Agilent Two-Color Microarray-Based Gene Expression Analysis (Low Input Quick Amp Labeling) Protocol

Contents

  1. Overview pp. 2-6
    Legal notices, what's new in version 2.0, and table of contents
  2. Overview pp. 7-9
    Procedural notes, safety notes, and general precautions
  3. Spec pp. 10-14
    Required equipment, reagents, hardware, and software for the protocol
  4. Procedure pp. 15-29
    Spike mix preparation, labeling reaction, purification, and cRNA quantification
  5. Procedure pp. 30-35
    Hybridization sample preparation, assembly, and hybridization steps
  6. Procedure pp. 36-42
    Wash buffer preparation and microarray slide washing procedure
  7. Procedure pp. 43-49
    Scanning slides and extracting data with Agilent Feature Extraction software
  8. Procedure pp. 50-50
    Two-color exon analysis low input quick amp WT labeling protocol
  9. Procedure pp. 51-68
    RNA purification, thermocycler protocols, quality assessment, and ozone prevention
  10. pp. 69-81
    Kit contents, supplemental user guides, microarray handling tips, and related reagents
  11. p. 11 table, 7 rows
  12. p. 10 table, 4 rows
  13. p. 12 table, 19 rows
  14. p. 13 table, 9 rows
  15. p. 59 table, 4 rows
  16. p. 64 table, 3 rows
  17. p. 81 table, 3 rows
  18. p. 19 table, 3 rows
  19. p. 23 table, 3 rows
  20. p. 24 table, 5 rows

Table 5 Required Equipment

DescriptionVendor and part number
Agilent Microarray ScannerAgilent p/n G4900DA, G2565CA or G2565BA
Hybridization Chamber, stainlessAgilent p/n G2534A
Hybridization gasket slides 1 microarray/slide, 5 slides/boxAgilent p/n G2534-60003
Hybridization gasket slides 2 microarrays/slide, 5 slides/boxAgilent p/n G2534-60002
Hybridization gasket slides 4 microarrays/slide, 5 slides/boxAgilent p/n G2534-60011
Hybridization gasket slides 8 microarrays/slide, 5 slides/boxAgilent p/n G2534-60014
Go to www.agilent.com/genomics to see all available kit configurations.

Table 1 Catalog SurePrint G3 Microarrays - Human

Part NumberDescription
G4832AG3 Human Exon 4×180K Microarray Kit (3 slides)
G4857A-028679G3 Human Exon 4×180K Microarray (1 slide)
G4848AG3 Human Exon 2×400K Microarray Kit (3 slides)
G4856A-028680G3 Human Exon 2×400K Microarray (1 slide)

Table 5 Required Equipment (continued)

DescriptionVendor and part number
Hybridization oven; temperature set at 65°CAgilent p/n G2545A
Hybridization oven rotator for Agilent Microarray Hybridization ChambersAgilent p/n G2530-60029
nuclease-free 1.5 mL microfuge tubeAmbion p/n 12400 or equivalent
magnetic stir bar (×2)Corning p/n 401435 or equivalent
magnetic stir plate (×2)Corning p/n 6795-410 or equivalent
circulating water baths or heat blocks set to 37°C, 40°C, 60°C, 65°C, 70°C, and 80°C,Corning p/n 6795-420 or equivalent
microcentrifugeEppendorf p/n 5417R or equivalent
sterile storage bottleNalgene 455-1000 or equivalent
spectrophotometerNanoDrop p/n ND-1000 UV-VIS or equivalent
micropipettorPipetman P-10, P-20, P-200, P-1000 or equivalent
slide-staining dish, with slide rack (×3)Thermo Shandon p/n 121 or equivalent
vacuum concentratorThermo Scientific p/n DNA 120-115 or equivalent
clean forceps
ice bucket
powder-free gloves
sterile, nuclease-free aerosol barrier pipette tips
vortex mixer
timer
nitrogen purge box for slide storage

Table 6 Required Reagents

DescriptionVendor and part or catalog number
Low Input Quick Amp WT Labeling Kit, Two-ColorAgilent p/n 5190-2944
RNA Spike-In Kit, Two-ColorAgilent p/n 5188-5279
Gene Expression Hybridization KitAgilent p/n 5188-5242
Gene Expression Wash Buffer KitAgilent p/n 5188-5327
DNase/RNase-free distilled waterInvitrogen p/n 10977-015
RNeasy Mini KitQiagen p/n 74104
ethanol (95% to 100% molecular biology grade)Sigma-Aldrich p/n E7023-6×500ML
Milli-Q water or equivalent
isopropyl alcohol (molecular biology grade)

Table 21 Analyzer and Kits

DescriptionVendor and part number
2100 BioanalyzerAgilent p/n G2938C or G2939A
RNA 6000 Nano LabChip KitAgilent p/n 5067-1511
RNA 6000 Pico LabChip KitAgilent p/n 5067-1513
SpectrophotometerNanoDrop p/n ND-1000 or equivalent

Table 23

Ozone level in your labWash ProcedureOzone-Barrier Slide Cover
< 5 ppbWash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 39No
> 5 ppb and < 10 ppbWash Procedure without Stabilization and Drying Solution. "Step 4. Wash the microarray slides" on page 39Yes
> 10 ppbWash Procedure with Stabilization and Drying Solution. See "Step 1. Prepare the Stabilization and Drying Solution" on page 65 and "Step 2. Wash with Stabilization and Drying Solution" on page 66.Yes

Related Microarray Reagents

DescriptionVendor and part number
Universal Human Reference RNAAgilent p/n 740000
Universal Mouse Reference RNAAgilent p/n 740100
Universal Rat Reference RNAAgilent p/n 740200

