FED (E2) 01 2025
Technical Reference Summary
Covered Models & Document Scope
This operating manual (FED / FED-UL (E2), issue 01/2025, Art. No. 7001-0070) covers the BINDER drying and heating ovens with forced convection and enhanced timer functions: FED 400 (E2), model version FED400-400V, Art. No. 9010-0216 / 9110-0216, and FED 400-UL (E2), model version FED400UL-208V, Art. No. 9010-0217 / 9110-0217. Options addressed include temperature safety device class 3.1, disconnectable audible over-temperature alarm, APT-COM™ 4 Multi Management Software, HEPA fresh air filter, analog temperature output, interior lighting, and connection to a suction plant.
Section-by-Section Technical Breakdown
The FED / FED-UL chambers are electrically heated drying and heating ovens with fan-assisted forced-air circulation, an electronic PID controller with digital display (accuracy one degree), and the APT.line™ preheating chamber system. The inner chamber, preheating chamber and door interiors are stainless steel V2A (1.4301 / AISI 304); the housing is powder-coated RAL 7035 or RAL 9003. A serial RS 422 interface is provided for computer communication, e.g. via APT-COM™ 4.
- Temperature range: 5 °C / 9 °F above room temperature up to 300 °C / 572 °F.
- Electronic PID controller with digital display, temperature indicated to one degree.
- Fan-assisted forced-air circulation with APT.line™ preheating chamber system.
- Stainless steel V2A (German material no. 1.4301, US equivalent AISI 304) inner chamber, preheating chamber and door interior.
- Housing powder-coated RAL 7035 or RAL 9003; all corners and edges fully coated.
- Serial interface RS 422 for computer communication (e.g. APT-COM™ 4).
- Above 150 °C, oxygen in the air may cause harmless yellowish-brown or blue discoloration of metallic surfaces.
- Equipment overview items: (1) display, (2) set-point value key, (3) selector keys, (4) time management key, (5) switch ON/OFF, (6) lever for ventilation slide, (7) safety device, (8) door handle, (9) interior lighting switch or buzzer switch, (10) main power switch.
The chambers are designed for indoor use inside a building. Installation in unventilated recesses is prohibited; sufficient ventilation must disperse heat, and prescribed minimum wall distances must be observed. The chamber must not be installed in damp areas, puddles, hazardous locations or potentially explosive areas.
- Wall distances: rear 100 mm / 3.9 in, sides 160 mm / 6.29 in.
- Do NOT install the chamber in unventilated recesses; ensure sufficient ventilation for heat dispersal.
- Indoor use only; do not install in damp areas or puddles; set up splash-proof.
- Do not install or operate in hazardous locations or potentially explosive areas.
- Keep combustible dust or air-solvent mixtures away from the chamber.
Electrical connection data are model-specific. FED 400 uses a 5-pole CEE plug on 3N~; FED 400-UL uses a NEMA L21-20P plug at 208 V / 60 Hz, 3N~. The type plate example for FED 400 states 3.40 kW nominal power, 7.4 A nominal current, 400 V / 50 Hz or 400 V / 60 Hz, 3N~, nominal voltage tolerance ±10%.
- FED 400: CEE plug 5-pole, current type 3N~.
- FED 400-UL: NEMA L21-20P, 208 V at 60 Hz, 3N~.
- FED 400 type plate example: nominal power 3.40 kW, nominal current 7.4 A.
- Nominal voltage 400 V / 50 Hz or 400 V / 60 Hz, 3N~, tolerance ±10% at the indicated power frequency.
- Observe local regulations when connecting.
Start-up is performed by pressing the switch until the display lights up. If the temperature safety device has tripped (chap. 6.3), the chamber is inactive (no heating) and the display shows an alarm value. The red heating control light in the display indicates heating operation.
- Press until the display lights up to turn on the chamber.
- If the safety device has responded, the chamber is inactive (no heating) and the display shows the alarm value.
- Red heating control light in the display indicates active heating.
Temperature and ventilation set-points are displayed and entered either without a ramp function or with a selected temperature ramp. Time functions provide Continuous operation and Timer operation, selectable and configurable via the time management key and user level settings.
- Display/entry of temperature and ventilation set-points without ramp function.
- Display/entry of temperature and ventilation set-points with selected temperature ramp.
- Time functions: Continuous operation and Timer operation; switching between them.
- Setting the timer values.
- User level settings include temperature unit change (°C / °F), entering a temperature ramp, chamber addressing, selecting the timer function, and setting the interface mode and printer interval.
