Engineering Digest Document Viewer Related Models

VD (E3.1) 07 2025

Technical Reference Summary

Covered Models & Document Scope

This operating manual (VD / VD-UL (E3.1), issue 07/2025, Art. No. 7001-0383) covers the BINDER Vacuum Drying Oven series VD 23, VD 56 and VD 115, each available in a 230 V version (VD023-230V Art. 9630-0001, VD056-230V Art. 9630-0002, VD115-230V Art. 9630-0003) and a UL 120 V version (VD023UL-120V Art. 9630-0005, VD056UL-120V Art. 9630-0006, VD115UL-120V Art. 9630-0007). Standard controller is the RD4 microprocessor program controller; the MB2 microprocessor program controller is optional. Options and accessories include APT-COM 4 software, analog outputs for temperature and pressure, a 9-pole measuring access port for vacuum, object temperature display with flexible Pt 100 sensor, LED interior lighting, pump module with pump stand and connection kit, and VD connection kits for VP 1 / VP 2.

Section-by-Section Technical Breakdown

pp. 29-36
Installation and Connections

Covers mechanical installation of vacuum expansion racks and rack holders, mounting of tilt protection holders, the pump module with pump stand and connection kit, the vacuum connection, inert gas supply connection, and electrical connection. It defines the physical interfaces required before start-up.

  • Vacuum expansion racks and rack holders are installed in the chamber (chap. 4.1).
  • Tilt protection holders are mounted to secure the chamber (chap. 4.2).
  • Pump module with pump stand and connection kit is an accessory (chap. 4.3).
  • Vacuum connection and VD connection kit for VP 1 or VP 2 (accessory) are described (chap. 4.4, 4.4.1).
  • Inert gas supply connection (chap. 4.5) and electrical connection (chap. 4.6) are covered.
pp. 37-48
Functional Overview and Menu Structure of the Controller

Describes the RD4 controller and the optional MB2 controller, including menu structures, access levels, display views, operating modes, and behavior during and after power failures. It is the reference for navigating controller functions.

  • RD4 controller menu structure and access levels (chap. 5.1).
  • MB2 controller normal display, program display, chart-recorder display, icon overview, operating modes, main menu, Settings submenu and Service submenu (chap. 5.2).
  • Performance during and after power failures for RD4 and MB2 controllers (chap. 5.3).
pp. 48-54
Start Up

Covers the condition after establishing power, standby mode on/off, controller settings at start-up, and operation of the vacuum system including evacuation, ventilation, ventilation after power failure, and inert gas operation. It also addresses avoidance of condensation in the interior during the process.

  • Condition after establishing the power connection (chap. 6.1).
  • Standby mode: turning the vacuum drying oven on and off (chap. 6.2).
  • Controller settings upon start-up for RD4 and MB2 (chap. 6.3).
  • Vacuum system: evacuation, ventilation after drying, ventilation/breaking vacuum in case of power failure, operation with inert gas (chap. 6.4).
  • Avoidance of condensation in the interior during the ongoing process (chap. 6.5).
pp. 55-57
Set-Point Entry

Describes how to enter temperature and pressure set-points on the RD4 controller and on the MB2 controller, including entry through the Setpoints menu and direct setpoint entry via the Normal display.

  • RD4 controller: temperature set-point entry and pressure set-point entry (chap. 7.1).
  • MB2 controller: set-point entry through the Setpoints menu and direct setpoint entry via Normal display (chap. 7.2).
pp. 57-69
Setting Special Controller Functions

Covers controller functions such as Functions on/off, Quick Access, Control on/off, turning off the vacuum pump via the pump module socket, using the optional GAS/AIR 2 universal connection for ventilation, closing all valves, activating/deactivating temperature and pressure control, drying monitoring, and cyclical rinsing.

