FP (E3.1) 06 2026

Technical Reference Summary

Covered Models & Document Scope

This operating manual (FP (E3.1), issue 06/2026, Art. No. 7001-0434) covers the BINDER Avantgarde.Line FP series of drying and heating ovens with forced convection, program functions and RD4 microprocessor temperature controller. Covered models: FP 56 (E3.1) FP056-230V Art. 9010-0363/9110-0363; FP 115 (E3.1) FP115-230V Art. 9010-0364/9110-0364; FP 260 (E3.1) FP260-230V Art. 9010-0365/9110-0365; FP 720 (E3.1) FP720-400V Art. 9010-0366/9110-0366; FP 056-UL (E3.1) FP056-120V Art. 9010-0367/9110-0367; FP 115-UL (E3.1) FP115-120V Art. 9010-0368/9110-0368; FP 260-UL (E3.1) FP260-240V Art. 9010-0369/9110-0369; FP 720-UL (E3.1) FP720-208V Art. 9010-0370/9110-0370. Options/accessories addressed include APT-COM 4 software, flexible Pt100 object temperature sensor, analog output, HEPA fresh air filter, mostly gas-tight version and inert gas connection.

Section-by-Section Technical Breakdown

pp. 17-20
Chamber description, rear panel and zero-voltage switching contacts

Describes the chamber overview, rear connections, triangular instrument panel, main power switch on size 720, and the zero-voltage switching contacts. The rear includes a SUB-D socket for ALARM/SWITCH CONTACT and an optional SUB-D socket for ANALOG OUTPUT. Pin allocation of the alarm/contact socket is given.

  • SUB-D socket 'ALARM / SWITCH CONTACT' provides zero-voltage switching contacts.
  • SUB-D socket 'ANALOG OUTPUT' is an option for analog temperature output.
  • Size 720 has a main power switch on the chamber rear.
  • Triangular instrument panel is described in chap. 2.3.
pp. 21-25
Installation, electrical connection and exhaust connection

Covers installation location and ambient conditions, electrical connection, optional connection to an exhaust/ventilation system, and insertion of racks. Minimum distances and ventilation requirements must be observed to avoid overheating.

  • Chamber must not be installed in unventilated recesses; sufficient ventilation for heat dispersal is required.
  • Prescribed minimum distances for installation are given in chap. 3.4.
  • Electrical connection per chap. 4.1; optional exhaust/ventilation connection per chap. 4.2.
  • Racks are inserted per chap. 4.3.
pp. 27-30
RD4 controller functional overview, menu structure and start-up

Describes the RD4 chamber controller menu structure and access levels, behavior during and after power failures, behavior when opening the door, and start-up including loading, controller settings and decimal place setting.

  • Menu structure and access levels per chap. 5.1.
  • Performance during and after power failures per chap. 5.2.
  • Behavior when opening the door per chap. 5.3.
  • Start-up settings and displayed decimal places per chap. 6.2 and 6.3.
pp. 31-32
Set-point entry: temperature, fan speed and air flap

Covers entry of the temperature set-point, fan speed set-point and adjustment of the air flap position. These settings define the thermal and airflow operating conditions of the chamber.

  • Temperature set-point entry per chap. 7.1.
  • Fan speed set-point entry per chap. 7.2.
  • Air flap position adjustment per chap. 7.3.
pp. 34-35
Authorization levels and password protection

Describes password request and assignment/modification of User and Admin passwords. These levels protect controller settings from unauthorized changes.

  • Password request per chap. 9.1.
  • User password assignment/modification per chap. 9.2.1.
  • Admin password assignment/modification per chap. 9.2.2.
pp. 35-40
Failsafe function and overtemperature protection

Covers the failsafe function, overtemperature protective device class 1, and overtemperature safety controller class 2/3.1 including selection, mode, setting value, alarm messages/measures and function check.

  • Overtemperature protective device class 1 per chap. 11.1.
  • Safety controller class 2 (temperature limiter) or class 3.1 selectable per chap. 11.2.1.
  • Safety controller mode and value setting per chap. 11.2.2 to 11.2.4.
  • Alarm state message and measures per chap. 11.2.5.
  • Function check per chap. 11.3.
pp. 40-43
General controller settings

Covers menu language, temperature unit, date, time, language selection at restart, chamber address and display brightness.

