Introduction and general information about the manual
Procedurepp. 17-24
Installation instructions and setup procedures
Procedurepp. 25-26
Operating instructions and procedures
Procedurepp. 27-28
Troubleshooting guide and fault diagnosis
Procedurepp. 29-32
Maintenance and service procedures
Orderingpp. 33-37
Spare parts list and replacement information
p. 42
Contact information for service and support
p. 28table, 16 rows
p. 18table, 28 rows
p. 19table, 12 rows
p. 36table, 14 rows
p. 33table, 18 rows
p. 37table, 4 rows
p. 12table, 8 rows
p. 14table, 10 rows
Table 3-1 Troubleshooting Guide
Fault
Probable Cause
Corrective Action
Poor system pressure
Leaks in system, virtual or real
Locate and repair
Poor ultimate pressure
Contaminated pump fluid
Examine and clean pump; replace fluid
Low heat input
Check voltage. Check for continuity, burned-out element, poor thermal contact
Excessive or too cold cooling
Check temperature and correct.
High forepressure
Check for leak in foreline, poor mechanical pump performance, breakdown of mechanical pump fluid
Low speed (Prolonged cycle after checking gas load conditions)
Low heat input
Check heaters
Low oil level
Add oil
Improperly located jets
Check and repair or replace
Inlet pressure surges
Incorrect heater voltage
Check and correct voltage
Fluid outgassing
Condition fluid by operating the pump for a 24 hours
Leak in system ahead of pump inlet
Check and correct
High chamber contamination of the pump fluid
Forepressure too high
Check for leak in foreline, poor mechanical pump performance, breakdown of pump fluid, and incorrect valve operation
Prolonged operation in overload range: 10-3 Torr
Adhere to operating procedures
Improper system operation and air release procedures
Adhere to operating procedures
Pump will not start
Safety circuits open
Inspect for electrical short or overtemperature condition
Main power off
Check main power source
Table 1-1 Operating Specifications
Specification
Units
AX-65 with cold cap
AX-65 with baffle
Operating Range
Torr
3x10-3 to <5x10-8
8.0x10-3 to <5x10-8
mbar
3.9x10-3 to <6.5x10-8
10.4x10-3 to <6.5x10-3
Pumping Speed, Maximum
l/s, Air
65
30
l/s, Helium
90
65
Throughput (Air), Max. Optimum Range
Torr-l/s
0.19
0.30
mbar-l/s
0.25
0.40
Overload Range (@ 1x10-2 Torr)
Torr-l/s
mbar-l/s
Forepressure, Max. No Load
Torr
0.75
0.60
mbar
1.00
0.78
Full Load
Torr
mbar
Compression Ratio Air
4x107
He
2x106
Backstreaming Rate at Pump Inlet
mg/cm2/min
<2x10-4
<2x10-5
Electrical Power
Watts
200 W nominal, 260 W maximum
250 W nominal, 320 W maximum
Warmup Time
minutes
7
Cooldown Time
minutes
10
Fluid Charge
cc
30
Recommended backing pump capacity
cfm (m3/hr)
1.5 (2.5)
Cooling Fan
cfm
30
Inlet Flange
N/A
NW50 or ISO-63-K
Body
N/A
Stainless steel
Jet Assembly
N/A
Three-stage, self-aligning, aluminum, w/aluminum cold cap
Net Weight
lbs (kg)
8 (3.63)
Environmental
Ambient temperature: 5 °C (41 °F) to 45 °C (113 °F); Maximum relative humidity: 80% for temperatures up to 31 °C (88 °F) decreasing linearly to 50% relative humidity at 40 °C (104 °F)
Installation
Indoor use, Installation Category II, Pollution Degree 2
Altitude
m (ft)
2000 m (6562’)
Table 1-2 AX-65 Electrical Specifications
Model
Voltage VAC, 50/60 Hz 1 phase, ±10%
Part Number Inlet Flange Type KF50
Part Number Inlet Flange Type ISO63
AX-65 with standard cold cap
115V, 250W
L9670301
L9707301
AX-65 with internal baffle
115V, 250W
L9670311
L9707311
AX-65 with standard cold cap
220V, 250W
L9670302
L9707302
AX-65 with internal baffle
220V, 250W
L9670312
L9707312
AX-65 with standard cold cap
115V, 200W
L9670303
L9707303
AX-65 with internal baffle
115V, 200W
L9670313
L9707313
AX-65 with standard cold cap
220V, 200W
L9670304
L9707304
AX-65 with internal baffle
220V, 200W
L9670314
L9707314
AX-65 with standard cold cap
90V, 250W
L9670305
L9707305
AX-65 with internal baffle
90V, 250W
L9670315
L9707315
AX-65 with standard cold cap
165V, 250W
L9670306
L9707306
AX-65 with internal baffle
165V, 250W
L9670316
L9707316
Table B-3 Ordering Information
Description
Contents
Agilent Part No.