  • Functions on/off and Quick Access menus for RD4; Functions on/off and Control on/off menus for MB2 (chap. 8.1).
  • Turning off the vacuum pump via the socket on the pump module (chap. 8.2).
  • Using the optional universal connection GAS/AIR 2 for ventilation (chap. 8.3).
  • Close all valves function (chap. 8.4).
  • Activating/deactivating temperature control and pressure control (chap. 8.5, 8.6).
  • Drying monitoring and cyclical rinsing (chap. 8.7, 8.8).
pp. 70-81
Authorization Levels and Password Protection

Describes password request, assignment and modification of User and Admin passwords on the RD4 controller, and user management, authorization levels, login/logout, user change, password assignment/change, deletion, and activation code on the MB2 controller.

  • RD4 controller: password request, assign/modify User password and Admin password (chap. 9.1).
  • MB2 controller: user management, authorization levels and password protection, log in/out, user change, password assignment and change, activation code (chap. 9.2).
pp. 82-91
General Controller Settings and Information

Covers controller language selection, temperature unit, date and time, chamber address, display brightness, display of inner chamber and heating temperature, display configuration, touchscreen calibration, event list, service contact page, current operating parameters, and technical chamber information.

  • RD4 controller: menu language, temperature unit, date, time, language selection at restart, chamber address, display brightness, display of inner chamber and heating temperature (chap. 10.1).
  • MB2 controller: menu language, date and time, temperature unit, display configuration, touchscreen calibration, event list, service contact page, current operating parameters, technical chamber information (chap. 10.2).
pp. 92-96
Temperature Safety Devices

Describes the over temperature protective device class 2 (thermal switch) and the overtemperature safety controller class 2, including safety controller mode, setting the safety controller on RD4 and MB2, messages and measures in the alarm state, and function check.

  • Over temperature protective device class 2 (thermal switch) (chap. 11.1).
  • Overtemperature safety controller class 2: safety controller mode, setting the safety controller on RD4 and MB2 (chap. 11.2).
  • Message and measures in the state of alarm for RD4 and MB2 (chap. 11.2.3).
  • Function check (chap. 11.2.4).
pp. 96-98
Tolerance Range Settings

Covers setting delay times and tolerance ranges for temperature and pressure alarms on the RD4 controller, and setting alarm delay times and tolerance ranges on the MB2 controller, including alarm condition.

  • RD4 controller: delay time for temperature tolerance range alarm, temperature tolerance range, delay time for pressure tolerance range alarm, pressure tolerance range, state of alarm (chap. 12.1).
  • MB2 controller: setting alarm delay times and tolerance ranges, alarm condition (chap. 12.2).
pp. 99-104
Notification and Alarm Functions

Describes information messages and alarm messages on RD4 and MB2 controllers, resetting an alarm, and activating/deactivating the audible alarm (buzzer).

  • Information messages for RD4 and MB2 (chap. 13.1).
  • Alarm messages for RD4 and MB2 (chap. 13.2).
  • Resetting an alarm (chap. 13.3).
  • Activating/deactivating the audible alarm (buzzer) (chap. 13.4).
pp. 104-106
RD4 Controller: Program Start / Stop Function

Covers starting a time program, starting a week program, stopping a program, and performance after power failures on the RD4 controller.

  • Start time program and start week program (chap. 14.1, 14.2).
  • Stop program (chap. 14.3).
  • Performance after power failures (chap. 14.4).
pp. 106-107
MB2 Controller: Timer Program (Stopwatch Function)

Covers starting a timer program, performance during program delay time, cancelling a running timer program, and performance after the end of the program on the MB2 controller.

  • Starting a timer program (chap. 15.1).
  • Performance during program delay time (chap. 15.1.1).
  • Cancelling a running timer program (chap. 15.2).
  • Performance after the end of the program (chap. 15.3).
pp. 108-120
MB2 Controller: Time Programs

Covers starting, stopping, pausing and cancelling time programs, creating new time programs, program editor and section editor management, value entry for program sections including duration, set-point ramp/step, special controller functions, setpoint, tolerance range, repeating sections, and saving the time program.

  • Starting an existing time program, performance during program delay time, stopping/pausing/cancelling a running time program, performance after the end of the program (chap. 16.1-16.3).
  • Creating a new time program and program editor management (chap. 16.4, 16.5).
  • Section editor: add, copy/insert/replace, delete program sections (chap. 16.6).
  • Value entry for a program section: section duration, set-point ramp and step, special controller functions, setpoint entry, tolerance range, repeating sections, saving the time program (chap. 16.7).
pp. 121-129
MB2 Controller: Week Programs

Covers starting and cancelling week programs, creating new week programs, program editor and section editor management, and value entry for a program section in the Section view including set-point ramp/step modes, weekday, start time, setpoint, and special controller functions.