  • Menu language selection per chap. 12.1.
  • Temperature unit selection per chap. 12.2.
  • Date and time setting per chap. 12.3 and 12.4.
  • Chamber address per chap. 12.6; display brightness per chap. 12.7.
pp. 44
Tolerance range settings

Covers setting the delay time for tolerance range alarm and setting the temperature tolerance range. These settings define when a tolerance alarm is triggered.

  • Delay time for tolerance range alarm per chap. 13.1.
  • Temperature tolerance range per chap. 13.2.
pp. 45-47
Notification and alarm functions

Describes information messages, alarm messages, activation/deactivation of the audible alarm buzzer, and the zero-voltage relay alarm contact.

  • Information messages per chap. 14.1.
  • Alarm messages per chap. 14.2.
  • Audible alarm on/off per chap. 14.3.
  • Zero-voltage relay alarm contact per chap. 14.4.
pp. 48-49
Program start/stop function

Covers start time program, start week program, stop program and performance after power failures.

  • Start time program per chap. 15.1.
  • Start week program per chap. 15.2.
  • Stop program per chap. 15.3.
  • Performance after power failures per chap. 15.4.
pp. 49-54
Ethernet network settings

Covers showing network settings (MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name) and changing the network configuration (automatic/manual IP and DNS assignment, IP address, subnet mask, gateway, DNS server).

  • Show MAC address per chap. 16.1.1.
  • Show IP address, subnet mask, gateway, DNS server and DNS chamber name per chap. 16.1.2 to 16.1.6.
  • Select automatic/manual IP assignment per chap. 16.2.1.
  • Assign IP address, subnet mask, gateway and DNS server per chap. 16.2.3 to 16.2.6.
pp. 54-56
Data recorder

Covers recorded data, storage capacity, setting the storage rate for the 'DL1' recorder data, and deleting the data recorder.

  • Recorded data per chap. 17.1.
  • Storage capacity per chap. 17.2.
  • Storage rate for 'DL1' recorder data per chap. 17.3.
  • Deleting the data recorder per chap. 17.4.
pp. 56-58
USB menu: data transfer via USB interface

Covers connecting a USB stick, import and export functions, ongoing data transfer, error during data transmission, and removing the USB stick.

  • Connecting the USB stick per chap. 18.1.
  • Import function per chap. 18.2; export functions per chap. 18.3.
  • Ongoing data transfer per chap. 18.4.
  • Error during data transmission per chap. 18.5; removing the USB stick per chap. 18.6.
pp. 58-60
Options and accessories

Describes APT-COM 4 Multi Management Software, object temperature display with flexible Pt100 sensor, analog output for temperature, HEPA fresh air filter, mostly gas-tight version, and inert gas connection with mostly gas-tight version.

  • APT-COM 4 software accessory per chap. 19.1.
  • Pt100 object temperature sensor option per chap. 19.2.
  • Analog output for temperature option per chap. 19.3.
  • HEPA fresh air filter option per chap. 19.4.
  • Mostly gas-tight version and inert gas connection per chap. 19.5 and 19.6.
pp. 62-64
Cleaning and decontamination

Covers cleaning and decontamination/chemical disinfection of the chamber. Cleaning must not be performed while the chamber is turned on, and water must not penetrate the chamber.

  • Cleaning per chap. 20.1.
  • Decontamination/chemical disinfection per chap. 20.2.
  • Cleaning while the chamber is turned on is listed as foreseeable misuse.
pp. 65-68
Maintenance and service, troubleshooting, repair, testing

Covers personnel qualification, maintenance intervals and service, service reminder, simple troubleshooting, and returning the chamber to BINDER. Annual maintenance and annual electrical repeat inspection are required.

  • Maintenance work at least once a year; warranty void if performed by non-authorized personnel.
  • Annual repeat electrical inspection and test before initial start-up and before resuming operation after maintenance/repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702.
  • Service reminder: keep OK button pressed 5 seconds to display remaining days until service due.
  • After one year of operation, 'Service due!' message appears; after one week it reappears.
  • Replace door gasket only when cold.
pp. 72-80
Technical description and technical data

Covers factory calibration and adjustment, definition of usable volume, overcurrent protection, FP technical data, equipment/options/accessories, spare parts, and dimensions for sizes 56, 115, 260 and 720.

