WIC Installation manual 01 2026

Technical Reference Summary

Covered Models & Document Scope

This installation manual (Art. no. 7001-0433, version 01/2026) covers the BINDER Environmental Walk-in Chambers WIC for stability tests and storage in three model versions: WIC 1 (WIC1230V, Art. no. 9800-0011), WIC 2 (WIC2230V, Art. no. 9800-0012) and WIC 3 (WIC3230V, Art. no. 9800-0013). It describes assembly of the chamber and installation of the conditioner (Art. no. 6011-0228), including optional door sill ramp, shelving systems and access port.

Section-by-Section Technical Breakdown

pp. 9-12
Installation site and floor conditions

Defines the required installation site, minimum clearances around the chamber and conditioner, ceiling load restrictions and floor load capacity. It also specifies flatness tolerances for the floor according to DIN 18202.

  • Wall clearance to rear (behind conditioner) 500 mm, side 100 mm; door side 1.5 m.
  • Free distance above device at least 1.5 m; 500 mm sufficient during operation.
  • Minimum ceiling height 2.59 m.
  • Chamber ceiling not suitable for additional loads other than its own weight.
  • Floor must support 0.75 kg/cm3 and be flat, level and stable.
  • Flatness tolerances per DIN 18202 extended table (rows 1-7).
pp. 13
Required tools and accessories

Lists the tools and accessories required for assembly, which are not included in the scope of delivery.

  • Torx and slotted screwdrivers, open-end wrench 13 mm, torque wrench, Allen key 8 mm.
  • Electric blind riveting gun, electric silicone tip, drilling machine, step and ladder.
  • Drill bits: D 2.9, D 3.8, D 12, D 16, D 5-6 with at least 110 mm spiral flute length.
  • Hole saw 62 mm for optional passage.
  • Water hose for fresh water connection and condensate drain with connections and hose clamps.
  • Blind rivet nut setting tool suitable for ⌀8 pull-in nuts.
pp. 16-17
Assembly of floor elements

Covers laying out and locking the floor elements and establishing equipotential bonding between them.

  • Lay out floors according to figures and check horizontal.
  • Push floor elements together and tighten eccentric locks clockwise with a size 8 Allen key.
  • Close eccentric lock holes with self-sealing stainless steel threaded plugs.
  • Equipotential bonding: drill D3 holes in adjacent floor elements, peel off film locally, connect with connecting plates with toothed washer and rivet.
pp. 18-22
Assembly of wall and ceiling elements

Describes the assembly sequence for wall and ceiling elements, including equipotential bonding between base plate, wall panels, wall elements, ceiling elements and between ceiling and wall elements.

  • Start at rear with a wall corner element; start with corner element for conditioner either front left or rear right.
  • Remove protective film from front sides of wall elements by approx. 100-150 mm.
  • Close eccentric locks (hook locks) on inside of each element; tighten clockwise with size 8 Allen key.
  • Leave an opening for ceiling element transport.
  • Equipotential bonding between base plate and wall panels before installing last wall elements.
  • Equipotential bonding between wall elements before installing ceiling elements; use toothed lock washers in addition to connection plates.
  • Equipotential bonding between ceiling elements and between ceiling and wall elements at one point.
  • Silicone all wall butt joints after assembly; fit plastic profiles to floor and ceiling.
pp. 23-25
Door adjustment and sheet metal cover

Covers adjustment of the revolving door and mounting of the sheet metal cover or optional door sill ramp.

  • Remove cover caps when door is open using pin or Allen key.
  • Fine adjustment of contact pressure via closing piston through slotted holes until door seals tightly.
  • Height fine adjustment of door leaf.
  • Sheet metal cover: pre-drill through top layer with Ø 3.2 mm drill bit, recut stainless steel top layer with galvanized screw, screw ramp with 4.2 x 25 mm screws.
  • Door sill ramp option: same pre-drilling and screws DIN 7982 4.2 x 25 mm.
pp. 26-29
Installing the conditioner

Describes checking the silicone gasket, removing conditioner covers, positioning and screwing the conditioner to the chamber, and connecting the water pipes.

