vac
0.5
1.16 * 10-1
3.41 * 10-5
4.20* 10-7
8.29 * 10-2
3.17 * 10-7
2.15 * 10-7
0.7
1.18 * 10-1
3.41 * 10-5
4.20 * 10-7
8.67 * 10-2
3.45 * 10-7
2.15 * 10-7
1
1.22 * 10-1
3.43 * 10-5
4.22 * 10-7
9.17 * 10-2
3.79 * 10-7
2.16 * 10-7
1.2
1.24 * 10-1
3.43 * 10-5
4.22 * 10-7
9.49 * 10-2
4.06 * 10-7
2.16 * 10-7
2
1.32 * 10-1
3.44 * 10-5
4.22 * 10-7
1.07 * 10-1
5.17 * 10-7
2.17 * 10-7
3
1.42 * 10-1
3.44 * 10-5
4.25 * 10-7
1.19 * 10-1
6.51 * 10-7
2.20 * 10-7
4
1.50 * 10-1
3.46 * 10-5
4.27 * 10-7
1.29 * 10-1
7.84 * 10-7
2.21 * 10-7
These pressures are approximate and will vary from instrument to instrument by as much as 30%.
Table 5 Temperatures for CI operation
Ion source
Analyzer
GC/MS interface
PCI
300 °C
150 °C
280 °C
NCI
150 °C
150 °C
280 °C
Table 6 Typical analyzer vacuum with reagent gas flow
MFC (%)
Collision cell gas flow on N2 = 1.5 mL/min — Rough Pump (mTorr)
Collision cell gas flow on N2 = 1.5 mL/min — Quadrupole (Torr)
Collision cell gas flow on N2 = 1.5 mL/min — Flight Tube (Torr)
Collision cell gas flow off N2 = 0 mL/min — Rough Pump (mTorr)
Collision cell gas flow off N2 = 0 mL/min — Quadrupole (Torr)
Collision cell gas flow off N2 = 0 mL/min — Flight Tube (Torr)
0
1.36e+02
3.62e-05
3.35e-07
9.13e+01
5.98e-07
1.64e-07
10
1.36+02
3.62e-05
3.37e-07
1.14e+01
1.27e-06
1.65e-07
15
1.43+02
3.66e-05
3.37e-07
1.23e+01
1.62e-06
1.67e-07
20
1.5+02
3.71e-05
3.39e-07
1.31e+01
1.96e-06
1.67e-07
25
1.57+02
3.73e-05
3.41e-07
1.39e+01
2.32e-06
1.70e-07
30
1.63+02
3.77e-05
3.41e-07
1.46e+01
2.64e-06
1.71e-07
35
1.69+02
3.81e-05
3.41e-07
1.52e+01
3.00e-06
1.71e-07
40
1.74+02
3.83e-05
3.43e-07
1.58e+01
3.34e-06
1.72e-07
MFC is calibrated for methane and the vacuum gauge is calibrated for nitrogen, so these measurements are not accurate, but are intended as a guide to typical observed readings. Measurements will vary by as much as 30% from one MS to the next.
Engineering Facts
13 facts · every claim linked to a page
Safety
Hydrogen carrier gas must never be used with this system, so the gas supply and safety plan must be built around an alternative carrier.
Carrier gas selection and lab safety design must exclude hydrogen before the instrument is specified or installed.
Source
“The use of hydrogen gas is specifically prohibited with this product.”
Installation
The column installation procedure is instrument-specific, so technicians must not reuse procedures from other MS systems.
Using a wrong installation procedure can damage the column or the MS during setup.
Source
“The column installation procedure for the 7200 Q-TOF GC/MS is different from that for other MSs.”
Installation
Carrier gas purity must be at least 99.9995%, which drives the choice of gas supply and traps.
Gas purity directly affects column bleed, ion source contamination, and MS sensitivity.
Source
“Always use carrier gas that is at least 99.9995% pure.”
Maintenance
Columns must be conditioned before connection to the MS interface to limit stationary-phase deposits in the ion source.
Skipping conditioning increases ion source cleaning frequency and reduces sensitivity.
Source
“Conditioning a column before it is connected to the GC/MS interface is essential.”
Installation
New ferrules can loosen after thermal cycling, so tightness must be rechecked after a few heating cycles.
Leak checks after the first heating cycles prevent vacuum loss and air leaks.
Source
“Because of thermal expansion, new ferrules may loosen after heating and cooling a few times.”
Operation
The GC/MS interface and oven must not be heated until carrier gas is on and the column is purged of air.
Heating without gas flow damages the column and contaminates the MS.
Source
“Ensure that the carrier gas is turned on and the column has been purged of air before heating the GC/MS interface or the GC oven.”
Operation
The analyzer cover should only be opened for pumpdown or sideplate thumbscrew access, not for general work.
Unnecessary opening risks vacuum loss and contamination of the analyzer.
Source
“You should not open the analyzer cover for any other reason.”
Operation
Reagent gas flow rates differ between PCI and NCI startup, so the flow controller must be set per mode.
Wrong reagent gas flow prevents stable CI tuning and sensitivity.
Source
“PCI mode set reagent gas flow to 20 (1 mL/min)”
Performance
EI performance must be verified before switching to CI, so a baseline EI check belongs in the workflow.
This confirms the system is healthy before introducing reagent gas and changing source conditions.
Source
“Always verify MS performance in EI before switching to CI operation.”
Maintenance
The RIS allows EI/CI source switching in under 30 minutes while keeping the instrument under vacuum.
This short cycle time affects method scheduling and uptime planning.
Source
“Switching back and forth between CI and EI ion sources takes less than 30 minutes with the new removable ion source (RIS).”
Selection
The system requires a methane/isobutane gas purifier for CI, which must be included in the installation scope.
The purifier removes oxygen, water, hydrocarbons, and sulfur compounds that would degrade CI performance.
Source
“For the CI system, a methane/isobutane gas purifier is provided and is required.”
Safety
The foreline pump exhaust must be vented outside or to a fume hood when toxic solvents or chemicals are analyzed.
Lab exhaust routing must be planned before installation to protect operators.
Source
“attach a hose (11-mm id) to the oil mist filter to take the foreline pump exhaust outside or to a fume (exhaust) hood.”
Installation
The instrument weighs 152 kg and the foreline pump adds 21.5 kg, so floor loading and rigging must be planned.
Bench strength and moving equipment must be sized for the total installed weight.
Source
“The weight of the instrument mainframe is 152 kg.”