Induztr
Engineering Digest Document Viewer Related Models

InfinityLab Assist Control Software Release Note

Contents

  1. Overview pp. 1-2
    Release note purpose and introduction to the software update
  2. Overview pp. 3-4
    New features and changes in this software release
  3. Spec p. 5
    Supported LC modules and firmware compatibility requirements
  4. Procedure p. 6
    Installation instructions and prerequisites for the software
  5. Overview pp. 8-9
    Detailed change history for versions 2.0.14 and 1.1.0
  6. Spec pp. 10-12
    Required InfinityLab Assist Control Software per LC module
  7. Warning p. 13
    Important notes and cautions for users
  8. p. 14
    Agilent contact and support information
  9. p. 5 table, 2 rows
  10. p. 11 table, 11 rows
  11. p. 12 table, 17 rows
  12. p. 13 table, 4 rows

Supported Operating Systems

SpecificationDescription Details
Operating system name, versionWindows 10 or 11, Enterprise of Professional, 64-bit
Web browserChromium-based browser (Chrome, Edge, Firefox, etc.) with a version higher than 132 Safari-based browser with a version higher than 17.6

Agilent LC – Pumps

Product NumberDescriptionmin. required version
G5654AAgilent 1260 Infinity II Bio-Inert Pump1.0
G7104AAgilent 1290 Infinity II Flexible Pump1.0
G7104CAgilent 1260 Infinity II Flexible Pump1.0
G7110BAgilent 1260 Infinity II Isocratic Pump1.0
G7111AAgilent 1260 Infinity II Quaternary Pump VL1.0
G7111BAgilent 1260 Infinity II Quaternary Pump1.0
G7112BAgilent 1260 Infinity II Binary Pump1.0
G7120AAgilent 1290 Infinity II High-Speed Pump1.0
G7131AAgilent 1290 Infinity II Bio Flexible Pump1.0
G7131CAgilent 1260 Infinity II Bio Flexible Pump1.0
G7132AAgilent 1290 Infinity II Bio High-Speed Pump1.0

Agilent LC – Detectors

Product NumberDescriptionmin. required version
G7114AAgilent 1260 Infinity II Variable Wavelength Detector1.0
G7114BAgilent 1290 Infinity II Variable Wavelength Detector1.0
G7115AAgilent 1260 Infinity II Diode Array Detector WR1.0
G7117AAgilent 1290 Infinity II Diode Array Detector FS1.0
G7117BAgilent 1290 Infinity II Diode Array Detector1.0
G7117CAgilent 1260 Infinity II Diode Array Detector HS1.0
G7121AAgilent 1260 Infinity II Fluorescence Detector1.0
G7121BAgilent 1260 Infinity II Fluorescence Detector Spectra1.0
G7162AAgilent 1260 Infinity II Refractive Index Detector1.0
G7162BAgilent 1290 Infinity II Refractive Index Detector1.0
G7165AAgilent 1260 Infinity II Multiple Wavelength Detector1.0
G6160BAgilent InfinityLab Pro iQ Mass Detector2.0
G6170AAgilent InfinityLab Pro iQ Plus Mass Detector2.0
G7123BAgilent 1290 Infinity III Fluorescence Detector2.2 (New)
G6495D6495D Triple Quadrupole LC/MS2.2 (New)
G6475A6475 Triple Quadrupole LC/MS2.2 (New)
G6465BUltivo Triple Quadrupole LC/MS2.2 (New)

Agilent LC – Clusters

ProductDescriptionDescription min. required version
Agilent Pump Valve Cluster DriverA combination of one of the following pumps with up to two G1170A and valves 5067-4147 or 5067-4159 Supported pumps: G7111x, G7112x, G5654x, G7104x, G7120x, G7131x, G7132x1.0
Agilent Valve-Thermostat Cluster DriverA combination of G7116B, G1170A for combined valve control plus a combination of G7116B and G7130A for combined temperature control. Supports up to 32 columns.2.2 (New)
Agilent 2D-LC System DriverCombination of one 2D-capable pump G7120A with up to four G1170A valve drives into a 2D-LC System Driver.2.1
Agilent 1200 Infinity Series High Dynamic Range DAD SolutionCombines two G7117A/B for high dynamic range DAD applications2.2 (New)

NOTE Modules which are part of Agilent Pump Valve Cluster and Agilent 2D-LC System are displayed with a cluster icon in status. Task execution for these clusters is not available in this version and will be added in a future release.

