Operating Instructions — VCS 100 vacuum pump (BA 237-EN, ed. 1.1 05/07/2022, EN)

Technical Reference Summary

Covered Models & Document Scope

This document covers the oil-flooded rotary vane vacuum pumps V-VCS 100, V-VCS 150, V-VCX 100 and V-VCX 150. V-VCS is the standard design; V-VCX is the variant with increased water vapour compatibility (more resistant functional parts, glass-fibre reinforced epoxy slide gates, increased gas ballast quantity, higher oil sump temperature).

Section-by-Section Technical Breakdown

pp. 14-16
Product overview and functioning

Describes the single-stage oil-flooded rotary vane design, main components and data plate. Explains the integrated mesh filter, oil mist separator, non-return valve, gas ballast valve and coupling drive. Covers the V-VCX variant features and the adjustable gas ballast valve.

  • Single-stage oil-flooded rotary vane vacuum pump; oil seals gaps, lubricates and dissipates compression heat.
  • Integrated mesh filter on suction side; oil and oil mist separator on outlet side returns oil to circulation.
  • Integrated non-return valve prevents ventilation of the evacuated system for approx. 2 minutes after stop.
  • Standard gas ballast valve prevents condensation of low quantities of water vapour at operating temperature.
  • V-VCX: functional parts more resistant to water steam and solvents; glass-fibre reinforced epoxy slide gates; increased gas ballast quantity; higher oil sump temperature.
  • V-VCX includes an additional adjustable gas ballast valve with ball valve.
pp. 16
Fields of application

Defines the permissible intake pressure ranges for continuous vacuum and evacuation of closed systems. Gives nominal pumping capacities and notes motor selection for continuous operation above 100 mbar (abs.).

  • Nominal pumping capacity at 50 Hz: 100 m³/h (VCS 100) and 150 m³/h (VCS 150).
  • Continuous vacuum intake pressure range: V-VCS 0.5 to 500 mbar (abs.); V-VCX 3.0 to 500 mbar (abs.).
  • For evacuating closed systems, volume to be evacuated must be no more than 2% of nominal pumping capacity.
  • For continuous operation > 100 mbar (abs.), use next higher motor output.
  • Frequent switching (approx. 10 times per hour) or high ambient/inlet temperatures may exceed motor winding and bearing temperature limits.
pp. 17-19
Installation

Covers preparation, mechanical installation, pipe connection, oil check and motor connection. Specifies space requirements, horizontal alignment, cooling air clearances and electrical protection requirements.

  • Provide at least 40 cm space around the machine for maintenance.
  • Machine must be set up horizontally; max. inclination 1° in all directions.
  • Cooling air inlets and outlets must be at least 30 cm from adjacent walls.
  • Ambient temperature: min. +12 °C, max. +40 °C.
  • Exhaust air opening must not be closed or restricted; no filter on exhaust side; back pressure only up to +100 mbar.
  • For parallel operation, install an external non-return valve in the suction line upstream of each pump.
  • Electrical installation per EN 60204 by qualified electrician; motor protection switch and main switch installed by operating company.
  • Motor protection class IP 55, insulation class F; permissible tolerances ±5% voltage and ±2% frequency.
pp. 20-22
Commissioning and decommissioning

Describes start-up checks, rotation direction check, oil level check, standstill measures, decommissioning and recommissioning. Emphasizes installation verification before start-up.

  • Installation check before start-up: no transport damage, safe horizontal standing, correct and leak-tight pipe connections, tight screw/flange connections, compliant electrical installation.
  • Check motor rotation direction against direction of rotation arrow on motor flange.
  • Check oil level through oil sight glass and top up if necessary.
  • For standstill times longer than two minutes, vent the connected suction line to atmospheric pressure.
pp. 23-31
Maintenance and repair

Provides maintenance table and procedures for cleaning, oil maintenance, oil separator elements, fan, motor and coupling. Specifies oil viscosity and maintenance intervals.

  • Lubricating oil viscosity must comply with ISO VG 100 as per DIN ISO 3448.
  • Clean mesh filter in suction flange and filter cartridge of gas ballast valve.
  • Change oil separator elements when dirty.
  • Check fan for dirt and damage; clean ventilation slots.
  • Check coupling and coupling sleeve for wear; replace if defective.
  • Repairs only by authorised workshops using original spare parts.
pp. 33-35
Errors

Contains a table of malfunctions with causes and remedies. Covers motor protection trips, insufficient pumping capacity, failure to reach final pressure, overheating, oil mist in exhaust, unusual noise and water in oil.

  • Never restart pump after automatic shut-off without finding and remedying the cause.
  • Motor protection switch must be suitable for short-term excess current at start-up (IEC 60947-4-1).
  • Oil viscosity must comply with ISO VG 100 as per DIN ISO 3448.
  • Back pressure in exhaust line must be checked if capacity or final pressure problems occur.
  • Water in oil: install water interceptor upstream or let pump run with closed suction side to evaporate water.
pp. 37
Technical data

Reference to chapter 10 for technical data including weight, sound pressure level and electrical data. Data sheet D 237 contains technical data and characteristic curves.