  • Starting an existing week program and cancelling a running week program (chap. 17.1, 17.2).
  • Creating a new week program and program editor management (chap. 17.3, 17.4).
  • Section editor: add, copy/insert/replace, delete program sections (chap. 17.5).
  • Value entry for a program section: set-point ramp and step modes, weekday, start time, setpoint entry, special controller functions (chap. 17.6).
pp. 129-136
Network and Communication

Covers Ethernet network settings on the RD4 controller including showing MAC address, IP address, subnet mask, standard gateway, DNS server address and DNS chamber name, and changing network configuration. Also covers MB2 Ethernet configuration, MAC address display, and E-Mail settings.

  • RD4 controller: showing network settings and changing configuration (chap. 18.1).
  • MB2 controller: Ethernet configuration, display of MAC address, E-Mail (chap. 18.2).
pp. 137-138
RD4 Controller: Data Recorder

Describes recorded data, storage capacity, setting the storage rate for the DL1 recorder data, and deleting the data recorder on the RD4 controller.

  • Recorded data and storage capacity (chap. 19.1, 19.2).
  • Setting the storage rate for the DL1 recorder data (chap. 19.3).
  • Deleting the data recorder (chap. 19.4).
pp. 139-146
USB Menu: Data Transfer via USB Interface

Covers connecting a USB stick, import and export functions, ongoing data transfer, and removing the USB stick on the RD4 controller.

  • Connecting the USB stick (chap. 20.1.1).
  • Import function and export functions (chap. 20.1.2, 20.1.3).
  • Ongoing data transfer and removing the USB stick (chap. 20.1.4, 20.1.6).
pp. 147-149
Options and Accessories

Lists options and accessories including APT-COM 4 Multi Management software, analog outputs for temperature and pressure, measuring access port for vacuum (9 poles), object temperature display with flexible Pt 100 temperature sensor including connection and display on RD4/MB2, and LED interior lighting.

  • APT-COM 4 Multi Management software (accessory) (chap. 23.1).
  • Analog outputs for temperature and pressure (option) (chap. 23.2).
  • Measuring access port for vacuum, 9 poles (option) (chap. 23.3).
  • Object temperature display with flexible Pt 100 temperature sensor: connection and display on RD4/MB2 (chap. 23.4).
  • LED interior lighting (chap. 23.5).
pp. 150-152
Cleaning and Decontamination

Covers cleaning and decontamination / chemical disinfection procedures for the chamber.

  • Cleaning (chap. 24.1).
  • Decontamination / chemical disinfection (chap. 24.2).
pp. 153-157
Maintenance and Service, Troubleshooting, Repair, Testing

Covers general information and personnel qualification, maintenance intervals and service, service reminder on RD4 and MB2, simple troubleshooting by operating personnel, and sending the chamber back to BINDER GmbH. Includes requirements for annual repeat electrical inspection and maintenance at least once a year.

  • Maintenance intervals and service: regular maintenance at least once a year; replace door gasket only when cold (chap. 25.2).
  • Service Reminder for RD4 and MB2 controllers (chap. 25.3).
  • Simple troubleshooting table covering power supply, heating, safety controller, overtemperature, set-point deviations, and calibration issues (chap. 25.4).
  • Electrical testing: annual repeat inspection and test prior to initial startup and after maintenance/repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702 (chap. 25.1).
pp. 162-169
Technical Description

Covers factory calibration and adjustment, over current protection, VD / VD-UL technical data, equipment/options/accessories extract, accessories and spare parts extract, and dimensions for VD 23, VD 56 and VD 115.

