Data Sheet Model VD 115
Technical Reference Summary
Covered Models & Document Scope
BINDER VD 115 vacuum drying chamber for non-flammable solvents. Two standard model versions: VD115-230V (Art.-No. 9630-0003, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug) and VD115UL-120V (Art.-No. 9630-0007, 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Interior volume | 119 L p.2 ↗ | VD115-230V |
| Temperature range | +9 °C above ambient temperature to 220 °C p.2 ↗ | VD115-230V |
| Temperature uniformity at 100°C | 2.9 ± K p.2 ↗ | VD115-230V |
| Temperature fluctuation at 100°C | 0.1 ± K p.2 ↗ | VD115-230V |
| Heating up time to 100°C | 130 min p.2 ↗ | VD115-230V |
| Leak rate | 0,01 bar/h p.2 ↗ | VD115-230V |
| Pressure fluctuation over time at 50 mbar | 1 ± mbar p.2 ↗ | VD115-230V |
| Pressure fluctuation over time at 400 mbar | 70 ± mbar p.2 ↗ | VD115-230V |
| Rated voltage | 200...230 V p.2 ↗ | VD115-230V |
| Rated voltage | 100...120 V p.2 ↗ | VD115UL-120V |
| Power frequency | 50/60 Hz p.2 ↗ | VD115-230V |
| Nominal power | 1,6 kW p.2 ↗ | VD115-230V |
| Unit fuse | 10,0 A p.2 ↗ | VD115-230V |
| Unit fuse | 20,0 A p.2 ↗ | VD115UL-120V |
| Phase (nominal voltage) | 1~ p.2 ↗ | VD115-230V |
| Vacuum connection with small flange | 16 DN mm p.2 ↗ | VD115-230V |
| Measuring access port with small flange | 16 DN mm p.2 ↗ | VD115-230V |
| Net weight of the unit (empty) | 146 kg p.2 ↗ | VD115-230V |
| Load per rack | 20 kg p.2 ↗ | VD115-230V |
| Permitted load | 70 kg p.2 ↗ | VD115-230V |
| Wall clearance back | 100 mm p.2 ↗ | VD115-230V |
| Wall clearance sidewise | 70 mm p.2 ↗ | VD115-230V |
| Width net | 743 mm p.3 ↗ | VD115-230V |
| Height net | 942 mm p.3 ↗ | VD115-230V |
| Depth net | 581 mm p.3 ↗ | VD115-230V |
Operating Limits
- Temperature range: +9 °C above ambient temperature to 220 °C p.2 ↗
- Permitted load: 70 kg p.2 ↗
- Load per rack: 20 kg p.2 ↗
- Wall clearance back: 100 mm p.2 ↗
- Wall clearance sidewise: 70 mm p.2 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-2 p.1 ↗ | VD | Product series: vacuum drying chamber |
| 3-5 p.1 ↗ | 115 | Model size (interior volume 119 L) |
| suffix p.1 ↗ | UL | UL variant with 100...120 V / NEMA 5-15 plug; absent = standard 200...230 V / CEE 7/7 |
| voltage field p.1 ↗ | 230V | Standard model version, 200...230 V 1~ 50/60 Hz |
| voltage field p.1 ↗ | 120V | UL model version, 100...120 V 1~ 50/60 Hz |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| VD115-230V p.1 ↗ | 9630-0003 | Standard, 200...230 V 1~ 50/60 Hz, CEE 7/7 plug | Standard model version | |
| VD115UL-120V p.1 ↗ | 9630-0007 | Standard, 100...120 V 1~ 50/60 Hz, NEMA 5-15 plug | UL model version |