Engineering Facts

13 facts · every claim linked to a page
Selection

VTC and HDR cluster configurations require Agilent InfinityLab LC and CE Drivers version 3.11 or later, so driver version must be verified before specifying these clusters.

Prevents selecting cluster hardware that cannot be fully controlled without the correct driver version.

Source
“Full support for Valve-Thermostat Cluster (VTC) and High Dynamic Range (HDR) detector cluster configurations requires Agilent InfinityLab LC and CE Drivers version 3.11 or later.”
Selection

The software requires a 64-bit Windows 10 or 11 Professional/Enterprise PC and a sufficiently recent browser, which must be confirmed before deployment.

Defines the minimum PC platform needed to run the browser-based user interface.

Source
“Windows 10 or 11, Enterprise of Professional, 64-bit”
Selection

The browser used to access the UI must be Chromium-based above version 132 or Safari-based above version 17.6.

An outdated browser will block access to the control software interface.

Source
“Chromium-based browser (Chrome, Edge, Firefox, etc.) with a version higher than 132”
Compliance

The minimum supported Agilent LC and CE Drivers version is 3.7 SR1, so older driver installations are not supported.

Ensures the CDS/driver stack meets the minimum compatibility baseline before installation.

Source
“3.7 SR1. Minimum supported version.”
Selection

OpenLab CDS must be at version 2.6 or higher (including 2.8 FP1, FP2 and Update 9 Client/Server) for browser access support.

Confirms which CDS versions provide the required browser access for the Assist software.

Source
“2.8 FP1, FP2 and Update 9 Client/Server (provides browser access)”
Selection

For Empower integration, the minimum is Empower 3 FR4 with Agilent Driver for Waters Empower 4.1 (Agilent LC Drivers 3.8).

Defines the minimum CDS and driver combination for Empower-based laboratories.

Source
“Min. Empower 3 FR4”
Selection

For Chromeleon integration, the Agilent Driver for Thermo Fisher Chromeleon 3.2 requires Agilent LC Drivers 3.11.

Prevents mismatched driver versions when connecting to a Chromeleon CDS.

Source
“Agilent Driver for Thermo Fisher Chromeleon 3.2 (Agilent LC Drivers 3.11)”
Installation

Installation and setup must follow the official Agilent InfinityLab LC Series Assist User Manual procedures.

Ensures correct installation, configuration and initial operation per Agilent specifications.

Source
“users are advised to follow the procedures outlined in the official documentation”
Maintenance

The Software Status Bulletin lists known limitations, incompatibilities and fixes/workarounds and should be consulted for this and previous versions.

Known limitations may affect commissioning and require workarounds during setup.

Source
“The Software Status Bulletin lists known limitations, incompatibility and information about available fixes or workarounds for this and previous versions.”
Maintenance

Agilent recommends using the latest driver, firmware and software updates, available from the InfinityLab Assist Updates site.

Keeping firmware and software current avoids known defects and ensures compatibility.

Source
“Agilent continuously improves its drivers, firmware and software and recommends using latest updates.”
Operation

Modules in the Pump Valve Cluster and 2D-LC System are shown with a cluster icon, but task execution for these clusters is not available in this version.

Prevents planning automated task workflows around clusters that cannot yet execute tasks.

Source
“Task execution for these clusters is not available in this version and will be added in a future release.”
Selection

The Valve-Thermostat Cluster supports up to 32 columns, which must be considered when sizing a multi-column switching system.

Defines the maximum column capacity of the VTC configuration for method planning.

Source
“Supports up to 32 columns.”
Selection

The 1290 Infinity III Hybrid Multisampler (G7137B) requires Assist Control Software version 2.0 or higher.

Confirms the minimum software version needed to support this multisampler.

Source
“Agilent 1290 Infinity III Hybrid Multisampler”