BINDER M 115
BINDER M 115 — Drying ovens with Temperature 23…300 °C.
The BINDER M 115 is a forced-convection drying and heating oven with a 113 L interior, operating from +23 to 300 °C via a 2.4 kW single-phase 230 V supply. It sits between the smaller M 56 and larger M 260/M 720 in the M series, distinguished by its 230 V single-phase connection and 105 air changes per hour. It is used for reproducible drying, heating, and thermal testing of laboratory and industrial loads up to 150 kg.
BINDER M Series — Model Comparison & Selection
Full specifications
| Temperature | 23...300 °C |
| Air change rate | 105 x/h |
| Heating time to 150°C | 11 min |
| Interior volume | 113 L |
| Load per rack | 30 kg |
| Nominal power | 2.4 kW |
| Permitted load | 150 kg |
| Temperature fluctuation | 0.2 K |
| Temperature uniformity | 1.0 K |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 6 min |
| Depth | 636 mm |
| Height | 732 mm |
| Width | 710 mm |
| Net weight of the unit (empty) mfr | 58 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 12.5 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
The M 115 (E3.1) is a forced-convection chamber with a 113 L usable volume, net dimensions of 710 mm wide × 732 mm high × 636 mm deep, and a net weight of 58 kg. Unlike the M 260 and M 720, which require a 400 V three-phase supply and have their main power switch on the rear panel, the M 115 operates on 230 V single-phase (1 N~, 50/60 Hz) with a nominal current of 10.5 A and a 12.5 A unit fuse. This makes it the largest M-series model that can be connected to a standard 230 V laboratory circuit. The M 56 shares the same +23 °C lower temperature limit, while the M 260 and M 720 extend down to +14 °C.
Thermal behavior and load capacity
- Temperature range: +23 to 300 °C; temperature fluctuation 0.2 K, uniformity 1.0 K.
- Heating-up time to 150 °C: 11 min; recovery time after 30 s door opening at 150 °C: 6 min.
- Air-change rate: 105 x/h, supporting rapid drying and solvent removal.
- Maximum load: 150 kg total, 30 kg per rack.
The chamber uses the MB2 microprocessor controller with selectable overtemperature protection class 2 (temperature limiter) or class 3.1, per DIN 12880. The high air-change rate and forced convection make the M 115 suitable for processes requiring efficient moisture removal, but the load must not contain substances that are combustible or explosive at the working temperature; the set-point must remain below the flash point or sublimation point of the charge.
Installation, control, and operation
The M 115 requires 100 mm side clearance and 160 mm rear clearance, and must not be installed in unventilated recesses. It is rated IP 20 and must not be exposed to water or installed in damp areas. The rear panel provides a SUB-D socket for a zero-voltage relay alarm contact (standard) and an optional analog output for temperature. The MB2 controller supports fixed-value operation, timer programs, time programs, and week programs, with user management and an event list. Optional accessories include APT-COM 4 software (networking up to 100 chambers), a flexible Pt100 object temperature sensor, HEPA fresh air filter, and a mostly gas-tight version with inert gas connection.
Integration & Operating Notes
The M 115 draws 10.5 A at 2.40 kW nominal. Verify that the branch circuit can supply this continuously and that the voltage stays within ±10% of 230 V at the indicated frequency; out-of-range supply can cause controller faults or heater damage. A 12.5 A unit fuse is fitted.
Selection & Integration Q&A
Choose M 115 when you need 113 L capacity and only have a 230 V single-phase supply; it is the largest M-series model on 230 V. The M 56 is smaller (56 L) but same temperature range; the M 260 and M 720 require 400 V three-phase and offer lower minimum temperature (+14 °C).
The rear provides a SUB-D socket for a zero-voltage relay alarm contact (standard) and an optional SUB-D analog output for temperature. For networked control, APT-COM 4 software can connect up to 100 chambers. Ensure the alarm contact is wired to a suitable monitoring device if unattended operation is required.
Annual maintenance by BINDER-authorized personnel and annual electrical safety testing per EN 50678/VDE 0701 are required. Base delivery excludes APT-COM 4 software, Pt100 object temperature sensor, analog output, HEPA fresh air filter, and gas-tight/inert gas options. Door gasket replacement only when cold.