NEOVAC SY
1 liter/1,000 cc
K6948301
NEOVAC SY
1 gallon/3.8 liters
K6948305
NEOVAC SY
5 gallons/18.9 liters
K6948315
DC-702 Silicone
500 cc
695472005
DC-702 Silicone
1 gallon/3.8 liters
695472008
DC-702 Silicone
5 gallons/18.9 liters
695472015
DC-704 Silicone
500 cc
695474005
DC-704 Silicone
1 gallon/3.8 liters
695474008
DC-704 Silicone
5 gallons/18.9 liters
695474015
DC-705 Silicone
500 cc
695475005
DC-705 Silicone
1 gallon/3.8 liters
695475008
Santovac 5
40 cc
695405001
Santovac 5
65 cc
695405002
Santovac 5
500 cc
695405005
Table B-1 AX-65 Diffusion Pump Replacement Parts List
Item
Part Number
Description
1
L9674301
Jet Assembly
2
L9675301
Cold Cap
3
L9706001
Inlet Baffle
4
L9995301
Jet Alignment Kit
5
L9994301
Heater Harness, 90V, 250W
L9994303
Heater Harness, 115V, 250W
L9994302
Heater Harness, 165V, 250W
L9994304
Heater Harness, 220V, 250W
L9994307
Heater Harness, 115V, 200W
L9994308
Heater Harness, 220V, 200W
6
Not Available
7
Not Available
8
661300138
Fan, Universal Voltage
9
L9676301
Foreline Baffle
10
L9680301
Centering Ring Assembly, NW50, with Viton O-ring
11
KC16AV
Centering Ring Assembly, NW16, with Viton O-ring
12
L9701002
Bottom Insulation (2 required with heater harness)
695405001
Diffusion Pump Fluid, Santovac 5, 40 cc
RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:
NORTH AMERICA:
PACIFIC RIM:
EUROPE:
Fax: 1 781 860 9252
please visit our website for individual office information http://www.DJLOHQW.com
Fax: 00 39 011 9979 330
Toll Free: 800 882 7426, Option 3
Fax Free: 00 800 345 345 00
vpl-ra@agilent.com
Toll Free: 00 800 234 234 00
vpt-customercare@agilent.com
Таблица контактов по регионам; в тексте также присутствует заголовок 'Request for Return Health and Safety Certification'
Table 1 General Hazards
Hazard
Suggested Corrective Action
Loss of utility: water and/or electricity
Provide sufficient backup water and power supply as necessary to effect a safe shutdown under worst case conditions
Overpressure in foreline
Provide an interlock to ensure that the power supply to the pump heater cannot be activated if the foreline pump is not running and/or the pressure in foreline is above 0.5 Torr (0.67 mbar)
Overtemperature
Fit temperature sensors and pump fluid level sensors with a feedback to an interlock on the heater power supply
Loss of electrical ground integrity
Incorporate ground fault interrupt circuit into heater power supply
Positive pressure in pumping system
Integrate pressure relief valve in vacuum system
High voltage
Prevent personnel contact with high voltages; design and attach warnings
Toxicity and Corrosivity
Vent toxic and/or corrosive gases to a safe location; ensure adequate dilution or scrubbing to safe levels; take all actions required to meet air quality standards
Explosion
Integrate pressure relief valves
Do not use hydrocarbon-based pumping fluids
Table 2 Explosive Conditions
Prohibited Action
Explosion-Causing Condition
Do not run pump without cooling water
Overtemperature
Do not run pump with low level of pump fluid
Overtemperature
Do not run pump without proper backing or holding pump
Overpressure
Do not run pump when not evacuated below 0.5 Torr (0.66 mbar)
Overpressure
Do not admit air to, or rough through, a pump with hot boiler
Overpressure plus strong oxidizer
Do not open drain or fill plug while pump is under vacuum, especially when it is hot
Overpressure plus strong oxidizer
Do not contaminate pump with explosive vapors
Lower explosive threshold of gas mixtures
Do not remove, defeat, or override safety counter-measures such as pressure and thermal switches and valve sequencer interlocks
Overtemperature, overpressure, more combustible mixtures
Do not machine or weld any part of the pump without removing all fluid or solvent residue in pump
Source of ignition
Do not use unsuitable pumping fluid
Lower explosive threshold of gas mixture
Engineering Facts
14 facts · every claim linked to a page
Selection
Choose polyphenyl ether fluid (e.g., Santovac 5) for systems with internal electronics and pair it with a 250 W heater; use lower-boiling fluids like DC-704 only with 200 W heaters.
