Shakers Flyer Innova S44i New Brunswick Innova 44 44R Stackable Incuba
Technical Reference Summary
Covered Models & Document Scope
Comparison leaflet of two stackable incubator shakers: Eppendorf Innova® S44i (flagship, incubated and refrigerated variants) vs. New Brunswick™ Innova® 44/44R (economical workhorse). Covers vessel capacity, temperature range, speed range, orbit options, platform, drive, dimensions, weight, user interface, programming, lighting and contamination options.
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Vessel capacity, 125 mL Erlenmeyer (universal/dedicated platform) | 81/86 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 125 mL Erlenmeyer (universal/dedicated platform) | 39/60 vessels p.2 ↗ | Innova 44/44R |
| Vessel capacity, 250 mL Erlenmeyer (universal/dedicated platform) | 49/53 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 250 mL Erlenmeyer (universal/dedicated platform) | 30/40 vessels p.2 ↗ | Innova 44/44R |
| Vessel capacity, 500 mL Erlenmeyer (universal/dedicated platform) | 36/39 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 500 mL Erlenmeyer (universal/dedicated platform) | 24/24 vessels p.2 ↗ | Innova 44/44R |
| Vessel capacity, 1 L Erlenmeyer (universal/dedicated platform) | 20/23 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 1 L Erlenmeyer (universal/dedicated platform) | 14/15 vessels p.2 ↗ | Innova 44/44R |
| Vessel capacity, 2 L Erlenmeyer (universal/dedicated platform) | 13/15 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 2 L Erlenmeyer (universal/dedicated platform) | 8/12 vessels p.2 ↗ | Innova 44/44R |
| Vessel capacity, 2.5 L Ultra Yield™ (universal/dedicated platform) | 13/15 vessels p.2 ↗ | Innova S44i |
| Vessel capacity, 2.5 L Ultra Yield™ (universal/dedicated platform) | 8/- vessels p.2 ↗ | Innova 44/44R |
| Maximum flask size | 5 L p.2 ↗ | |
| Temperature range, incubated | 10 °C above ambient to 80 °C p.2 ↗ | |
| Temperature range, refrigerated | 20 °C below ambient (min. 4 °C) to 80 °C p.2 ↗ | |
| Temperature uniformity | ±0.25 °C p.2 ↗ | Innova S44i, at 37 °C and 250 rpm |
| Temperature uniformity | ±0.25 °C p.2 ↗ | Innova 44/44R, at 37 °C |
| Speed range | 20 – 400 rpm p.2 ↗ | Innova S44i, 2.5 cm orbit |
| Speed range | 20 – 300 rpm p.2 ↗ | Innova S44i, 5.1 cm orbit |
| Speed range | 25 – 400 rpm p.2 ↗ | Innova 44/44R, 2.5 cm orbit |
| Speed range | 25 – 300 rpm p.2 ↗ | Innova 44/44R, 5.1 cm orbit or 2 stacked units |
| Speed range | 25 – 250 rpm p.2 ↗ | Innova 44/44R, 3 stacked units |
| Platform size | 76 × 62 (30 × 24 in) cm p.2 ↗ | Innova S44i |
| Platform size | 76 × 46 (30 × 18 in) cm p.2 ↗ | Innova 44/44R |
| External height (with lid/door open) | 114.1 (44.9 in) excluding base cm p.2 ↗ | Innova S44i |
Operating Limits
- Temperature range, incubated: 10 °C above ambient to 80 °C p.2 ↗
- Temperature range, refrigerated: 20 °C below ambient (min. 4 °C) to 80 °C p.2 ↗
- Maximum flask size: 5 L p.2 ↗
- Stackability: up to 3 units p.2 ↗
- Max speed, Innova 44/44R, 3 stacked units: 250 rpm p.2 ↗