  • Check silicone seal on back of conditioner; fix with tube silicone if necessary.
  • Open rear panel by loosening two tension locks on each side; remove sheet metal cover and slide air filter (blue) up and out.
  • Place conditioner at back of chamber, position 4 supplied vibration pads under feet.
  • Screw conditioner to chamber with supplied screws; tightening torque 30 Nm.
  • Water connections: fresh water inlet 1/2" female NPT; waste water connection for condensate drain inlet 1/2" female NPT.
pp. 30-33
Installing air baffles and sensor box

Covers mounting the air baffles in sequence and mounting the sensor box on the first air baffle.

  • Air baffle types: vertical air baffle with sensor box position, vertical air baffle, air baffle for deflection, air baffle for distribution (WIC 1: 1x, WIC 2: 2x, WIC 3: 3x).
  • Pull off protective film in area of screw connections and connection points.
  • Screw first air baffle with stainless steel self-tapping screws (Art. no. 3001-0198).
  • Mount sensor box with self-tapping 6 stainless steel screws on first air baffle.
pp. 34-35
Junction box and distribution boxes

Describes assembling the junction box, cable ducts, distribution boxes and temperature limiter.

  • Attach two cable ducts above and to the right of conditioner; horizontal cable duct length corresponds to conditioner width; remaining approx. 1.5 m for vertical cable duct.
  • Drill 2 holes each with diameter 2.9 mm for each distribution box and temperature limiter.
  • Attach distribution boxes at 200 mm from front edge of chamber and 100 mm from each other.
  • Temperature limiter already connected to door module; screw to ceiling element and wire according to circuit diagram.
  • Temperature limiter wired and screwed to supplied M20 cable gland.
pp. 36-37
Temperature and humidity sensor installation

Covers assembly of the sensor unit and its installation in the chamber ceiling.

  • Sensor connection: plug with cable attached via connecting piece with threaded collar; sensor recesses two at top, one at bottom.
  • Drill 16 mm hole in ceiling near first rear air baffle.
  • Fit M16 cable glands with rubber seals on outside and inside of ceiling panel.
  • Place sensor horizontally and allow cable drip loop; route cable in air ducts using cable ties.
pp. 38-40
Lighting equipment installation

Describes mounting the light with clips, cable entry and motion detector adjustment.

  • Attach two mounting clips to air duct, 100 mm from both ends of lamp.
  • Drill 16 mm hole in ceiling; guide light cable through cover and fasten with cable gland on outside and inside.
  • Light has integrated motion detector; detection range, hold time and brightness threshold adjustable via DIP switches.
  • Attach motion detector near door so cable is on opposite side.
pp. 40-43
Optional shelving system

Lists shelf sets for each WIC size and describes installation and equipotential bonding of shelves.

  • WIC 1 shelf set 8012-2240: height 1.80 m, 6x uprights, minimum 2x cross braces, 5 sets in widths 1000 mm (1x), 1100 mm (1x), 1200 mm (3x).
  • WIC 2 shelf set 8012-2241: height 1.80 m, 8x uprights, minimum 3x cross braces, 7 sets in widths 1000 mm (1x), 1100 mm (1x), 1125 mm (2x), 1200 mm (3x).
  • WIC 3 shelf set 8012-2242: height 1.80 m, 10x uprights, minimum 5x cross braces, 9 sets in widths 1000 mm (1x), 1100 mm (1x), 1125 mm (4x), 1200 mm (3x).
  • Detailed installation instructions included with shelving system.
pp. 46-51
Controller box and electrical connections

Covers attaching the controller box with controller and main power switch / Emergency Off (EMO) switch, and electrical connections of door frame heater, chamber light, conditioner and junction box to customer power supply.

  • Placement and fastening of mounting plate.
  • Attaching controller box and main power switch / emergency Off switch.
  • Installing cover of controller box and main power switch.
  • Connecting chamber door frame heater to junction box.
  • Connecting chamber light to junction box.
  • Connecting conditioner to junction box.
  • Connecting junction box to customer's power supply.
pp. 53
Optional access port

Describes installation of the optional access port.