Fluid choice is tied to heater wattage and system electronics; wrong pairing risks overheating or poor performance.
Source
“When a diffusion pump is integrated into a system that has internal electronics, use a polyphenyl ether diffusion pump fluid such as Santovac 5.”
Installation
Mount the pump body vertical and plumb with the inlet mating flange horizontal within ±1° so the boiler plate stays level and avoids overheating.
Uneven fluid level from tilt can overheat the boiler plate and damage the pump.
Source
“Install the diffusion pump must be installed with the body vertical and plumb. Ensure that the pump inlet mating flange on the system is horizontal within ± 1°.”
Installation
Provide at least 1 inch clearance between the cooling fan and any closet surface, and mount the fan 90° from the foreline.
Insufficient airflow causes overheating; fan orientation affects cooling of the pump body.
Source
“Achieve adequate air flow by ensuring there is a minimum clearance of 1" between the fan and the closet surface.”
Installation
Install supplementary fusing no larger than 240 V, 15 A fast-acting for overcurrent protection when the pump is integrated into a system.
Correct fusing protects the heater circuit and prevents fire or equipment damage.
Source
“For overcurrent protection, the pump must have supplementary fusing when installed in a system. The fuse cannot be larger than 240 V, 15 A, fast acting.”
Operation
Never turn on the heater without fluid in the pump, and never run the heater unless the cooling fan is running.
Dry running or loss of cooling destroys heaters and overheats the pump fluid.
Source
“Do not turn on the heater without fluid in the pump. This may ruin the heaters and damage the pump.”
Operation
Keep forepressure below approximately 0.5 Torr during operation; higher forepressure causes poor pumping and excessive backstreaming.
Exceeding tolerable forepressure prevents proper operation and contaminates the chamber.
Source
“Do not operate the pump with forepressure (outlet pressure) above approximately 0.5 Torr”
Operation
After shutdown, let the fan run about 10 minutes (or 30 minutes if fan and heater are wired together) and confirm the body is at most 130 °F before venting.
Venting a hot boiler exposes hot fluid to oxidizer and can cause explosion or fluid breakdown.
Source
“The temperature of the pump body just above the boilerplate should be a maximum of 130 °F before the pump is vented.”
Maintenance
While running, keep fluid level within the FULL HOT zone; if low, cool down and add fluid to the FULL COLD zone (30 cm³ charge for an empty pump).
Low fluid level causes overheating and loss of pumping performance.
Source
“When the pump is running, the fluid level should remain within the zone indicated as FULL HOT.”
Maintenance
Do not use solvents on O-rings; wipe them with a clean lint-free cloth or a small amount of diffusion pump oil instead.
Solvent attack on O-rings causes leaks and loss of vacuum integrity.
Source
“Solvents degrade O-ring materials reducing their ability to hold vacuum. Do not use solvents on O-rings.”
Safety
Provide an interlock so heater power cannot be activated unless the foreline pump is running and foreline pressure is below 0.5 Torr.
Prevents overpressure operation that can cause explosion and equipment damage.
Source
“Provide an interlock to ensure that the power supply to the pump heater cannot be activated if the foreline pump is not running and/or the pressure in foreline is above 0.5 Torr (0.67 mbar)”
Safety
Never pressurize any part of the vacuum system for testing or any other purpose; vacuum components are not designed for positive pressure.
Pressurization can burst the pump body and expel shrapnel at lethal velocity.
Source
“Never pressurize any part of a vacuum system for test or any other purpose.”
Safety
Boiler surfaces reach 530 °F (275 °C); ensure surfaces have cooled to near room temperature before touching.
Prevents serious burns during maintenance and operation.
Source
“Boiler temperatures reach 530 °F (275 °C) which can cause serious burns.”
Performance
The gas load at the inlet must not exceed the maximum throughput of 0.30 Torr-liters/sec, or the diffusion pump will not function.
Exceeding throughput causes loss of pumping action and relies only on the backing pump.
Source
“The gas load must be reduced to the maximum throughput specification (0.30 Torr-liters/sec) or less.”
Selection
For leak detector applications, Agilent recommends using Santovac 5 fluid only.
Ensures compatibility and performance in sensitive leak detection systems.
Source
“For leak detectors, Agilent recommends the use of Santovac 5 only.”