Data Sheet Model M 115

Technical Reference Summary

Covered Models & Document Scope

Model M 115 drying and heating chamber with forced convection and advanced program functions; standard version M115-230V, 230 V 1~ 50/60 Hz, Art.-No. 9010-0372

Technical Data

Parameter Value Condition
Temperature range +23 °C above ambient temperature to 300 °C p.2 ↗ M115-230V
Temperature uniformity at 150°C 1.0 ± K p.2 ↗ M115-230V
Temperature fluctuation at 150°C 0.2 ± K p.2 ↗ M115-230V
Heating up time to 150°C 11 min p.2 ↗ M115-230V
Recovery time after door was opened for 30 s at 150°C 6 min p.2 ↗ M115-230V
Air change rate at 100°C 105 x/h p.2 ↗ M115-230V
Rated voltage 230 V p.2 ↗ M115-230V
Power frequency 50/60 Hz p.2 ↗ M115-230V
Nominal power 2,4 kW p.2 ↗ M115-230V
Unit fuse 12,5 A p.2 ↗ M115-230V
Phase (Nominal voltage) 1~ p.2 ↗ M115-230V
Interior volume 113 L p.2 ↗ M115-230V
Net weight of the unit (empty) 58 kg p.2 ↗ M115-230V
Load per rack 30 kg p.2 ↗ M115-230V
Permitted load 150 kg p.2 ↗ M115-230V
Wall clearance back 160 mm p.2 ↗ M115-230V
Wall clearance sidewise 100 mm p.2 ↗ M115-230V

Operating Limits

  • Temperature range: +23 °C above ambient temperature to 300 °C p.2 ↗
  • Permitted load: 150 kg p.2 ↗
  • Load per rack: 30 kg p.2 ↗
  • Wall clearance back: 160 mm p.2 ↗
  • Wall clearance sidewise: 100 mm p.2 ↗

Type Code Decoder

Position Token Meaning
1-4 p.1 ↗ M115 Model series and size (M series, 115 size)
field 2 p.1 ↗ 230V Rated voltage 230 V, 1~, 50/60 Hz

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
M115-230V p.1 ↗ 9010-0372 Standard, 230 V 1~ 50/60 Hz Standard model version