  • Hole saw 62 mm for optional passage.
  • Silicon access port d50 with plug and sealing ring included in accessory pack.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Minimum wall clearances: 500 mm to rear (behind conditioner), 100 mm to side, 1.5 m on door side. p.9 ↗ Ensure installation site provides these clearances for air circulation and access. WIC 1, WIC 2, WIC 3
Minimum ceiling height: 2.59 m. p.9 ↗ Verify room height before installation. WIC 1, WIC 2, WIC 3
Floor must support 0.75 kg/cm3. p.11 ↗ Check structural load capacity of installation floor. WIC 1, WIC 2, WIC 3
Conditioner screwed to chamber with tightening torque 30 Nm. p.28 ↗ Use torque wrench to ensure proper sealing and mechanical connection. WIC 1, WIC 2, WIC 3
Water connections: fresh water inlet and waste water connection for condensate drain both 1/2" female NPT. p.29 ↗ Provide matching 1/2" NPT fittings for water supply and drain. WIC 1, WIC 2, WIC 3
Air baffle for distribution quantity: WIC 1: 1x, WIC 2: 2x, WIC 3: 3x. p.30 ↗ Select correct number of air baffles based on chamber size. WIC 1, WIC 2, WIC 3
Distribution boxes attached at 200 mm from front edge of chamber and 100 mm from each other. p.35 ↗ Position distribution boxes according to these dimensions. WIC 1, WIC 2, WIC 3
Drill 16 mm hole in ceiling near first rear air baffle for sensor unit. p.36 ↗ Use 16 mm drill bit for sensor cable entry. WIC 1, WIC 2, WIC 3
Mounting clips for lamp fitted 100 mm from both ends of the lamp. p.38 ↗ Position clips symmetrically for secure lamp mounting. WIC 1, WIC 2, WIC 3
Shelf set WIC 1 8012-2240: height 1.80 m, 6x uprights, minimum 2x cross braces, 5 shelf sets in widths 1000 mm (1x), 1100 mm (1x), 1200 mm (3x). p.42 ↗ Order correct shelf set for WIC 1 and plan shelf widths. WIC 1
Shelf set WIC 2 8012-2241: height 1.80 m, 8x uprights, minimum 3x cross braces, 7 shelf sets in widths 1000 mm (1x), 1100 mm (1x), 1125 mm (2x), 1200 mm (3x). p.42 ↗ Order correct shelf set for WIC 2 and plan shelf widths. WIC 2
Shelf set WIC 3 8012-2242: height 1.80 m, 10x uprights, minimum 5x cross braces, 9 shelf sets in widths 1000 mm (1x), 1100 mm (1x), 1125 mm (4x), 1200 mm (3x). p.42 ↗ Order correct shelf set for WIC 3 and plan shelf widths. WIC 3
BINDER WIC accessory pack part numbers: WIC 1 8009-1123, WIC 2 8009-1172, WIC 3 8009-1201. p.14 ↗ Order correct accessory pack for each chamber size. WIC 1, WIC 2, WIC 3
Connection box Art. no. 5001-0055 includes 2 x M25 cable glands for control sensor and 2 x M16 cable glands for lighting equipment. p.14 ↗ Use appropriate cable glands for sensor and lighting cables. WIC 1, WIC 2, WIC 3
Conditioner Art. no. 6011-0228. p.15 ↗ Use this part number when ordering replacement conditioner. WIC 1, WIC 2, WIC 3
Fastening elements basic package: 3001-0198 (53 pcs), 3001-0240 (14 pcs), 3001-0112 (26 pcs); expansion package adds 8 pcs of 3001-0198. p.15 ↗ Verify screw quantities for assembly; expansion package for additional air baffles. WIC 1, WIC 2, WIC 3
Temperature limiter Jumo ATH-70 and Pt 100 temperature sensor included in mounting material. p.14 ↗ Use specified temperature limiter and sensor for safety and control. WIC 1, WIC 2, WIC 3
LED light RZB Planox ECO with motion detector, IP66, 37 W, 4950 lumen, operating temperature -20 to +50 °C. p.14 ↗ Ensure ambient temperature within range and IP66 protection maintained. WIC 1, WIC 2, WIC 3