Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide

Contents

  1. Overview pp. 1-5
    Document purpose, notices, and table of contents
  2. Procedure pp. 6-7
    Guidelines for reliable UV-Visible measurements and PC requirements
  3. Warning pp. 8-9
    Safety notices and fast installation overview
  4. Procedure pp. 10-16
    Installing preconfigured and non-preconfigured systems and software
  5. Spec pp. 17-19
    Environmental, power, and site requirements for the spectrophotometer
  6. Procedure pp. 20-30
    Unpacking, installing, starting, and warming up the instrument
  7. Procedure pp. 31-44
    Cell materials, optical specifications, handling, and sample preparation
  8. Procedure pp. 45-58
    Installing thermostattable cell holder, stirring module, long path cell holder, and filter wheel
  9. Procedure pp. 59-71
    Installing Agilent ChemStation software and configuring computer details
  10. Config pp. 72-74
    Configuring the system and autosamplers
  11. Procedure pp. 75-96
    TCP/IP setup, LAN interface installation, BootP service, and troubleshooting
  12. pp. 97-104
    UV light source safety information and document index
  13. p. 19 table, 3 rows
  14. p. 60 table, 3 rows
  15. p. 56 table, 4 rows
  16. p. 83 table, 4 rows
  17. p. 39 table, 25 rows
  18. p. 51 table, 4 rows
  19. p. 61 table, 2 rows

Table 1 Suitable environmental conditions for the 8454 UV-Vis

ConditionNon-operating (Storage or shipping)Operating within specifications
AltitudeUp to 4,600 m (14,950 ft.)Up to 2,000 m (6,500 ft.)
Temperature- 40 and 70 °C (-4 to 158 °F)5-40 °C (41-104 °F)
Humidity (%RH) non-condensing90%<95%

Table 5 Minimum PC Requirements

PC configurationWindows 7 Professional 64-bit (SP1)Windows 10 Professional 64-bit
Graphics resolutionThe recommended graphics resolutions for video graphics adapter are: 1024 x 768 with large fonts or 1280 x 1024 with large fonts
Processor3 GHz
Screen resolutionSuper VGA (1280 × 1024)

The software will run with the minimum amount of RAM installed, but system performance will depend on what other processes are running.

Table 4 Filter wheel positions

PositionFilter
0None
1UV roll-off
2265 nm UV cut-off (at 50%T)
3295 nm UV cut-off (at 50%T)

Table 6 Example

DeviceAddress
PC10.10.10.1
Spectrophotometer10.10.10.101
Subnet mask255.255.255.0
Gateway10.10.10.1

Table 2 Lower limit of UV transmission for some common solvents

Lower LimitSolvent
180–195 nmSulfuric acid (96%)
Water
Acetonitrile
200–210 nmCyclopentane
n-Hexane
Glycerol
2,2,4-Trimethylpentane
Methanol
210–220 nmn-Butyl alcohol
Isopropyl alcohol
Cyclohexane
Ethyl ether
245–260 nmChloroform
Ethyl acetate
Methyl formate
265–275 nmCarbon tetrachloride
Dimethyl sulfoxide
Dimethyl formamide
Acetic acid
280–290 nmBenzene
Toluene
m-Xylene
Above 300 nmPyridine
Acetone
Carbon disulfide

Table 3 Tubing connections for a cell stirring module

ConnectionComment
A (in), B (out)Without stirring
C (in), D (out)Without cell thermostatting
A (in), B to D, C (out)In series with cell thermostatting
A and C (in), B and D (out)In parallel with cell thermostatting

Table 5 Minimum PC Requirements (continued)

CharacteristicRequirement
Mass storage40 GB Hard-Disk and CD-ROM
MemoryMinimum of 4 GB of RAM for UV-Visible ChemStation software and a single instrument. The minimum recommended amount of memory for a single-bath dissolution system (G1118AA) and kinetic measurements (G1117AA) is 3 GB of RAM. Add 256 MB of RAM for each additional bath in a multi-bath dissolution system (G1118AA). In all cases, an appropriate amount of virtual memory should be configured.

A maximum of four instruments may be interfaced to each Agilent ChemStation. If you are upgrading from a previous version of Agilent ChemStation, you must uninstall the old version of ChemStation first.

