BINDER MKT 720
BINDER MKT 720 — Climate chambers with Temperature -70…180 °C.
The BINDER MKT 720 is a 734 L alternating climate chamber with a -70...180 °C range, using a cascade refrigeration system for deep-temperature cycling. As the largest MKT model, it delivers 13 kW nominal power and 4.5 K/min average heating between -40 and 180 °C, with 4.0 K/min cooling between 180 and -40 °C. It is intended for material and component testing requiring rapid temperature changes under program control.
BINDER MKT Series — Model Comparison & Selection
Full specifications
| Temperature | -70...180 °C |
| Interior volume | 734 L |
| Load per rack | 40 kg (MKT720-400V, 50 Hz) |
| Nominal power | 13 kW |
| Article Number mfr | 9020 |
| Depth | 1140 mm (MKT720-400V, 50 Hz) |
| Height | 2005 mm (MKT720-400V, 50 Hz) |
| Interior height | 1020 mm (MKT720-400V, 50 Hz) |
| Interior width | 1200 mm (MKT720-400V, 50 Hz) |
| Width | 1580 mm (MKT720-400V, 50 Hz) |
| Net weight of the unit (empty) mfr | 610 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 32.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
| Cooling down time from 180°C to -70°C mfr | 120 min |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 4.5 K/min |
Description
Construction and series positioning
The MKT 720 is the largest chamber in the MKT series, offering 734 L of interior volume with dimensions of 1,200 mm width, 1,020 mm height, and 1,140 mm depth. Unlike the smaller MKT 115 and MKT 240, which share the same -70...180 °C range, the MKT 720 is designed for higher throughput and larger test loads, with a net weight of 610 kg and a load capacity of 40 kg per rack. The chamber is equipped with a 13 kW nominal power system and operates on a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. The standard model (MKT720-400V, Art.-No. 9020-0387) is distinguished from the -C variant, which includes a voltage and frequency changer for 480 V 60 Hz operation.
Thermal behavior and control
Temperature uniformity ranges from 0.1 to 2.0 K and fluctuation from 0.1 to 0.5 K, depending on the set value. The chamber achieves an average heating rate of 4.5 K/min between -40 °C and 180 °C and an average cooling rate of 4.0 K/min between 180 °C and -40 °C, with a cooling-down time of 120 minutes from 180 °C to -70 °C. These performance figures follow IEC 60068-3-5 and are supported by a cascade refrigeration system using R452a and R23 refrigerants. The MB2 controller provides program control, fixed-value operation, timer and week programs, and a chart recorder display for documentation.
Installation and operation
Installation requires sufficient ventilation and adherence to minimum distances to walls and ceilings, as the chamber must not be installed in unventilated recesses. The chamber is not explosion-protected and must not be operated in hazardous locations or with combustible materials. Electrical connection must be made by qualified personnel, and the chamber includes a rear power switch (for sizes 115 and above) in addition to the main power switch. The Ethernet interface allows networking with APT-COM™ 4 software for remote monitoring and data logging. Optional RS485, analog temperature outputs, and a flexible Pt 100 sensor are available for integration into existing systems.
Integration & Operating Notes
The MKT 720 draws 13 kW nominal power and must be connected to a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. Verify that the site provides this capacity and that the over-current protection is sized accordingly; a dedicated circuit is recommended to avoid nuisance trips during heating and cooling cycles.
The chamber has no explosion protection measures. It must not be operated in potentially explosive areas, and no combustible dust or air-solvent mixtures may be present near the chamber. Ensure that test materials do not release flammable vapors at working temperatures, as this could create an ignition hazard.
A maintenance message appears after 8760 operating hours or two years. After confirmation, the message reappears every two weeks until reset by BINDER Service. Schedule service intervention before this interval to avoid recurring messages and to maintain warranty and performance. Annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required.
The chamber is equipped with an Ethernet serial interface and can be networked with BINDER APT-COM™ 4 Multi Management Software (accessory). This allows remote monitoring, data logging, and integration into laboratory management systems. Network configuration is described in chapter 14.5 of the manual; APT-COM™ 4 requires separate registration.
Selection & Integration Q&A
Choose the MKT 720 when you need a large interior volume (734 L) for testing bulky components or high-throughput batches. The MKT 115 and MKT 240 offer the same temperature range but smaller volumes (115 L and 240 L respectively). The MKT 720 also provides higher nominal power (13 kW) to handle larger loads, but requires a 400 V three-phase supply with 32 A fusing, which may not be available in all labs.
The MKT 720 requires a 400 V, 3 N/PE, 50 Hz supply with a 32 A unit fuse. It includes an Ethernet interface for networking with APT-COM™ 4 software, enabling integration with building management or laboratory monitoring systems. Optional RS485 and analog temperature outputs are available for connection to PLCs or data loggers.
Install the chamber with sufficient ventilation and observe minimum distances to walls and ceilings as specified in chapter 3.4 of the manual. Do not install in unventilated recesses or damp areas. The chamber must not be operated in explosive atmospheres or hazardous locations. Ambient temperature and humidity limits are given in the technical data; ensure the site meets these to avoid condensation and overheating.
A maintenance message appears after 8760 operating hours or two years, requiring service by BINDER. Annual electrical safety testing is mandatory. Base delivery does not include the Ethernet cable for APT-COM™ 4, RS485 interface, analog outputs, or the flexible Pt 100 sensor; these must be ordered separately. Also, the compressed air connection and dryer are optional for specific test profiles.