  • Factory calibration and adjustment (chap. 27.1).
  • Over current protection (chap. 27.2).
  • VD / VD-UL technical data (chap. 27.3).
  • Equipment, options, and accessories extract; accessories and spare parts extract (chap. 27.4, 27.5).
  • Dimensions for VD 23, VD 56, VD 115 (chap. 27.6).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
The chamber is available in 230 V versions (VD023-230V Art. 9630-0001, VD056-230V Art. 9630-0002, VD115-230V Art. 9630-0003) and UL 120 V versions (VD023UL-120V Art. 9630-0005, VD056UL-120V Art. 9630-0006, VD115UL-120V Art. 9630-0007). p.1 ↗ Select the correct voltage variant for the site supply; troubleshooting includes checking power supply for voltage of 115 V or 230 V. VD 23, VD 56, VD 115 (all variants)
Standard controller is the RD4 microprocessor program controller; the MB2 microprocessor program controller is optional. p.1 ↗ Controller-specific procedures (menus, passwords, programs, data recorder, USB) differ between RD4 and MB2 and must be followed per the installed controller. Whole series
The chamber is equipped with an Ethernet serial interface and a USB interface; APT-COM 4 Multi Management Software permits networking. p.2 ↗ Network configuration (IP address, subnet mask, gateway, DNS) and USB data transfer are available for data logging and remote management. Whole series
The inner chamber is made of especially corrosion resistant stainless material. p.2 ↗ Suitable for laboratory drying applications; corrosive damage caused by certain ingredients is excluded from BINDER liability. Whole series
Regular maintenance work must be performed at least once a year. p.25 ↗ Plan annual maintenance; warranty becomes void if maintenance work is conducted by non-authorized personnel. Whole series
An annual repeat inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or repair are required for electrical safety; testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. p.25 ↗ Include electrical safety testing in commissioning and annual maintenance schedules. Whole series
The door gasket must be replaced only when cold; otherwise the door gasket may become damaged. p.25 ↗ Allow the chamber to cool before door gasket replacement. Whole series
The chamber must NOT become wet during operation or maintenance work; do NOT remove the rear panel; disconnect the power plug before conducting maintenance work. p.25 ↗ Observe electrical safety precautions during maintenance; only licensed electricians or BINDER-authorized experts may perform maintenance. Whole series
Troubleshooting includes checking the chamber fuse if the chamber is permanently turned off. p.25 ↗ A blown chamber fuse is a first-line check before contacting BINDER Service. Whole series
If the alarm message “Safety controller” appears, the inner chamber temperature has reached the safety controller value; the safety controller value may be set too low or the temperature set-point too high, or there may be an error in the heating system. p.25 ↗ Let the chamber cool down, acknowledge the alarm, and check temperature setpoint and safety controller value (chap. 11.2). Whole series
If the alarm message “Overtemperature” appears, the overtemperature protective device class 2 (thermal switch) has activated due to an error in the heating system. p.25 ↗ Let the chamber cool down, acknowledge the alarm and restart; if the error repeats, turn off the chamber and contact BINDER Service. Whole series
Deviations from the temperature set-point in equilibrated state can be caused by invalid calibration; use only the delivered expansion racks and do NOT change between aluminum and stainless-steel racks. p.25 ↗ Maintain consistent rack material for valid calibration and temperature performance. Whole series
If the reference temperature sensor has insufficient contact to the expansion rack, fix it with thermal conductive paste or adhesive aluminum tape. p.25 ↗ Ensure good thermal contact for accurate reference measurements during calibration. Whole series
Leakage current can occur when using a thermo element that is not electrically isolated; mount the thermo element electrically isolated from the rack. p.25 ↗ Use electrically isolated thermocouples to avoid leakage current during measurements. Whole series
The chamber has IP protection 20 according to standard EN 60529 (as indicated on the type plate). p.13 ↗ Protect the chamber from water ingress; do not operate in wet environments. Whole series
Temperature safety device according to DIN 12880, class 2.0. p.13 ↗ The chamber is equipped with a temperature safety device class 2.0 per DIN 12880. Whole series