The chambers are equipped with a temperature safety device class 2 according to DIN 12880:2007, with visual and audible (buzzer) signals indicating temperature exceeding. A temperature safety device class 3.1 (DIN 12880) and a disconnectable audible over-temperature alarm are available as options.
- Standard: temperature safety device class 2 acc. to DIN 12880:2007.
- Visual and audible (buzzer) signals indicate temperature exceeding.
- Option: temperature safety device class 3.1 acc. to DIN 12880.
- Option: disconnectable audible over-temperature alarm.
- Type plate example: temp. safety device DIN 12880, class 2.0.
Documented options include APT-COM™ 4 Multi Management Software, HEPA fresh air filter, analog output for temperature, temperature safety device class 3.1, disconnectable audible over-temperature alarm, interior lighting, and connection to a suction plant.
- APT-COM™ 4 Multi Management Software (option).
- HEPA fresh air filter (option).
- Analog output for temperature (option).
- Temperature safety device class 3.1 (DIN 12880) (option).
- Disconnectable audible over-temperature alarm (option).
- Interior lighting (option) with switch (9).
- Connection to a suction plant (optional).
The manual provides cleaning and decontamination instructions. The chamber must not become wet during operation, cleaning or maintenance; water penetration creates a deadly electric shock hazard. Cleaning while the chamber is turned on is listed as foreseeable misuse.
- The chamber must NOT become wet during operation, cleaning, or maintenance.
- Do NOT spray with water.
- Cleaning activity while the chamber is turned on is prohibited.
- Protect the interior from contamination by toxic, infectious or radioactive substances.
- Take suitable protective measures when introducing and removing toxic, infectious or radioactive material.
Regular maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. Before maintenance, turn off the chamber at the main power switch and disconnect the power plug; the rear panel must not be removed. An annual repeat electrical inspection and tests before initial start-up and before resuming operation after maintenance or repair are required, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702.
- Maintenance at least once a year by licensed electricians or experts authorized by BINDER.
- Before maintenance: turn off at main power switch and disconnect power plug.
- Do NOT remove the rear panel of the chamber.
- The chamber must NOT become wet during operation or maintenance work.
- Annual repeat electrical inspection required; also test before initial start-up and before resuming operation after maintenance or repair.
- Recommended electrical tests: EN 50678/VDE 0701 and EN 50699/VDE 0702.
- Replace the door gasket only when cold; otherwise the door gasket may become damaged.
- Warranty becomes void if maintenance work is conducted by non-authorized personnel.
- After maintenance, the chamber must be tested prior to resuming operation.
The technical description covers factory calibration and adjustment, definition of usable volume, overcurrent protection, FED / FED-UL technical data, equipment and options extract, and accessories and spare parts extract. The type plate example for FED 400 lists nominal temperature 300 °C / 572 °F, IP protection 20, nominal power 3.40 kW, nominal current 7.4 A, and 400 V / 50 Hz or 60 Hz, 3N~.
- Nominal temperature 300 °C / 572 °F.
- IP protection 20 acc. to EN 60529.
- Nominal power 3.40 kW; nominal current 7.4 A (FED 400 type plate example).
- Nominal voltage 400 V / 50 Hz or 400 V / 60 Hz, 3N~, ±10%.
- Temperature safety device DIN 12880, class 2.0.
- Factory calibration and adjustment, definition of usable volume, and overcurrent protection are described.
