Peter Huber Kältemaschinenbau CC-108A
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-108A

Peter Huber Kältemaschinenbau CC-108A — Thermostatic baths.

The CC-108A is a 6-litre polycarbonate-bath immersion circulator with a variable-speed pressure/suction pump and Pilot ONE controller, rated 25–100 °C without cooling, 20–100 °C with water cooling and 15–100 °C with a refrigerator. It sits between the CC-106A and CC-110A in bath volume within the series. It serves as a temperature-control source for external circuits and open baths up to 100 °C.

Direct drop-in replacements

Eppendorf
MixMate ®
53%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-108A Current Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
CC-118A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Bath volume 6 l
Filling volume 4 l (min.)
Maximum delivery flow rate 27 l/min (pressure pump; also 0, 7 bar)
suction pump; also 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 147 x 407 x 330 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 130 x 210 / 150
Dimensions, Weight & Mechanical Design
Weight 6 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch circuit must be UL 489 protected and fused 10–16 A

The 200–240 V 1~/2~ supply draws up to 10 A; the manufacturer requires external branch circuit protection according to UL 489 and a fuse between 10 A and 16 A. Undersized protection will trip at run-up, oversized protection leaves the unit unprotected.

CC-108A Immersion Circulator Technical Data, page 2
Electrical
Voltage and frequency tolerances must not oppose each other

A ±10 % voltage tolerance is permitted only if the frequency deviation does not run in the opposite direction; a −10 % voltage combined with +3 % frequency is not allowed. Verify site power quality before commissioning, otherwise output data are not guaranteed.

CC-108A Immersion Circulator Technical Data, page 2
Process Limits
Polycarbonate bath caps operation at 100 °C

The transparent polycarbonate bath is limited to a maximum of 100 °C; exceeding it risks deformation and loss of containment. The adjustable overtemperature protection to DIN 12876 must be set below this limit for the bath and external circuit.

CC-108A Immersion Circulator Technical Data, page 1
Process Limits
Minimum 4 l fill required for safe pump operation

With a 6 l bath volume and 4 l minimum filling capacity, the bath must not be run below 4 l; low level risks pump cavitation and poor temperature control. Account for the volume held in external tubing and the circuit when filling.

CC-108A Immersion Circulator Technical Data, page 1
Cooling & Media
Low-temperature reach depends on the cooling option fitted

Without cooling the unit operates 25–100 °C; water cooling extends the low end to 20 °C and a refrigerator to 15 °C. Specify the cooling method before sizing the external circuit, since the achievable setpoint changes with it.

CC-108A Immersion Circulator Technical Data, page 1
Cooling & Media
No polyglycol with a Huber pump adapter fitted

When a Huber pump adapter is used, polyglycol is not permissible as heat-transfer fluid; using it can damage the pump circuit or create a safety hazard. Select an alternative fluid compatible with polycarbonate and the external loop.

CC-108A Immersion Circulator Technical Data, page 2
Mounting
IP20 unit needs a dry site with 5–40 °C ambient

Degree of protection IP20 offers no water protection, so the installation location must be dry and within 5–40 °C ambient. Respect the manufacturer's space requirements for ventilation and access before siting the unit.

CC-108A Immersion Circulator Technical Data, pages 1–2
Electrical
Country-specific plug must be stated at order

Single- and two-phase devices are delivered with a power cable and a country-specific plug; the plug type must be specified when ordering to avoid a mismatch with the site socket and a rework delay.

CC-108A Immersion Circulator Technical Data, page 2
Control & I/O
Class A equipment may need extra EMC measures

The unit is classified Class A, Group 1 to EN55011; in residential or EMC-sensitive surroundings additional filtering or shielding may be required. Plan this before installing near sensitive analytical instrumentation.

CC-108A Immersion Circulator Technical Data, page 2

Selection & Integration Q&A

When should I choose the CC-108A rather than the CC-106A or CC-110A in the same series?

Pick the CC-108A when the bath inventory plus external circuit fits its 6 l volume (4 l minimum fill) and the 130 x 210 mm opening suits the workload. Step to the CC-106A for smaller baths or the CC-110A when more volume is needed; all share the same 100 °C polycarbonate limit.

What must the external circuit and power installation provide for this unit?

A 200–240 V 1~/2~ 50/60 Hz supply with UL 489 branch protection and a 10–16 A fuse, plus a country-specific plug specified at order. The external loop must tolerate up to 27 l/min at 0.7 bar pressure side and 22 l/min at 0.4 bar suction side.

What mounting and duty-cycle limits apply from the manual?

Install dry, at 5–40 °C ambient, respecting the manufacturer's space requirements; IP20 gives no water protection. Keep the bath at or above 4 l fill and at or below 100 °C, with overtemperature protection to DIN 12876 set accordingly.

What is not included in base delivery and what service items should be planned?

The country-specific plug must be specified at order; the M16x1 male pump connection is an option (#19606), not standard. No cooling unit is included unless ordered, so the 15–20 °C range requires a refrigerator or water cooling to be added.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-108A Immersion Circulator Technical Data
Datasheet ·110,4 КБ ·2 pp.
View PDF