  • Technical data are provided in chapter 10 of the manual.
  • Data sheet D 237 shows dependency of pumping capacity on intake pressure.
  • Sound pressure level measured as per EN ISO 3744.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Permissible ambient and inlet temperature range is 12 to 40 °C. p.8 ↗ Operation outside this range requires consultation with the manufacturer; ensure installation room ventilation and temperature control. V-VCS 100/150 and V-VCX 100/150
Continuous intake pressure range: V-VCS 0.5 to 500 mbar (abs.); V-VCX 3.0 to 500 mbar (abs.). p.16 ↗ Select pump variant based on required process vacuum; operating outside these ranges risks oil leakage through outlet. V-VCS and V-VCX
For evacuating closed systems, the volume to be evacuated must be no more than 2% of the nominal pumping capacity. p.16 ↗ Size the pump so that the connected closed volume does not exceed 2% of the pump's nominal capacity to avoid excessive evacuation time or oil carry-over. V-VCS 100/150 and V-VCX 100/150
For continuous operation > 100 mbar (abs.), the next higher motor output should be used. p.16 ↗ When specifying a pump for continuous duty above 100 mbar, select a larger motor to avoid overheating. V-VCS 100/150 and V-VCX 100/150
Maximum permissible back pressure on the exhaust side is +100 mbar. p.18 ↗ Exhaust piping and any connected equipment must not create back pressure exceeding +100 mbar; no filter may be installed on the exhaust side. V-VCS 100/150 and V-VCX 100/150
Machine must be installed horizontally with maximum inclination of 1° in all directions. p.17 ↗ Ensure the floor is plane and even; secure the machine against tipping over and falling. V-VCS 100/150 and V-VCX 100/150
Cooling air inlets and outlets must be at least 30 cm away from adjacent walls. p.17 ↗ Maintain clearance to prevent recirculation of hot exhaust air and ensure sufficient cooling. V-VCS 100/150 and V-VCX 100/150
Provide at least 40 cm space around the machine for maintenance work. p.17 ↗ Installation layout must allow access to suction flange, oil filling point, oil sight glass, oil drain, gas ballast and de-oiler housing. V-VCS 100/150 and V-VCX 100/150
Motor electrical data: protection class IP 55, insulation class F, designed per DIN EN 60034. p.19 ↗ Verify motor data plate matches mains supply; use appropriate motor protection switch. V-VCS 100/150 and V-VCX 100/150
Permissible voltage deviation ±5% and frequency deviation ±2%. p.19 ↗ Ensure mains supply quality is within these tolerances to avoid loss of power or motor damage. V-VCS 100/150 and V-VCX 100/150
Lubricating oil viscosity must comply with ISO VG 100 as per DIN ISO 3448. p.26 ↗ Use only oil of the specified viscosity grade; incorrect viscosity can cause overheating, insufficient lubrication or poor vacuum. V-VCS 100/150 and V-VCX 100/150
For parallel operation of multiple pumps, an external non-return valve must be installed in the suction line upstream of each pump. p.18 ↗ Prevents a switched-off pump from being damaged by suction capacity of operating pumps. V-VCS 100/150 and V-VCX 100/150
For standstill times longer than two minutes, vent the connected suction line to atmospheric pressure. p.18 ↗ Provide an air release position (e.g. ball valve) in the suction line to avoid possible damage to the machine. V-VCS 100/150 and V-VCX 100/150
Nominal pumping capacity at 50 Hz: 100 m³/h (VCS 100) and 150 m³/h (VCS 150). p.16 ↗ Use these values for sizing; actual capacity depends on intake pressure (see data sheet D 237). V-VCS 100 and V-VCS 150
Storage temperature range: -10 °C to +60 °C; relative air humidity max. 80%; store dust-free, dry, vibration-free and protected from sun radiation. p.13 ↗ For long-term storage (>3 months), use preservation oil instead of operating oil; do not store for more than 6 months. V-VCS 100/150 and V-VCX 100/150
Motor protection switch should have overload-dependent delayed switch off considering short-term excess current at start-up (version with short circuit and overload trigger as per IEC 60947-4-1). p.33 ↗ Select motor protection switch suitable for cold-start excess current to avoid nuisance tripping. V-VCS 100/150 and V-VCX 100/150
An output reduction is noticeable when installed at more than 1000 m above sea level. p.17 ↗ Contact manufacturer for installations at high altitude to account for reduced cooling and motor performance. V-VCS 100/150 and V-VCX 100/150
The non-return valve prevents ventilation of the evacuated system for approx. 2 minutes after the pump has stopped. p.15 ↗ Do not rely on the non-return valve for long-term isolation; use external valves if longer hold times are required. V-VCS 100/150 and V-VCX 100/150

