BINDER FED 56
BINDER ·Made in Germany · In Active Production

BINDER FED 56

BINDER FED 56 — Drying ovens with Temperature 10…300 °C.

The BINDER FED 56 is a forced-convection drying and heating oven in the Avantgarde.Line, with a 60 L interior, +10 to 300 °C range, and 1.1 kW single-phase 230 V supply. Its 80 air changes per hour and 15 min heat-up to 150 °C distinguish it from natural-convection ED siblings, while the 56 size is the smallest of the FED series. It is intended for exact tempering, drying, and heat treatment of non-flammable solids and bulk materials in laboratory and production environments.

Temperature
10…300 °C

BINDER FED Series — Model Comparison & Selection

Model Temperature, °CAir change rate, 1/hourDepth, mmHeating up time to 150 c, minHeight, cmInterior volume, L
FED 56 Current 10–300805651562.560
FED 115 10–300326051870.5116
FED 260 10–30097601994259
FED 400 5–3001876529102.5400
FED 720 12–300587025159741

Full specifications

Core Performance & Functional Specifications
Temperature 10...300 °C
Air change rate 80 x/h
Heating time to 150°C 15 min
Interior volume 60 L
Load per rack 30 kg (FED056-230V / FED056UL-120V)
Nominal power 1.1 kW
Permitted load 70 kg (FED056-230V / FED056UL-120V)
Temperature fluctuation 0.3 K
Temperature uniformity 1.4 K
Article Number mfr 9010
Recovery time after door was opened for 30 s at 150°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 565 mm (FED056-230V / FED056UL-120V)
Height 625 mm (FED056-230V / FED056UL-120V)
Width 560 mm (FED056-230V / FED056UL-120V)
Net weight of the unit (empty) mfr 39 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Electrical
Supply must stay within ±10 % of 230 V nominal

The FED 56 is rated for 230 V 1~ 50 Hz with a nominal power of 1.1 kW and an external 6.3 A time-lag fuse. Voltage fluctuations beyond ±10 % can cause malfunction or damage; verify the supply quality and fuse rating before connection. The chamber uses an IEC connector plug with shockproof plug, so a compatible outlet and upstream circuit protection must be provided.

Operating Manual 07/2025, Section 'Electrical connection data and installation', p.18; Troubleshooting, p.70
Mounting
Install indoors with sufficient ventilation and wall clearance

The chamber must be placed on a non-flammable surface in a well-ventilated indoor area, not in unventilated recesses or explosive atmospheres. Ambient temperature must remain between +18 °C and +40 °C, humidity ≤70 % r.h. non-condensing, and installation altitude ≤2000 m. Observe minimum wall distances to ensure airflow and access.

Operating Manual 07/2025, Section 'Installation location', p.20; Quick Start Guide 03/2023, p.18
Control & I/O
Set safety controller 2–5 °C above working set-point

The safety controller acts as an absolute maximum temperature limit. After any change to the temperature set-point, re-adjust the safety controller set-point to approximately 2 °C to 5 °C above it. The permissible range is 10 °C to 300 °C with 1 degree accuracy. Failure to set it correctly may cause unnecessary tripping or inadequate protection.

Operating Manual 07/2025, Section 'Overtemperature protection', p.20; Quick Start Guide 03/2023, p.20
Maintenance
Annual maintenance and electrical safety test required

The chamber must be maintained at least once a year by licensed electricians or BINDER-authorized experts. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, and a test is required before initial startup and after any maintenance or repair. Warranty may be void if maintenance is performed by non-authorized personnel.

Operating Manual 07/2025, Section 'Maintenance and service', p.69
Process Limits
Charging material must be non-flammable and below flash point

The chamber has no explosion protection. Do not introduce combustible or explosive substances, dust, or air-solvent mixtures. The operating temperature must remain below the flash point or sublimation point of the material. Corrosive ingredients such as acids and halides that may damage stainless steel, aluminum, or copper are also prohibited.

Operating Manual 07/2025, Section 'Intended use', p.12-13

Selection & Integration Q&A

When should I choose the FED 56 over a larger FED model or an ED natural-convection oven?

Choose FED 56 for small loads up to 60 L requiring forced convection and adjustable fan speed for uniform drying. For larger volumes, select FED 115/260/400/720. If natural convection is sufficient and fan adjustability is not needed, the ED series is an alternative, but FED offers enhanced timer functions and standard Ethernet.

What maintenance is required and what accessories are not included?

Annual maintenance by authorized personnel and electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 are required. Base delivery excludes accessories such as APT-COM 4 software, Pt100 sensor, analog output, HEPA filter, and inert gas connection. Cleaning agents (art. 1002-0016, 1002-0022) must be ordered separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model FED 56 en
Datasheet ·388,3 КБ ·8 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD BF ED FD FED (E3.1) 07 2025 en
Manual ·18,7 МБ ·122 pp.
View PDF