Peter Huber Kältemaschinenbau CC-118A — Thermostatic baths.

The CC-118A is a polycarbonate-bath thermostat with an 18-litre working volume and a Pilot ONE controller, available in three cooling configurations that set the lower temperature limit: 25 °C without cooling, 20 °C with water cooling, and 15 °C with a refrigerator. Its variable-speed pressure/suction pump delivers up to 27 l/min at 0.7 bar, making it suited to closed-loop temperature control of external circuits in laboratory processes.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection

Model ControllerInterfaceMax delivery flow rate, liter/minutePump connectionWeight, kgAmbient temperature limit, °C
CC-118A Current Pilot ONEEthernet, USB (Host u. D22M16x1 AG male (Option, #85–40
CC-104A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-106A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male55–40
CC-108A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #65–40
CC-110A Pilot ONEEthernet, USB (Host u. D22M16x1 AG male65–40
CC-112A Pilot ONEEthernet, USB (Host u. D27M16x1 AG male (Option, #85–40
KISS 104A KISSUSB (Device), RS232 Inte14M16x1 AG male55–40
KISS 106A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #55–40
KISS 108A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 110A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #65–40
KISS 112A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40
KISS 118A KISSUSB (Device), RS232 Inte14M16x1 AG male (Option, #85–40

Full specifications

Core Performance & Functional Specifications
Bath volume 18 l
Dimensions (w × D × H) 333 x 520 x 335 mm
Maximum delivery flow rate 27 l/min (pressure pump; 0, 7 bar)
suction pump; 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19606) (Option, #19606)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...100 °C
with cooling unit: 15...100 °C · with water cooling: 20...100 °C
Temperature stability mfr ±0, 02 K
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 275 x 321 / 150 mm (WxD/depth triplet)
Filling volume min. mfr 12, 5 l (minimum)
Dimensions, Weight & Mechanical Design
Weight 8 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max.)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behaviour and media

Installation, control and operation

Integration & Operating Notes

Electrical
Branch protection must be sized between 10 A and 16 A

The datasheet specifies a minimum fuse of 10 A and a maximum fuse of 16 A. Installing a breaker outside this window either fails to protect the unit or permits currents above its rating; external UL 489 branch circuit protection is mandatory.

CC-118A Technical Data Sheet, Technical Data
Electrical
Voltage and frequency tolerances must not oppose each other

A ±10% voltage tolerance is allowed only when the frequency tolerance does not run in the opposite direction. Supply quality should be verified before commissioning to avoid operating outside the permitted envelope.

CC-118A Technical Data Sheet, Technical Data
Process Limits
Lower temperature limit depends on the cooling method ordered

The bath reaches 25 °C without auxiliary cooling, 20 °C with water cooling, and 15 °C with a refrigerator. The process minimum must be defined before ordering, because the cooling configuration cannot be changed in the field without additional hardware.

CC-118A Technical Data Sheet, Technical Data
Process Limits
Minimum filling capacity is 12.5 l in an 18 l bath

Operating below 12.5 l risks pump cavitation and unstable temperature control. The working fluid charge must be at least this volume, and the bath bridge bore hole for an immersion cooler must remain accessible.

CC-118A Technical Data Sheet, Technical Data
Cooling & Media
Polyglycol is not permitted with the Huber pump adapter

When the Huber pump adapter is installed, polyglycol must not be used as heat-transfer fluid. Fluid selection must account for this restriction to avoid pump damage or safety compromise.

CC-118A Technical Data Sheet, Technical Data
Mounting
Country-specific plug must be specified at ordering

Single- and two-phase units are supplied with a power cable and a country-specific plug. The plug type must be stated on the order; otherwise the unit may arrive with a plug unsuitable for the site.

CC-118A Technical Data Sheet, Technical Data
Control & I/O
External Pt100 sensor can be connected to the Pilot ONE

The controller accepts an external Pt100 sensor, enabling temperature control at a point outside the bath. This is useful when the process temperature must be regulated at the external circuit rather than in the bath.

CC-118A Technical Data Sheet, Technical Data

Selection & Integration Q&A

When should I choose the CC-118A rather than the smaller CC-104A, CC-106A, CC-108A or CC-110A?

Choose the CC-118A when the application needs the largest bath volume in the series (18 l, minimum fill 12.5 l) or the 275 × 321 mm opening. The smaller siblings suit lower-volume work; all share the Pilot ONE controller and variable-speed pump.

What external circuit components must be compatible with the CC-118A pump?

External circuits must tolerate up to 27 l/min at 0.7 bar from the pressure pump or 22 l/min at 0.4 bar from the suction pump. The pump connection is M16×1 male (option #19606); polyglycol is not permitted with the Huber pump adapter.

What are the electrical and environmental limits for installing the CC-118A?

Supply 200–240 V, 1~ or 2~, 50/60 Hz, max. 10 A, with branch protection between 10 A and 16 A and external UL 489 protection. Ambient must stay 5–40 °C; IP20 and Class A Group 1 per EN55011 apply.

What is not included in the base delivery and must be ordered or provided separately?

The country-specific plug must be specified at ordering. The M16×1 pump connection is an option (#19606). An immersion cooler (TC45–TC100E) or refrigerator is required for operation below 25 °C and is not part of the base unit.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-118A Technical Data Sheet
Datasheet ·79,1 КБ ·2 pp.