Table 10 Dilutions of Spike A Mix for cyanine 3 and Spike B Mix for cyanine 5 labeling

Starting Amount of RNA Total RNA (ng)Serial Dilution FirstSerial Dilution SecondSerial Dilution ThirdSerial Dilution FourthSpike Mix Volume to be used in each labeling reaction (µL)
251:201:401:161:82
501:201:401:161:42
1001:201:401:161:22

Table 11 WT Primer MasterMix

ComponentVolume (µL) per reactionVolume (µL) per 5 reactionVolume (µL) per 10 reactions
WT Primer1510
Diluted Spike A Mix and Spike B Mix21020
Total Volume31530

Table 12 cDNA Master Mix

ComponentVolume (µL) per reactionVolume (µL) per 5 reactionVolume (µL) per 10 reactions
5× First Strand Buffer (green cap)21020
0.1 M DTT (white cap)1510
10 mM dNTP Mix (green cap)0.52.55
Affinity Script RNase Block Mix (violet cap)1.2612
Total Volume4.723.547

Engineering Facts

14 facts · every claim linked to a page
Selection

The protocol requires the Low Input Quick Amp WT Labeling Kit, Two-Color (p/n 5190-2944), not the standard Low Input Quick Amp Labeling kit.

Using the wrong labeling kit will invalidate the workflow and Agilent will not guarantee performance.

Source
“Low Input Quick Amp WT Labeling Kit, Two-Color”
Selection

Agilent Exon microarrays are only available on SurePrint G3 formats, so a SureScan or Agilent C scanner is mandatory.

Scanner selection must be confirmed before committing to the Exon microarray platform.

Source
“A SureScan or Agilent C microarray scanner is required for SurePrint G3 formats.”
Selection

The starting total RNA input is 25–100 ng, with 50 ng recommended for 2-, 4-, and 8-pack arrays and 100 ng for 1-pack arrays.

Sample input capacity determines whether low-input samples are compatible with the chosen array format.

Source
“The starting input for the Low Input Quick Amp WT Labeling Kit, Two-Color ranges from 25 ng to 100 ng of total RNA.”
Installation

After opening the microarray pouch foil, slides must be stored at room temperature in the dark under vacuum desiccation or nitrogen purge, never in open air.

Improper storage after opening causes slide degradation and invalidates experiments.

Source
“Do not store microarray slides in open air after breaking foil.”
Installation

All unused wells in the SureHyb gasket slide must be filled with 1× Hybridization Buffer because empty wells cause hybridization failure.

Filling unused wells is a mandatory assembly step to prevent run failure.

Source
“Make sure all wells contain sample or 1× Hybridization Buffer. Empty wells can cause failure in hybridization.”
Operation

The array slide must be placed active side down with the Agilent barcode facing down and the numeric barcode facing up, and must not be dropped onto the gasket.

Correct orientation and gentle placement prevent cross-contamination between gasket wells.

Source
“Do not drop the array slide onto the gasket. Doing so increases the chances of samples mixing between gasket wells.”
Operation

Gene Expression Wash Buffer 2 must be prewarmed to 37°C the night before washing arrays.

Wash Buffer 2 temperature directly affects wash stringency and signal quality.

Source
“Tightly cap the sterile storage bottle and put in a 37°C water bath the night before washing arrays.”
Operation

The array-gasket sandwich must be pried open from the barcode end only, and only the barcode portion or edges of the microarray slide may be touched.

Opening from the wrong end or touching the array area damages features and contaminates the slide.

Source
“pry the sandwich open from the barcode end only”
Operation

For uniform washing, no more than eight microarray slides should be processed per disassembly/wash group.

Exceeding eight slides per group causes non-uniform washing and inconsistent data.

Source
“For uniform washing, do up to a maximum of eight disassembly procedures yielding eight microarray slides.”
Maintenance

Staining dishes, racks, and stir bars must be solvent-washed with acetonitrile (if exposed to Stabilization and Drying Solution) or isopropyl alcohol (if not) to avoid wash artifacts.

Residual contaminants on glassware produce image artifacts that corrupt feature extraction.

Source
“Wash staining dishes, racks and stir bars with acetonitrile or isopropyl alcohol to avoid wash artifacts on your slides and images.”
Performance

The two-color platform is robust only at ozone levels of 5 ppb or less; above 10 ppb the Stabilization and Drying Solution wash procedure is required.

Lab ozone level dictates which wash procedure and ozone-barrier accessories must be used.

Source
“The Agilent two-color platform is robust in environments where the ozone level is 5 ppb (approximately 10 μg/m3) or less.”
Performance

The recommended specific activity for hybridization is 15 pmol Cy3 per μg cRNA.

Specific activity is a key QC acceptance criterion for labeled cRNA before hybridization.

Source
“The recommended Specific Activity for hybridization is increased to 15 pmol Cy3 per μg cRNA.”
Compliance

Agilent does not guarantee microarray performance or provide technical support when non-Agilent protocols are used to process Agilent microarrays.

Deviating from the protocol voids Agilent performance guarantees and support.

Source
“Agilent cannot guarantee microarray performance and does not provide technical support to those who use non-Agilent protocols in processing Agilent microarrays.”
Compliance

Agilent guarantees data quality only when Agilent microarrays are scanned on Agilent scanners.

Using a non-Agilent scanner removes the data quality guarantee and requires compatibility verification.

Source
“Agilent can guarantee the quality of data only if the data comes from Agilent microarrays scanned on Agilent scanners.”