- Equipment and options extract and accessories/spare parts extract are provided.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Temperature range: 5 °C / 9 °F above room temperature up to 300 °C / 572 °F. p.15 ↗ | The chamber cannot control below 5 °C above ambient; applications requiring lower temperatures are not supported. | FED / FED-UL series |
| Wall distances: rear 100 mm / 3.9 in, sides 160 mm / 6.29 in. p.18 ↗ | Minimum clearances must be maintained for heat dispersal and to avoid overheating damage. | FED / FED-UL series |
| FED 400 electrical connection: CEE plug 5-pole, current type 3N~. p.19 ↗ | Site must provide a compatible 5-pole CEE receptacle for the FED 400 variant. | FED 400 (E2), FED400-400V |
| FED 400-UL electrical connection: NEMA L21-20P, 208 V at 60 Hz, 3N~. p.19 ↗ | UL variant requires a 208 V / 60 Hz three-phase supply with NEMA L21-20P receptacle. | FED 400-UL (E2), FED400UL-208V |
| FED 400 type plate example: nominal power 3.40 kW, nominal current 7.4 A, nominal voltage 400 V / 50 Hz or 400 V / 60 Hz, 3N~, tolerance ±10%. p.8 ↗ | Supply voltage must remain within ±10% of nominal at the indicated power frequency; size upstream protection and wiring for 7.4 A. | FED 400 (E2) |
| Nominal temperature 300 °C / 572 °F; IP protection 20 acc. to EN 60529; temperature safety device DIN 12880 class 2.0. p.8 ↗ | The chamber is rated to 300 °C and IP20 (no water protection); the standard safety device is class 2.0. | FED 400 (E2) type plate example |
| Temperature is indicated with an accuracy of one degree. p.15 ↗ | Display resolution is 1 °C; finer set-point resolution is not available from the display. | FED / FED-UL series |
| Inner chamber, preheating chamber and door interiors are stainless steel V2A (German material no. 1.4301, US equivalent AISI 304). p.15 ↗ | Chamber materials are 304 stainless steel; above 150 °C harmless discoloration may occur. | FED / FED-UL series |
| Above 150 °C, oxygen in the air may cause yellowish-brown or blue discoloration of metallic surfaces by natural oxidation; these colorations are harmless and do not impair function or quality. p.15 ↗ | Discoloration above 150 °C is not a defect and requires no corrective action. | FED / FED-UL series |
| Serial interface RS 422 is provided for computer communication, e.g. via APT-COM™ 4 Multi Management Software. p.15 ↗ | Data logging and remote monitoring require RS 422-compatible connection and APT-COM™ 4 software. | FED / FED-UL series |
| Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. p.39 ↗ | Plan annual service; non-authorized maintenance voids the warranty. | FED / FED-UL series |
| Annual repeat electrical inspection required, plus tests before initial start-up and before resuming operation after maintenance or repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.39 ↗ | Electrical safety testing is mandatory at commissioning, annually, and after service. | FED / FED-UL series |
| Before maintenance work, turn off the chamber at the main power switch and disconnect the power plug; do NOT remove the rear panel. p.39 ↗ | Live maintenance is prohibited; rear panel removal is not permitted by operating personnel. | FED / FED-UL series |
| Replace the door gasket only when cold. p.39 ↗ | Replacing the gasket while hot may damage it. | FED / FED-UL series |
| If the temperature safety device has responded, the chamber is inactive (no heating) and the display shows the alarm value. p.20 ↗ | No heating after a safety trip is expected behavior; investigate the over-temperature cause before resetting. | FED / FED-UL series |
| The chamber does not dispose of any measures of explosion protection; no combustible or explosive substance at working temperature may be introduced. p.11 ↗ | Solvent content must not be explosive or flammable and operating temperature must lie below the flash point or sublimation point of the charging material. | FED / FED-UL series |
| Loading material shall not contain corrosive ingredients, in particular acids and halides. p.11 ↗ | Corrosive charge materials can damage chamber components and are excluded from BINDER liability. | FED / FED-UL series |
| Persons with active implants (e.g. pacemakers, defibrillators) should keep a distance of 30 cm from field sources. p.15 ↗ | Maintain a 30 cm separation between implant wearers and electrical equipment such as motors and power cables. | FED / FED-UL series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.40 ↗ | No power supply. | Check connection to power supply. Check whether the chamber is turned on at the main power switch. | Chap. 10.3 |
| Chamber without function. p.40 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.1). | Chap. 10.3 |
| Chamber without function. p.40 ↗ | Chamber fuse has responded. | Contact BINDER service. | Chap. 10.3 |
| Chamber without function. p.40 ↗ | Controller defective. | Contact BINDER service. | Chap. 10.3 |
| Set-point temperature is not reached after specified time. p.40 ↗ | Chamber door not properly closed. | Completely close chamber door. | Chap. 10.3 |
| Set-point temperature is not reached after specified time. p.40 ↗ | Door gasket defective. | Replace door gasket. | Chap. 10.3 |
| Set-point temperature is not reached after specified time. p.40 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chap. 10.3 |
| Chamber heating permanently, set-point not held. p.40 ↗ | Controller defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber heating permanently, set-point not held. p.40 ↗ | Pt 100 sensor defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber heating permanently, set-point not held. p.40 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber heating permanently, set-point not held. p.40 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chap. 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is lit. p.40 ↗ | Heating element defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is lit. p.40 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.40 ↗ | Timer has run off. | Program the timer or change to time function Continuous operation (chap. 6.3). | Chap. 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.40 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chap. 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.40 ↗ | Controller defective. | Contact BINDER service. | Chap. 10.3 |