  • Factory calibration and adjustment per chap. 23.1.
  • Definition of usable volume per chap. 23.2.
  • Overcurrent protection per chap. 23.3.
  • FP technical data per chap. 23.4.
  • Dimensions for sizes 56, 115, 260 and 720 per chap. 23.7 to 23.10.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal temperature of the FP 115 is 300 °C (572 °F). p.1 ↗ Maximum intended operating temperature for this model is 300 °C; process requirements above this are not supported. FP 115 (E3.1)
FP 115 (E3.1) nominal power 1.40 kW, nominal current 6.10 A at 230 V / 50 Hz; also rated 230 V / 60 Hz, 1 N~. p.1 ↗ Electrical supply must be 230 V single-phase with ±10% tolerance at the indicated power frequency; size the circuit for at least 6.10 A. FP 115 (E3.1)
FP 115-UL (E3.1) nominal power 1.40 kW, nominal current 11.7 A at 120 V / 50 Hz, 1 N~. p.1 ↗ UL variant requires a 120 V single-phase supply sized for 11.7 A. FP 115-UL (E3.1)
IP protection type 20 according to EN 60529. p.1 ↗ Chamber is protected against solid objects >12.5 mm but not against water; do not install in damp areas or puddles and do not spray with water. whole series
Temperature safety device according to DIN 12880, class 2/3.1 (selectable). p.1 ↗ Overtemperature protection can be configured as class 2 temperature limiter or class 3.1 safety controller; selection affects alarm behavior and reset requirements. whole series
Nominal voltage tolerance ±10% at the indicated power frequency. p.1 ↗ Supply voltage must remain within ±10% of nominal; out-of-tolerance supply may affect temperature performance and safety. whole series
Chamber must NOT be installed in unventilated recesses; prescribed minimum distances per chap. 3.4 must be observed. p.1 ↗ Provide sufficient ventilation for heat dispersal and maintain wall clearances to prevent overheating damage. whole series
Chambers are designed for indoor use only. p.1 ↗ Installation site must be inside a building and meet the ambient conditions of chap. 3.4. whole series
Operating temperature must lie below the flash point or below the sublimation point of the charging material; no explosive mixture with air may form. p.1 ↗ Verify solvent and material properties before loading; the chamber has no explosion protection measures. whole series
Loading material must not contain corrosive ingredients such as acids and halides that may damage stainless steel, aluminum and copper components. p.1 ↗ Corrosive damage caused by such ingredients is excluded from BINDER liability; select compatible charge materials. whole series
Maintenance work must be performed at least once a year; warranty becomes void if maintenance is conducted by non-authorized personnel. p.21 ↗ Schedule annual maintenance by licensed electricians or BINDER-authorized experts to preserve warranty. whole series
Annual repeat electrical inspection and a test prior to initial start-up and prior to resuming operation after maintenance or repair are required; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.21 ↗ Plan electrical safety testing at commissioning, after service, and annually. whole series
Service reminder: keep the OK button pressed down for 5 seconds to display remaining days until maintenance is due. p.21 ↗ Operating personnel can check the service countdown without service intervention. whole series
After the recommended maintenance interval (one year of operation) the message 'Service due!' appears; after one week of operation the message reappears. p.21 ↗ The controller provides a recurring annual service reminder that must be confirmed with OK. whole series
Replace the door gasket only when cold. p.21 ↗ Hot gasket replacement may damage the gasket; allow the chamber to cool before service. whole series
Chamber must NOT become wet during operation or maintenance; do NOT remove the rear panel. p.21 ↗ Disconnect power before maintenance and avoid water ingress to prevent deadly electric shock. whole series
Connect only external devices to the Ethernet and USB interfaces that comply with EN 61010-1:2010 or EN 60950-1:2006 mod. p.1 ↗ Verify compliance of connected IT equipment to maintain safety of the system. whole series
The chambers are not classified as medical devices per Regulation (EU) No 2017/745 and are not qualified to sterilize medical devices. p.1 ↗ Do not use for medical device sterilization applications. whole series
Zero-voltage switching contacts are available via SUB-D socket 'ALARM / SWITCH CONTACT'; optional analog output via SUB-D socket 'ANALOG OUTPUT'. p.2 ↗ External alarm or monitoring systems can be wired to the zero-voltage contacts; analog temperature output requires the option. whole series
Size 720 has a main power switch on the chamber rear. p.2 ↗ For size 720, both the on/off switch and the rear main power switch must be considered for disconnection before maintenance. FP 720 (E3.1) and FP 720-UL (E3.1)

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Service due! p.66 ↗ Recommended maintenance interval (one year of operation) has elapsed. Press the OK button to confirm the message. The message reappears after one week of operation. chap. 21.3