BINDER FED 400
BINDER ·Made in Germany · In Active Production

BINDER FED 400

BINDER FED 400 — Drying ovens with Temperature 5…300 °C.

The FED 400 is a forced-convection drying and heating oven with a 400 L stainless-steel interior (V2A 1.4301/AISI 304) and APT.line™ preheating chamber, rated +5 to 300 °C. Within the FED series it is the 400 L modification with 3.4 kW heating and 18 x/h air change, distinguished from smaller siblings by its 1,000 × 800 × 765 mm working volume and 90 kg total load capacity. It serves batch drying, heating and thermal testing where uniform temperature distribution (3.8 K) and short recovery after door opening are required.

Temperature
5…300 °C

BINDER FED Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FED 400 Current 5–3001876529102.5400
FED 115 10–300326051870.5116
FED 260 10–30097601994259
FED 56 10–300805651562.560
FED 720 12–300587025159741

Full specifications

Core Performance & Functional Specifications
Temperature 5...300 °C
Air change rate 18 x/h
Heating time to 150°C 29 min
Interior volume 400 L
Load per rack 35 kg (FED400-400V / FED400UL-208V)
Nominal power 3.4 kW
Permitted load 90 kg (FED400-400V / FED400UL-208V)
Temperature fluctuation 0.7 K
Temperature uniformity 3.8 K
Rear wall clearance 160 mm (FED400-400V / FED400UL-208V)
Side wall clearance 100 mm (FED400-400V / FED400UL-208V)
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 765 mm (FED400-400V / FED400UL-208V)
Height 1025 mm (FED400-400V / FED400UL-208V)
Interior height 800 mm (FED400-400V / FED400UL-208V)
Interior width 1000 mm (FED400-400V / FED400UL-208V)
Width 1235 mm (FED400-400V / FED400UL-208V)
Net weight of the unit (empty) mfr 145 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE

Description

Construction and series positioning

Behavior under load and in operation

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must stay within ±10% of 400 V at 50/60 Hz

The type plate specifies 400 V 3N~ with a ±10% tolerance at the indicated power frequency. Operation outside this band can cause controller faults, contactor chatter or heating underperformance; verify site voltage and frequency before connecting the 5-pole CEE plug.

Operating Manual FED/FED-UL (E2), Electrical connection, p. 19; Type plate, p. 8
Mounting
Keep 160 mm rear and 100 mm side clearance

The chamber dissipates heat through its surfaces; insufficient clearance in recesses or against walls causes overheating and possible damage. Installation in unventilated recesses is prohibited, and the unit must not be placed in damp areas, puddles, hazardous locations or explosive atmospheres.

Operating Manual FED/FED-UL (E2), Installation location and ambient conditions, p. 18
Process Limits
No explosive or corrosive charge materials permitted

The chamber has no explosion protection; combustible or explosive substances at working temperature must not be introduced, and the operating temperature must remain below the flash point or sublimation point of the charge. Acids, halides and other corrosive ingredients are excluded and can damage chamber components.

Operating Manual FED/FED-UL (E2), Intended use, p. 11
Control & I/O
Safety trip disables heating and shows alarm value

If the temperature safety device (class 2 standard) responds, the chamber becomes inactive with no heating and the display shows the alarm value. This is expected behavior; investigate and correct the over-temperature cause before resetting, otherwise the trip will recur.

Operating Manual FED/FED-UL (E2), Start up, p. 20; Temperature safety devices
Maintenance
Annual service by BINDER-authorized electricians required

Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. Before any maintenance, switch off at the main power switch and disconnect the power plug; the rear panel must not be removed. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required, plus tests before initial start-up and after repair.

Operating Manual FED/FED-UL (E2), Maintenance intervals, service, p. 39
Maintenance
Replace door gasket only when cold

Replacing the door gasket while the chamber is hot may damage the new gasket and compromise the seal. Allow the chamber to cool to ambient before replacement, and never clean the chamber while it is switched on (listed as foreseeable misuse).

Operating Manual FED/FED-UL (E2), Maintenance intervals, service, p. 39; Cleaning and decontamination
Process Limits
Discoloration above 150 °C is normal, not a defect

Natural oxidation of stainless steel surfaces above 150 °C produces yellowish-brown or blue tints. These are harmless and do not impair function or quality; no corrective action or warranty claim is warranted for this appearance.

Operating Manual FED/FED-UL (E2), Chamber description, p. 15

Selection & Integration Q&A

When should I choose the FED 400 over the FED 260 or FED 720?

Choose the FED 400 for 400 L batches needing +5 °C start and 90 kg total load; the FED 260 offers less volume, while the FED 720 gives more capacity but starts at +12 °C. Select by load size and lowest required temperature.

What supply and mating connection does the FED 400 require on site?

The FED400-400V needs 400 V 3N~ 50/60 Hz within ±10%, 7.4 A nominal, protected at 16 A, with a 5-pole CEE receptacle. The UL variant needs 208 V 3~ 60 Hz and a NEMA L21-20P receptacle.

What ambient and duty-cycle limits apply to installation?

Indoor use only, no unventilated recesses, damp areas or explosive atmospheres. Maintain 160 mm rear and 100 mm side clearance. The chamber controls from +5 °C above ambient to 300 °C; IP20 means no water exposure.

What maintenance and accessories are required beyond base delivery?

Annual service by BINDER-authorized electricians and annual electrical inspection per EN 50678/50699 are mandatory. RS 422 cable and APT-COM™ 4 software, HEPA filter, class 3.1 safety device, analog output and interior lighting are options not included.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FED 400 en
Datasheet ·1,1 МБ ·7 pp.
View PDF
PDF
FED (E2) 01 2025 en
Manual ·6,0 МБ ·61 pp.
View PDF