Induztr
Engineering Digest Document Viewer Related Models

Data Sheet Model KBF S ECO 240

Technical Reference Summary

Covered Models & Document Scope

Model KBF-S ECO 240 constant climate chamber with thermoelectric cooling. Two standard variants: KBFSECO240-230V (Art.-No. 9020-0416, 200...230 V, CEE 7/7 plug) and KBFSECO240UL-120V (Art.-No. 9020-0420, 100...120 V, NEMA 5-20 plug).

Technical Data

Parameter Value Condition
Interior volume 247 L p.1 ↗ KBF-S ECO 240, 50/60 Hz
Temperature range 0...70 °C p.2 ↗ max. 28 °C under ambient temperature
Max. heat compensation at 40°C 250 W p.2 ↗
Temperature range with humidity 5...70 °C p.2 ↗ max. 28 °C under ambient temperature
Humidity range 10...80 % RH p.2 ↗
Temperature uniformity at 25°C and 60% RH 0.1 ± K p.2 ↗
Temperature uniformity at 40°C and 75% RH 0.2 ± K p.2 ↗
Temperature fluctuation at 25°C and 60% RH 0.1 ± K p.2 ↗
Temperature fluctuation at 40°C and 75% RH 0.1 ± K p.2 ↗
Humidity fluctuation at 25°C and 60% RH 0.3 ± % RH p.2 ↗
Humidity fluctuation at 40°C and 75% RH 0.3 ± % RH p.2 ↗
Rated voltage 200...230 V p.2 ↗ KBFSECO240-230V
Rated voltage 100...120 V p.2 ↗ KBFSECO240UL-120V
Power frequency 50/60 Hz p.2 ↗
Nominal power 0,8 kW p.2 ↗
Unit fuse 16,0 A p.2 ↗
Phase (nominal voltage) 1~ p.2 ↗
Net weight of the unit (empty) 146 kg p.2 ↗
Load per rack 30 kg p.2 ↗
Permitted load 100 kg p.2 ↗
Width net 932 mm p.3 ↗
Height net 1,461 mm p.3 ↗
Depth net 796 mm p.3 ↗
Sound-pressure level 46 dB(A) p.3 ↗
Number of shelves (std./max.) 2/9 p.3 ↗

Operating Limits

  • Max. ambient temperature for full temperature range: 28 °C p.2 ↗
  • Wall clearance back: 100 mm p.3 ↗
  • Wall clearance sidewise: 180 mm p.3 ↗

Type Code Decoder

Position Token Meaning
1-4 p.1 ↗ KBFS Series: constant climate chamber KBF-S
5-7 p.1 ↗ ECO Model version: thermoelectric cooling
8-10 p.1 ↗ 240 Size: interior volume 247 L
11-15 p.1 ↗ -230V Voltage variant: 200...230 V, CEE 7/7 plug
11-15 p.1 ↗ UL-120V Voltage variant: 100...120 V, NEMA 5-20 plug

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
KBFSECO240-230V p.1 ↗ 9020-0416 200...230 V 1~ 50/60 Hz, CEE 7/7 plug, Standard
KBFSECO240UL-120V p.1 ↗ 9020-0420 100...120 V 1~ 50/60 Hz, NEMA 5-20 plug, Standard

Related models