Engineering Facts

14 facts · every claim linked to a page
Installation

The spectrophotometer accepts any line voltage in the 100–240 V AC range with no voltage selector, so no manual voltage setting is needed at installation.

Confirms universal input voltage and power budget for site power planning.

Source
“Your spectrophotometer can operate on a line voltage of 100–240 V AC with a line frequency of 50 or 60 Hz. Maximum power consumption is 220 VA.”
Safety

Disconnecting line power requires unplugging the power cord, since the supply still draws power with the front switch off.

Defines the true disconnect device and warns of standby power for safe servicing.

Source
“To disconnect the instrument from line power, disconnect the power cord from the supply. The power supply still uses some power, even if the power switch on the front panel is turned off.”
Installation

The instrument needs 2.5 cm clearance on each side for cooling airflow and about 8 cm at the rear for electrical connections.

Sets minimum bench clearance for thermal management and cable access.

Source
“It needs an additional 2.5 cm (1.0 inches) of space on either side to allow for the circulation of air for cooling, and approximately 8 cm (3.1 inches) in the rear for electrical connections.”
Installation

The waste tubing must run with a continuous downward slope from the instrument to the waste container.

Prevents liquid backing up into the sample area in case of leaks or spills.

Source
“Make sure that the waste tubing has a constant down slope from the spectrophotometer to the waste container.”
Operation

Allow at least 1 hour warm-up for optical stabilization, and the automated self-test can only run after at least one hour of warm-up.

Defines the minimum stabilization time before valid measurements or self-test.

Source
“This self-test can only be executed after the instrument has warmed up for at least one hour.”
Operation

Leaving the instrument powered on extends its life, but the lamps should be switched off when not in use because lamp life decreases with use.

Guides power/lamp management to maximize instrument and lamp lifetime.

Source
“In general, the life of your spectrophotometer will be increased by leaving it on all the time but switching off the lamps when the instrument is not in use.”
Selection

Quartz cells are required for the full 190–1100 nm range, while glass cells suffice for 350–1100 nm work.

Determines the correct cuvette material for the intended wavelength range.

Source
“Quartz sample cells (cuvettes) or sample cells with quartz face plates are required if you want to use the full 190 to 1100 nm wavelength range of your spectrophotometer.”
Maintenance

New cells must be passivated with cleaning/passivating fluid to prevent sticky air bubbles on the windows.

Specifies the passivation step and part number needed before first use of new cells.

Source
“To prevent the formation of sticky bubbles, rinse the cell with cleaning and passivating fluid (part number 5062-8529).”
Operation

For measurements above 400 nm the D2 lamp can be turned off, since the tungsten lamp gives sufficient signal-to-noise over 400–1100 nm.

Reduces photodegradation and extends D2 lamp life for visible-range work.

Source
“For application where readings are taken at wavelengths above 400 nm, the D2-Lamp can be turned off.”
Installation

Installing the filter wheel on the long path cell holder requires a 2.5-mm hex wrench instead of a Phillips screwdriver.

Ensures the correct tool is on hand for accessory installation.

Source
“However, if you install the filter wheel on the long path cell holder, instead of a screwdriver, a 2.5-mm hex wrench is required.”
Compliance

Scheduled processes such as virus checking or defragmenting can abort automated measurements and must be disabled.

Prevents communication faults and aborted runs on the controlling PC.

Source
“Some scheduled processes, like virus checking or defragmenting, may cause instrument communication problems.”
Compliance

The decimal separator must be set to a decimal point to avoid wrong displayed or printed values.

Prevents misformatted numeric results in reports and displays.

Source
“Set the Number Format Decimal Separator to "." (decimal point) in Control Panel > Region and Settings.”
Compliance

A maximum of four instruments may be interfaced to each Agilent ChemStation.

Sets the scaling limit for multi-instrument configurations.

Source
“A maximum of four instruments may be interfaced to each Agilent ChemStation.”
Installation

On an isolated LAN using the default IP 192.168.254.11, BootP is not required.

Simplifies network setup for standalone/isolated installations.

Source
“If the system is on an isolated LAN and the default IP address 192.168.254.11 for the spectrophotometer is used, BootP is not necessary.”