Data Sheet Model KB ECO 1020

Technical Reference Summary

Covered Models & Document Scope

Data sheet for BINDER KB ECO 1020 cooling incubator (thermoelectric cooling), interior volume 1,020 L. Two standard variants: KBECO1020-230V (Art.-No. 9020-0425, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug) and KBECO1020UL-120V (Art.-No. 9020-0428, 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug). Accessories and services are listed separately.

Technical Data

Parameter Value Condition
Temperature range 0...70 °C p.2 ↗ KBECO1020-230V; max. 26 °C under ambient temperature
Temperature range 0...70 °C p.2 ↗ KBECO1020UL-120V; max. 26 °C under ambient temperature
Temperature uniformity at 37°C 0.1 ± K p.2 ↗ KBECO1020-230V
Temperature fluctuation at 37°C 0.1 ± K p.2 ↗ KBECO1020-230V
Max. heat compensation at 40°C 500 W p.2 ↗ KBECO1020-230V
Recovery time after door was opened for 30 s at 37°C 4 min p.2 ↗ KBECO1020-230V
Rated voltage 200...230 V p.2 ↗ KBECO1020-230V
Rated voltage 100...120 V p.2 ↗ KBECO1020UL-120V
Power frequency 50/60 Hz p.2 ↗ both variants
Nominal power 1,3 kW p.2 ↗ both variants
Unit fuse 16,0 A p.2 ↗ both variants
Phase (nominal voltage) 1~ p.2 ↗ both variants
Interior volume 1,020 L p.2 ↗ both variants
Net weight of the unit (empty) 327 kg p.2 ↗ both variants
Load per rack 45 kg p.2 ↗ both variants
Permitted load 150 kg p.2 ↗ both variants
Wall clearance back 100 mm p.2 ↗ both variants
Wall clearance sidewise 100 mm p.2 ↗ both variants
Width net 1,250 mm p.2 ↗ both variants
Height net 1,925 mm p.2 ↗ both variants
Depth net 1,149 mm p.2 ↗ both variants
Interior width 973 mm p.2 ↗ both variants
Interior height 1,250 mm p.2 ↗ both variants
Interior depth 836 mm p.3 ↗ both variants
Sound-pressure level 47 dB(A) p.3 ↗ both variants

Operating Limits

  • Temperature range: 0...70 °C p.2 ↗
  • Max. temperature below ambient temperature: 26 °C p.2 ↗
  • Max. heat compensation at 40 °C: 500 W p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 45 kg p.2 ↗
  • Wall clearance back: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-2 p.1 ↗ KB Product series: cooling incubator
3-6 p.1 ↗ ECO Model version: thermoelectric cooling, energy-efficient
7-10 p.1 ↗ 1020 Size / interior volume class: 1,020 L
suffix p.1 ↗ -230V Mains variant: 200...230 V 1~ 50/60 Hz, CEE 7/7 plug
suffix p.1 ↗ UL-120V Mains variant: 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
KBECO1020-230V p.1 ↗ 9020-0425 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard Standard model version
KBECO1020UL-120V p.1 ↗ 9020-0428 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard UL model version