Agilent Cary 630 FTIR Spectrometer User's Guide

Contents

  1. Overview pp. 7-10
    Overview of the spectrometer, information symbols, and safety conventions
  2. Overview pp. 11-12
    Initial setup guidance and first-use notes
  3. Procedure pp. 13-16
    Connecting power, power cord selection, and environmental requirements
  4. Procedure p. 17
    Procedure for safely shutting down the instrument
  5. Procedure pp. 18-20
    Diagnostic values and verification of instrument performance
  6. Procedure pp. 21-22
    Collecting sample and background spectra, interchanging accessories
  7. Procedure pp. 23-52
    Operation, cleaning, and background collection for each accessory type
  8. Procedure pp. 53-60
    Replacing windows and source, maintenance cautions and notes
  9. Ordering pp. 61-62
    List of spare parts and consumable items
  10. Spec p. 63
    Instrument technical specifications
  11. Warning pp. 64-66
    Safety warnings and regulatory information
  12. p. 19 table, 5 rows
  13. p. 61 table, 15 rows
  14. p. 17 table, 5 rows
  15. p. 21 table, 10 rows

Table 2 Diagnostic values

ValueOptimal (green)Marginal (yellow)Critical (red)Comment
Energy (center-burst)28,000-15,000>28,000 or <10000> 30,000 or < 3000Indicates the overall alignment and proper gain adjustment for the system.
Source1.9 A>2.2 A or <1.6 A>2.4 A or <1.4 AIndicates a problem with the infrared source control voltage, or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal15,000-3,000>15,000 or <3,000>17,000 or < 2,000Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature38–42<38 or >42<36 or >44Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature10–75>75>80Indicates the ambient temperature over the specified range.

When using the Diffuse Reflectance sampling accessory, the appropriate reflectance disk must be in place to make these measurements. The diffuse gold reference disk should be used with the Diffuse Reflectance sampling accessory. This disk is permanently mounted into Position 1 of the sampling tray.

Table 3 Spare parts

Part numberDescription
G8043-67483Replacement flat tip for ATR
G8043-67484Replacement pellet tip for ATR
G8043-67405Polystyrene film for ATR/Transmission sampling accessory
G8043-67406Polystyrene film for TumblIR/DialPath sampling accessory
PIKE-162-5450NIST traceable polystyrene from PIKE approximately 38 µm (for IQOQ)
925-0128NIST traceable polystyrene from ICL approximately 35 µm (for IQOQ)
G8043-67400Infrared source with purge fitting
G8043-67401Desiccant with purge fitting
G8043-67402KBr window with removal tool
G8043-67403ZnSe window with removal tool
G8043-67003Cary FTIR Swivel Diamond ATR Press
G8043-67004Cary FTIR Swivel Ge-ZnSe Press
G8043-67012Purge kit for engine and Transmission sampling accessory
430-013515VDC 2.66A Power Supply
430-0018FTIR USB Cable

For service contracts, repairs and refurbishment services, contact Agilent.

Table 1 Power switch LED status

StatusLED-colorAction (% duty cycle)
System OFFNo LED illuminationN/A
System OFFRed100%
System startupRed / greenGreen 0.5 s on / red 0.5 s on
System ONGreen100%
Firmware updateRed / greenGreen blinks rapidly twice then red blinks rapidly twice

Analyzing samples

РазделСтраница
Interchanging Sampling Accessories22
Diamond ATR Sampling Accessory23
ZnSe ATR Sampling Accessory27
Germanium ATR Sampling Accessory31
5 Bounce Zinc Selenide ATR Sampling Accessory34
Transmission Sampling Accessory36
10 and 45 Degree Specular Reflectance Sampling Accessories39
TumblIR Sampling Accessory42
DialPath Sampling Accessory45
Diffuse Reflectance Sampling Accessory49

Engineering Facts

14 facts · every claim linked to a page
Installation

The instrument must be allowed to reach room temperature before unpacking to prevent condensation on components.

Prevents condensation damage during initial setup and installation.

Source
“place the shipping case in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature”
Compliance

Shipping the instrument in anything other than the original container voids the warranty.

Determines how the unit must be returned or relocated to preserve warranty coverage.

Source
“Shipping the Cary 630 FTIR in packaging other than the original shipping container will void the warranty.”
Installation

The sampling accessory must be installed for the instrument to function properly.

Confirms the accessory is mandatory, not optional, for operation.

Source
“Attaching the sampling accessory or interface is necessary in order to ensure proper working order of the Cary 630 FTIR.”
Installation

Power cords must be routed away from walkways to avoid strain damage to connectors.

Guides cable routing during installation to prevent apparatus or adapter damage.

Source
“All cords must be kept away from high traffic areas.”
Operation

Leaving the spectrometer powered on is recommended to keep moisture out.

Affects daily operating practice and desiccant/humidity management.

Source
“To help keep moisture out of the Cary 630 FTIR, it is recommended to leave the spectrometer on, even when not in use.”
Operation

The power button must be held for two seconds to switch the system on or off.

Prevents confusion about momentary-switch behavior during operation.

Source
“In order to turn the system on or off, hold down the button for two (2) seconds.”
Performance

A background spectrum should be collected before every sample to avoid environmental drift effects.

Sets the measurement workflow needed for accurate results.

Source
“it is recommended that the system is configured in the software to collect a background spectrum before every sample is analyzed”
Selection

The diamond ATR accepts samples across the full pH range 1 to 14.

Determines accessory choice for strong acids or abrasive samples.

Source
“it can accept samples with a pH range from 1 to 14”
Selection

The ZnSe ATR is limited to samples with pH between 5 and 9 and must not be used with hard or abrasive samples.

Prevents crystal damage and guides accessory selection by sample chemistry.

Source
“Only measure samples with a pH between 5 and 9.”
Selection

The Ge ATR provides shallower penetration, making it suited to highly absorbing carbon-black samples.

Guides accessory selection for highly absorbing or carbon-black-containing samples.

Source
“This is important when measuring samples containing carbon black, such as tires, rubbers and o-rings”
Maintenance

The desiccant should be replaced once humidity reaches 30% or the indicator turns pink.

Defines the maintenance trigger for desiccant replacement to protect optics.

Source
“Changing the desiccant is recommended once the humidity level has reached 30% or the color indicator is pink.”
Maintenance

After opening the unit, allow 30 minutes to 2 hours for warm-up and equilibration before use.

Sets the waiting period after service before reliable measurements.

Source
“can take 30 minutes to 2 hours”
Maintenance

The infrared source has a typical lifetime of three years and may need replacement.

Informs spare-part planning and expected service intervals.

Source
“The typical lifetime of the infrared source is three years.”
Safety

The system is not intrinsically safe and must only be used in atmospheres tested for flammable materials.

Critical for site selection and hazardous-area compliance.

Source
“The Cary 630 FTIR systems are NOT intrinsically safe.”