Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model FED 56

Technical Reference Summary

Covered Models & Document Scope

BINDER FED 56 drying and heating chamber (Avantgarde.Line), interior volume 60 L, in two standard variants: FED056-230V (230 V 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9010-0295) and FED056UL-120V (120 V 1~ 60 Hz, NEMA 5-15 plug, Art.-No. 9010-0296). Document also lists optional accessories and services.

Technical Data

Parameter Value Condition
Temperature range +10 °C above ambient temperature to 300 °C p.2 ↗ FED056-230V / FED056UL-120V
Temperature uniformity at 150°C 1.4 ± K p.2 ↗ FED056-230V / FED056UL-120V
Temperature fluctuation at 150°C 0.3 ± K p.2 ↗ FED056-230V / FED056UL-120V
Heating up time to 150°C 15 min p.2 ↗ FED056-230V / FED056UL-120V
Recovery time after door was opened for 30 s at 150°C 4 min p.2 ↗ FED056-230V / FED056UL-120V
Air change rate at 100°C 80 x/h p.2 ↗ FED056-230V / FED056UL-120V
Rated voltage 230 V p.2 ↗ FED056-230V
Rated voltage 120 V p.2 ↗ FED056UL-120V
Power frequency 50/60 Hz p.2 ↗ FED056-230V
Power frequency 60 Hz p.2 ↗ FED056UL-120V
Nominal power 1,1 kW p.2 ↗ FED056-230V
Nominal power 1,2 kW p.2 ↗ FED056UL-120V
Unit fuse 6,3 A p.2 ↗ FED056-230V
Unit fuse 12,5 A p.2 ↗ FED056UL-120V
Phase (nominal voltage) 1~ p.2 ↗ FED056-230V / FED056UL-120V
Interior volume 60 L p.2 ↗ FED056-230V / FED056UL-120V
Net weight of the unit (empty) 39 kg p.2 ↗ FED056-230V / FED056UL-120V
Load per rack 30 kg p.2 ↗ FED056-230V / FED056UL-120V
Permitted load 70 kg p.2 ↗ FED056-230V / FED056UL-120V
Width net 560 mm p.2 ↗ FED056-230V / FED056UL-120V
Height net 625 mm p.2 ↗ FED056-230V / FED056UL-120V
Depth net 565 mm p.2 ↗ FED056-230V / FED056UL-120V
Sound-pressure level 43 dB(A) p.3 ↗ FED056-230V / FED056UL-120V
Energy consumption at 100°C 180 Wh/h p.3 ↗ FED056-230V / FED056UL-120V
Energy consumption at 150°C 290 Wh/h p.3 ↗ FED056-230V / FED056UL-120V

Operating Limits

  • Temperature range: +10 °C above ambient temperature to 300 °C p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗
  • Permitted load: 70 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ FED056 Model series FED (drying/heating chamber) with size 56 (60 L)
suffix p.1 ↗ 230V Mains variant 230 V 1~ 50/60 Hz, CEE 7/7 plug, Art.-No. 9010-0295
suffix p.1 ↗ UL-120V Mains variant 120 V 1~ 60 Hz, NEMA 5-15 plug, Art.-No. 9010-0296

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
FED056-230V p.1 ↗ 9010-0295 230 V 1~ 50/60 Hz, CEE 7/7, Standard Standard model version
FED056UL-120V p.1 ↗ 9010-0296 120 V 1~ 60 Hz, NEMA 5-15, Standard Standard model version