Agilent Cary 635 FTIR

Contents

  1. Overview pp. 1-3
    Introduces the Cary 635 FTIR and its versatile sampling capabilities
  2. Config pp. 4-5
    Shows how to configure the system and lists available components
  3. Spec p. 6
    Describes the DialPath sampling module and its specifications
  4. Spec p. 7
    Covers ATR sampling modules and their features
  5. Spec p. 8
    Describes reflectance sampling modules and related specifications
  6. Spec p. 9
    Details the 45-degree specular reflectance module
  7. p. 10
    Provides additional information and contact details
  8. p. 6 table, 3 rows
  9. p. 8 table, 3 rows
  10. p. 9 table, 1 rows

DialPath Modules

ModuleWavelength RangeEffective PathlengthSample Type
DialPath Modules5,100 to 600 cm–130, 50, 100 µm (± 0.25 µm repeatability)Liquids (varying viscosities and volatilities), thin polymer films (< 50 µm thick)
50, 100, 200 µm (± 0.25 µm repeatability)
Custom up to 1,000 µm

ATR Module

Multireflection ZnSe ATR Module
Wavelength Range5,100 to 600 cm–1
Effective Pathlength5.5 μm at 4,000 cm–1 13.0 μm at 1,700 cm–1 36.5 μm at 600 cm–1
Sample TypeLower-concentration components, more dilute solutions; pastes, gels, liquids; not strong acids or bases

45-degree specular reflectance module

ModuleWavelength RangeMeasurement/Sample Type
45-Deg Specular Reflectance ModuleZnSe system: 5,100 to 600 cm-1 KBr system: 7,000 to 350 cm-1Reflectance at angle; mirrors, glass, windows, and thin-film coatings on reflective substrates

Engineering Facts

13 facts · every claim linked to a page
Selection

The ZnSe engine covers most FTIR applications and tolerates humid environments, while the KBr engine gives a broader range but needs humidity control.

Engine optics choice determines spectral range and whether the lab must control humidity.

Source
“ZnSe is inert to moisture, which allows for instrument operation even in humid environments.”
Selection

The KBr engine extends the spectral range down to 350 cm-1 but requires a humidity-controlled environment.

Selecting KBr optics imposes an environmental requirement on the installation site.

Source
“KBr is susceptible to moisture and therefore requires operation in a humidity-controlled environment.”
Installation

Modules attach to the front of the engine and are recognized automatically, so no user optical alignment is needed when swapping them.

Simplifies installation and module changes without alignment procedures.

Source
“Permanently aligned optics require zero user alignment when changing modules.”
Operation

Sampling modules are auto-identified by RFID and all relevant parameters are applied without user intervention.

Reduces operator setup errors and training burden during routine operation.

Source
“The modules are automatically recognized by the Agilent MicroLab software using radio frequency identification (RFID).”
Selection

The ZnSe single-reflection ATR handles only aqueous solutions within pH 5 to 9 and is not for strong acids or bases.

Crystal choice must match sample chemistry to avoid damaging the ATR sensor.

Source
“It can analyze aqueous solutions with pH values ranging from 5 to 9.”
Selection

Diamond ATR suits hard solids, strong acids and bases across all pH ranges.

Diamond is the safe choice when sample chemistry or hardness is aggressive.

Source
“Diamond (Di) is frequently used due to its physical and chemical resistance.”
Selection

Germanium ATR is intended for highly absorbing or highly scattering samples such as carbon black-filled polymers.

Ge's shallow penetration is needed for dark, scattering materials that defeat other crystals.

Source
“It is suitable for analyzing materials that are highly absorbing or have highly scattering components”
Operation

The multireflection ZnSe ATR must not be used with solid samples or a pressure clamp.

Prevents misuse and crystal damage during measurement setup.

Source
“This module is not recommended for use with solid materials; therefore, it is not used with a pressure clamp.”
Selection

DialPath provides fixed-pathlength liquid transmission with three preset pathlengths and 0.25 µm repeatability, avoiding flow cells.

Offers a leak-free alternative to demountable cells for quantitative liquid work.

Source
“30, 50, 100 µm (± 0.25 µm repeatability)”
Operation

The diffuse reflectance module uses a four-position holder whose first position holds a gold mirror for background collection.

Operators must reserve position one for background or results will be invalid.

Source
“The first position of the holder contains a diffuse reflectance gold mirror, which is used for background collection.”
Installation

Specular reflectance accessories slide into the card slide mount holder of the transmission module.

Requires the transmission module as a host, affecting configuration planning.

Source
“Easily slide this accessory into the card slide mount holder in the transmission module of the Cary 635 FTIR spectrometer.”
Maintenance

The light source carries a 5-year warranty while the interferometer and laser diode carry 10-year warranties.

Long component warranties reduce lifecycle cost and downtime risk.

Source
“the spectrometer is backed by a 5-year warranty on the light source, as well as 10-year warranties on the interferometer and the laser diode.”
Performance

A 25 mm optical aperture and short internal path deliver performance comparable to larger lab FTIR systems.

Sets expectations for sensitivity and throughput in a compact footprint.

Source
“A large 25 mm optical aperture and short internal optical path provide performance levels associated with far larger lab FTIR systems.”