UL chambers are certified by Underwriters Laboratories according to ANSI/UL/IEC EN 61010-1, 3rd Edition, 2012-05, Rev. 2 No. 61010-1. p.1 ↗ UL variants are suitable for use in North American markets where UL certification is required. VD 23-UL, VD 56-UL, VD 115-UL
The chamber is delivered with a factory calibration and adjustment (chap. 27.1). p.27 ↗ Initial calibration is performed by the factory; field calibration may be required after maintenance or if deviations occur. Whole series
The chamber has over current protection (chap. 27.2). p.27 ↗ Refer to technical data for over current protection ratings when planning electrical supply. Whole series
The vacuum system supports evacuation, ventilation after drying (flooding with ambient air or inert gas), ventilation/breaking vacuum in case of power failure, and operation with inert gas. p.6 ↗ Plan vacuum pump and inert gas supply connections accordingly; ventilation after power failure is possible. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Chamber permanently turned off. p.155 ↗ No power supply. Check connection to power supply. Chap. 25.4
Chamber permanently turned off. p.155 ↗ Wrong voltage. Check power supply for voltage of 115V or 230V. Chap. 25.4
Chamber permanently turned off. p.155 ↗ Chamber fuse has responded. Check chamber fuse. Chap. 25.4
Chamber permanently turned off. p.155 ↗ Controller defective. Contact BINDER Service. Chap. 25.4
Chamber doesn’t heat up. p.155 ↗ Pt 100 sensor defective. Contact BINDER service. Chap. 25.4
Chamber doesn’t heat up. p.155 ↗ Heating element defective. Contact BINDER service. Chap. 25.4
Chamber doesn’t heat up. p.155 ↗ Semiconductor relay defective. Contact BINDER service. Chap. 25.4
Chamber doesn’t heat up. Alarm message “Safety controller” on the controller display p.155 ↗ Safety controller has responded: Inner chamber temperature has reached the safety controller value. Safety controller value set too low or temperature set-point too high, or error in the heating system. Let the chamber cool down. Acknowledge the alarm on the controller. Check the settings of the temperature setpoint and safety controller value. If appropriate, select suitable safety controller value (chap. 11.2). Chap. 25.4
Chamber doesn’t heat up. Alarm message “Safety controller” on the controller display p.155 ↗ Safety controller defective. Contact BINDER service. Chap. 25.4
Chamber doesn’t heat up. Alarm message “Overtemperature” on the controller display p.155 ↗ Overtemperature protective device class 2 (thermal switch) activated. Error in the heating system. Let the chamber cool down. Acknowledge the alarm on the controller and restart the chamber. If the error repeats, turn off the chamber and contact BINDER service. Chap. 25.4
Set-point temperature is not reached after specified time. p.155 ↗ Chamber door not properly closed. Completely close chamber door. Chap. 25.4
Set-point temperature is not reached after specified time. p.155 ↗ Door gasket defective. Replace door gasket. Chap. 25.4
Set-point temperature is not reached after specified time. p.155 ↗ Controller not adjusted. Calibrate and adjust controller. Chap. 25.4
Chamber heating permanently, set-point not maintained. p.155 ↗ Controller defective. Contact BINDER Service. Chap. 25.4
Chamber heating permanently, set-point not maintained. p.155 ↗ Pt 100 sensor defective. Contact BINDER Service. Chap. 25.4
Chamber heating permanently, set-point not maintained. p.155 ↗ Semiconductor relay defective. Contact BINDER Service. Chap. 25.4
Chamber heating permanently, set-point not maintained. p.155 ↗ Controller not adjusted. Calibrate and adjust controller. Chap. 25.4
Deviations from the indicated heating-up times. p.155 ↗ Oven fully loaded. Charge the oven less or consider longer heating-up times. Chap. 25.4
Deviations from the temperature set-point in equilibrated state. p.155 ↗ Invalid calibration. Use the delivered expansion racks only. Do NOT change between aluminum and stainless-steel racks. Chap. 25.4
Deviations from the temperature set-point in equilibrated state. p.155 ↗ Wrong temperature value measured during calibration. Reference temperature sensor has insufficient contact to expansion rack. Fix the reference temperature sensor with thermal conductive paste or adhesive aluminum tape. Chap. 25.4
Deviations from the temperature set-point in equilibrated state. p.155 ↗ Leakage current when using a thermo element not electrically isolated. Mount a thermo element electrically isolated from the rack. Chap. 25.4