CC-118A Technical Data Sheet

Contents

  1. Spec p. 2
    Technical specifications according to DIN 12876 standard
  2. p. 2 table, 2 rows

Technical data according to DIN 12876

ParameterValue
max. ambient temperature40 °C
from Serial-No.:S408372

1.0/21

Engineering Facts

12 facts · every claim linked to a page
Selection

The operating temperature range depends on the cooling method, so the choice of water cooling or refrigerator must match the required process temperature.

Determines whether the bath can reach the required low temperatures without an external cooler.

Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation

The unit requires a 200–240 V single- or two-phase supply with a minimum 10 A fuse and maximum 16 A fuse.

Electrical supply and branch protection must be sized accordingly before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances must not run in opposite directions, so supply quality must be checked before commissioning.

Prevents invalid supply combinations that could damage the unit or void compliance.

Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Safety

External branch circuit protection according to UL 489 is required for this equipment.

The installer must provide a listed branch circuit breaker to meet the required protection.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be ordered to avoid rework at installation.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

Polyglycol must not be used as heat transfer fluid when the Huber pump adapter is installed.

Using the wrong fluid with the pump adapter can damage the pump or compromise safety.

Source
“When using Huber pump adapter: Polyglycol is not permissible to be used as a heat transfer fluid”
Operation

The bath has a minimum filling capacity of 12.5 l within an 18 l bath volume, so the working fluid charge must be at least this amount.

Underfilling can cause pump cavitation and poor temperature control.

Source
“Bath volume 18 l min. filling capacity 12,5 l”
Installation

The ambient temperature must stay between 5 °C and 40 °C for correct operation.

Site conditions outside this range affect performance and reliability.

Source
“min. ambient temperature 5 °C”
Performance

The pressure pump delivers up to 27 l/min at 0.7 bar, and the suction pump up to 22 l/min at 0.4 bar.

External circuit pressure drop must be checked against these pump limits.

Source
“max. delivery 27 l/min max. delivery pressure 0,7 bar”
Compliance

The equipment is classified as Class A, Group 1 for electromagnetic disturbance to EN55011.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.

Confirms the safety category and the need to set overtemperature protection for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Installation

The bath bridge has a bore hole for a cooling probe, allowing an immersion cooler to be fitted.

Enables lower-temperature operation via an external immersion cooler when no refrigerator is used.

Source
“Bath bridge with bore hole for cooling probe (e.g. for immersion cooler TC45-TC100E).”