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Machine is switched off by the motor protection switch p.33 ↗ Mains voltage/ Frequency does not correspond with the motor data Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.33 ↗ Connection to motor terminal board is not correct Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.33 ↗ Motor protection switch is not set correctly Check by qualified electrician Chapter 5.5
Machine is switched off by the motor protection switch p.33 ↗ Motor protection switch is triggered too quickly Use a motor protection switch with an overload-dependent delayed switch off that takes into consideration the short term excess current at start up (version with short circuit and overload trigger as per IEC 60947-4-1)
Machine is switched off by the motor protection switch p.33 ↗ Vacuum pump or its oil is too cold Note the ambient temperature and the inlet temperature Chapter 2.4
Machine is switched off by the motor protection switch p.33 ↗ The lubricating oil is too viscous The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 Chapter 7.6
Machine is switched off by the motor protection switch p.33 ↗ The oil separator elements are dirty. Change the oil separator elements Chapter 7.7
Machine is switched off by the motor protection switch p.33 ↗ The back pressure in the exhaust line is too high Check the hose and/or the pipe Chapter 5.3
Machine is switched off by the motor protection switch p.33 ↗ Permanent operation >100 mbar (abs.). Use next largest motor output Chapter 4.4
Pumping capacity is insufficient p.33 ↗ The inlet pipe is too long or too narrow Check the hose and/or the pipe Chapter 5.3
Pumping capacity is insufficient p.33 ↗ Leak on the inlet side of the vacuum pump or in the system Check the pipework and screw connections for leaks and check for tight fit Chapter 7.2
Pumping capacity is insufficient p.33 ↗ The intake filter is dirty Clean or replace the intake filter Chapter 7.5
Final pressure (max. vacuum) is not reached p.34 ↗ Leak on the inlet side of the vacuum pump or in the system Check the pipework and screw connections for leaks and check for tight fit Chapter 7.2
Final pressure (max. vacuum) is not reached p.34 ↗ Incorrect oil viscosity The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 Chapter 7.6
Machine gets too hot p.34 ↗ Ambient or inlet temperatures too high Ensure proper use Chapter 2.4
Machine gets too hot p.34 ↗ Cooling air supply is obstructed Check ambient conditions Chapter 5.1
Machine gets too hot p.34 ↗ Cooling air supply is obstructed Clean ventilation slots Chapter 7.8
Machine gets too hot p.34 ↗ Fan dirty and damaged Clean the fan, replace the fan, if necessary Chapter 7.8
Machine gets too hot p.34 ↗ The lubricating oil is too viscous The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 Chapter 7.6
Machine gets too hot p.34 ↗ The oil separator elements are dirty. Change the oil separator elements Chapter 7.7
Machine gets too hot p.34 ↗ The back pressure in the exhaust line is too high Check the hose and/or the pipe Chapter 5.3
Exhaust air contains visible oil mist p.34 ↗ The oil separator elements are not inserted correctly or the O rings are missing. Check that it is correctly seated Chapter 7.7
Exhaust air contains visible oil mist p.34 ↗ Unsuitable oil is being used Use suitable types Chapter 7.6
Exhaust air contains visible oil mist p.34 ↗ The oil separator elements are dirty Change the oil separator elements Chapter 7.7
Exhaust air contains visible oil mist p.34 ↗ Ambient or inlet temperatures too high Ensure proper use Chapter 2.4
Exhaust air contains visible oil mist p.34 ↗ Cooling air supply is obstructed Check ambient conditions Chapter 5.1
Exhaust air contains visible oil mist p.34 ↗ Cooling air supply is obstructed Clean ventilation slots Chapter 7.2
The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.34 ↗ The pump housing is worn (chatter marks) Repair by manufacturer or authorised workshop Elmo Rietschle Service
The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.34 ↗ The vacuum adjustment valve (if available) is vibrating Replace the valve Chapter 7.11
The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.34 ↗ The Blades are damaged Repair by manufacturer or authorised workshop Elmo Rietschle Service
The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.34 ↗ Vacuum pump or its oil is too cold Note the ambient temperature and the inlet temperature Chapter 2.4
The machine makes a strange noise (The blades making a hammering noise when starting from cold is normal if it disappears within two minutes as the operating te p.34 ↗ The lubricating oil is too viscous The oil viscosity must comply with ISO VG 100 as per DIN ISO 3448 Chapter 7.6
The machine makes a strange noise p.34 ↗ Coupling defective or coupling sleeve worn out Check and replace coupling / coupling sleeve Chapter 7.9.2
Water in lubricating oil p.35 ↗ Pump sucks in water Install water interceptor upstream of the pump —
Water in lubricating oil p.35 ↗ The pump sucks in more water vapour than is suitable for its water vapour compatibility Contact the manufacturer for increased gas ballast —
Water in lubricating oil p.35 ↗ Pump only works for a short time and therefore does not reach its normal operating temperature Let the pump continue to run with a closed suction side after extracting the water vapour until the water has evaporated from the oil —