Product introduction and general description of the CC-905 unit
Specpp. 1-2
Technical data, performance figures and operating parameters
Specp. 2
DIN 12876 compliant technical data table
p. 2table, 18 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Pump connection
M16x1 male
max. permissible kin. viscosity
50 mm²/s
Bath volume
26 l
min. filling capacity
19 l
Bath capacity with displacement rack
15 l
Width bath opening WxD / bath depth
260 x 260 / 200 mm
Height of bath opening
937 mm
sound pressure level +/- 4 dB(A)
64 dB(A)
Overall dimensions WxDxH **
605x706x1136 mm
Net weight
171 kg
Power supply requirement (3 phase)
400V 3~N 50Hz
max. current (3 Phase)
14 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
S538539
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-90...200 °C
Temperature stability at -10°C
0,02 K
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
3 kW
Cooling power at 200°C
2 kW
Cooling power at 100°C
2 kW
Cooling power at 20°C
2 kW
Cooling power at 0°C
2 kW
Cooling power at -20°C
1,9 kW
Cooling power at -40°C
1,7 kW
Cooling power at -60°C
1 kW
Cooling power at -80°C
0,34 kW
Cooling power at -90°C
0,13 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,5 kg
CO2 equivalent
3,2 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-508B (A1, H280)
Global Warming Potential (GWP)
13396
Refrigerant quantity 2nd stage
0,42 kg
CO2 equivalent 2nd stage
5,6 t
UN-number
UN 2857
Pressure pump
yes
max. delivery
25 l/min
max. delivery pressure
0,7 bar
Suction pump
yes
max. delivery (suction)
18,5 l/min
max. delivery pressure (suction)
0,4 bar
Order-No.: CC-905; 2027.0001.01
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator with a working range of -90 to 200 °C, so it can cover both deep-freeze and high-temperature duties in one bath.
Defines whether the circulator can meet the process temperature window before any other sizing is done.
Source
“Operating temperature range -90...200 °C”
Performance
Cooling capacity falls sharply at low temperatures, from 2 kW at 20 °C to only 0.13 kW at -90 °C, so low-temperature heat loads must be checked against the curve.
Prevents oversizing the process load at the cold end and underestimating pull-down or hold time.
Source
“at -90°C 0,13 kW”
Installation
The unit requires a 400 V three-phase supply with 14 A maximum current and 3x16 A fusing, so the electrical infrastructure must be planned accordingly.
Determines the site power connection, cable sizing and breaker rating before delivery.
Three-phase devices below 63 A are supplied with cable but without a plug, so the customer must provide and fit the site plug.
Avoids a commissioning delay caused by a missing mains plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit is rated IP20 and must be installed where ambient temperature stays between 5 and 40 °C, with space requirements respected.
An unsuitable room temperature or cramped location will reduce cooling capacity and may trip the machine.
Source
“max. ambient temperature 40 °C”
Performance
Cooling capacity data are valid at 20 °C room temperature and may drop if ambient temperature rises, so hot rooms erode the published duty.
Requires derating of the cooling capacity when the installation room is warmer than 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Safety
The refrigerant is R-452A with GWP 2141 and 1.5 kg charge (3.2 t CO2e), plus R-508B in the second stage with GWP 13396 and 0.42 kg (5.6 t CO2e), so refrigerant reporting and handling obligations apply.
Affects F-gas compliance, leak-check intervals and the environmental footprint of the chosen unit.
Source
“Global Warming Potential (GWP) 2141”
Safety
The refrigerant is classified A1 with H280 (pressurised gas), and the unit falls under pressure equipment category Art. 4.3 PED.
Informs site hazard labelling, ventilation and pressure-equipment compliance checks.
Source
“R-452A (A1, H280)”
Operation
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, which sets the external circuit pressure drop the loop may have.
Used to check that the external application and hoses do not exceed the available pump head.
Source
“max. delivery pressure 0,7 bar”
Operation
The pump connection is M16x1 male and the maximum permissible kinematic viscosity is 50 mm²/s, so hoses and the heat-transfer fluid must match these limits.
Prevents selecting a fluid too viscous at low temperature or fittings that do not mate with the unit.
Source
“max. permissible kin. viscosity 50 mm²/s”
Selection
Bath volume is 26 l with a minimum filling of 19 l, and only 15 l remains when a displacement rack is fitted.
Sets the fluid charge to purchase and confirms whether a displacement insert is needed for smaller samples.
Source
“min. filling capacity 19 l”
Installation
The unit measures 605x706x1136 mm and weighs 171 kg, so floor loading and access route must be verified.
Needed for room layout, door widths and lifting/handling planning.
Source
“Overall dimensions WxDxH ** 605x706x1136 mm”
Compliance
The unit is EMC class A group 1 to EN55011, which is industrial-grade emission classification rather than light-industrial.
May restrict installation in residential or sensitive EMC environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation
Voltage and frequency tolerances are ±5 % and ±2 % but must not be combined in the unfavourable direction, e.g. -5 % voltage with +2 % frequency is not allowed.
Guides the site power quality check before commissioning the compressor-driven unit.