Data Sheet Model MKT 720

Technical Reference Summary

Covered Models & Document Scope

BINDER MKT 720 dynamic climate chamber, 734 L interior volume, temperature range -70...180 °C. Two variants: MKT720-400V (standard, 400 V 3~/N/PE 50 Hz, Art.-No. 9020-0387) and MKT720-480V-C (with voltage and frequency converter, 480 V 3~ 60 Hz, Art.-No. 9020-0365). Document also lists accessories and services for MKT 720.

Technical Data

Parameter Value Condition
Temperature range -70...180 °C p.2 ↗ MKT720-400V, 50 Hz
Temperature uniformity dependent on set value 0.1...2.0 ± K p.2 ↗ MKT720-400V, 50 Hz
Temperature fluctuation dependent on set value 0.1...0.5 ± K p.2 ↗ MKT720-400V, 50 Hz
Cooling down time from 180°C to -70°C 120 min p.2 ↗ MKT720-400V, 50 Hz
Rated voltage 400 V p.2 ↗ MKT720-400V, 50 Hz
Power frequency 50 Hz p.2 ↗ MKT720-400V
Nominal power 13 kW p.2 ↗ MKT720-400V, 50 Hz
Unit fuse 32,0 A p.2 ↗ MKT720-400V, 50 Hz
Phase (nominal voltage) 3~/N/PE p.2 ↗ MKT720-400V, 50 Hz
Interior volume 734 L p.2 ↗ MKT720-400V, 50 Hz
Net weight of the unit (empty) 610 kg p.2 ↗ MKT720-400V, 50 Hz
Load per rack 40 kg p.2 ↗ MKT720-400V, 50 Hz
Permitted load 160 kg p.2 ↗ MKT720-400V, 50 Hz
Wall clearance back 300 mm p.2 ↗ MKT720-400V, 50 Hz
Wall clearance sidewise 200 mm p.2 ↗ MKT720-400V, 50 Hz
Width net 1,580 mm p.2 ↗ MKT720-400V, 50 Hz
Height net 2,005 mm p.2 ↗ MKT720-400V, 50 Hz
Depth net 1,140 mm p.2 ↗ MKT720-400V, 50 Hz
Interior width 1,200 mm p.3 ↗ MKT720-400V, 50 Hz
Interior height 1,020 mm p.3 ↗ MKT720-400V, 50 Hz
Interior depth 600 mm p.3 ↗ MKT720-400V, 50 Hz
Unit doors 1 p.3 ↗ MKT720-400V, 50 Hz
Sound-pressure level 65 dB(A) p.3 ↗ MKT720-400V, 50 Hz
Number of shelves (std./max.) 1/11 p.3 ↗ MKT720-400V, 50 Hz
Rated voltage 480 V p.2 ↗ MKT720-480V-C, 60 Hz

Operating Limits

  • Temperature range: -70...180 °C p.2 ↗
  • Wall clearance back: 300 mm p.2 ↗
  • Wall clearance sidewise: 200 mm p.2 ↗
  • Permitted load: 160 kg p.2 ↗
  • Load per rack: 40 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-3 p.1 ↗ MKT Series: dynamic climate chamber for rapid temperature changes with extended low temperature range
4-6 p.1 ↗ 720 Model size / interior volume class (734 L)
7-9 p.1 ↗ 400 Rated voltage variant: 400 V
7-9 p.1 ↗ 480 Rated voltage variant: 480 V

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
MKT720-400V p.1 ↗ 9020-0387 400 V 3~/N/PE 50 Hz, standard version CEE 32 / 6 H plug 5-pin
MKT720-480V-C p.1 ↗ 9020-0365 480 V 3~ 60 Hz, with voltage and frequency converter CEE 32 / 6 H plug 5-pin