Contents

  1. Overview p. 1
    Product identification and general description of the CC-108A unit
  2. Spec p. 2
    Technical data per DIN 12876 including temperature and performance ratings
  3. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range25...100 °C
with water cooling20...100 °C
with refrigerator15...100 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Absolute accuracysetup for calibration
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Heating power at 200V1,5 kW
Pressure pump max. delivery27 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)22 l/min
max. delivery pressure (suction)0,4 bar
Pump connenction (optional)M16x1 male
Bath volume6 l
min. filling capacity4 l
Width bath opening WxD/ bath depth130x210/ 150 mm
Overall dimensions WxDxH **147x407x330 mm
Net weight6 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current10 A
min. Fuse10A
max. Fuse16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method: 25–100 °C without cooling, 20–100 °C with water cooling, and 15–100 °C with a refrigerator.

Determines whether the bath can reach the required low temperatures and which cooling option must be specified.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A and a fuse rated between 10 A and 16 A.

Electrical supply and branch-circuit protection must be sized accordingly before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are ±10 % only if the frequency deviation does not run in the opposite direction, so a −10 % voltage with +3 % frequency is not allowed.

Site power quality must be checked; combined out-of-tolerance voltage and frequency can invalidate the output data.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Compliance

External branch circuit protection according to UL 489 is required.

The installation must include a UL 489-listed branch circuit protector to meet the manufacturer's requirement.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified when ordering.

The correct plug type must be stated on the order to avoid a mismatch with the site socket.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The bath has a 6 l volume with a minimum filling capacity of 4 l.

The minimum fill level must be respected for safe pump operation and temperature control.

Source
“Bath volume 6 l min. filling capacity 4 l”
Operation

The pressure pump delivers up to 27 l/min at 0.7 bar, and the suction pump delivers up to 22 l/min at 0.4 bar.

Pump performance determines whether the external circuit can achieve the required flow and pressure drop.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Safety

The unit has adjustable overtemperature protection according to DIN 12876.

Overtemperature protection must be set correctly to protect the bath and external circuit from overheating.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The bath is made of polycarbonate and is limited to a maximum temperature of 100 °C.

The bath material limits the maximum operating temperature and must not be exceeded.

Source
“transparent baths made of polycarbonate with temperature range up to max. 100°C”
Operation

Polyglycol must not be used as a heat transfer fluid when a Huber pump adapter is used.

Using the wrong heat transfer fluid with the pump adapter can damage the system or cause a safety hazard.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Installation

The unit has degree of protection IP20 and must be operated at ambient temperatures between 5 °C and 40 °C.

The installation location must be dry and within the ambient temperature range; IP20 offers no water protection.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Space requirements around the unit must be respected according to the operating conditions.

Adequate clearance is needed for ventilation and access, and must be verified before siting the unit.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Compliance

The unit is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”