Peter Huber Kältemaschinenbau CC-905 Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau CC-905 Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 64 dB.
The CC-905 is a refrigerated/heating circulator with a 26-litre bath and a working range of -90 to 200 °C, driven by a two-stage cascade using R-452A and R-508B. It is the deep-cold modification of the Cc series, distinguished by its -90 °C lower limit, and serves as a temperature-control source for external circuits and bath duties.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| CC-905 Kälte-Badumwälzthermostat Current | 64 | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 171 | 0.019 |
| CC-410wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -45–200 | 71 | 0.017 |
| CC-415wl Kälte-Badumwälzthermostat | — | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -40–200 | 61 | 0.004 |
| CC-508w Kälte-Badumwälzthermostat | 57 | 208-230V 1~/2~ 60Hz | M16x1 AG male | -55–200 | — | 0.004 |
| CC-510w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -50–200 | 106 | 0.016 |
| CC-515w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 105 | 0.016 |
| CC-520w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 161 | 0.015 |
| CC-525w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -55–200 | 163 | 0.015 |
| CC-905w Kälte-Badumwälzthermostat | — | 400V 3~N 50Hz | M16x1 AG male | -90–200 | 175 | 0.019 |
| CC-906w Kälte-Badumwälzthermostat | — | 460V 3~ 60Hz | M16x1 AG male | -90–200 | 234 | 0.025 |
| CC-K12 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 29 | 0.007 |
| CC-K15 Kälte-Badthermostat | — | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| CC-K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.012 |
| CC-K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.017 |
| CC-K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | — |
| Variostat Kältethermostat für variable Badgefäße | 54 | 220-240V 1~/2~ 50/60Hz | M16x1 AG male | -30–150 | 25 | — |
Full specifications
| Noise level | 64 dB(A) |
| Bath volume | 26 l |
| Dimensions (w × D × H) | 605 x 706 x 1136 mm |
| Filling volume | 19 l (min) |
| Heating capacity | 3 kW |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| External sensor connection | Pt100 |
| Temperature range | -90...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Delivery capacity pressure pump max. mfr | 25 l/min; 0, 7 bar |
| Delivery capacity suction pump max. mfr | 18, 5 l/min; 0, 4 bar (suction pump) |
| Bath opening WxD/BD mfr | 260 x 260 / 200 |
| Weight | 171 kg |
| Current consumption | 14 A (max) |
| Power supply | 400V 3~N 50Hz |
| Fuse mfr | 3x16A |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- The CC-905 is a refrigerated bath circulator with a 26 l bath (19 l minimum filling; 15 l with displacement rack) and a 260 x 260 mm opening, 200 mm deep.
- It is the deep-freeze member of the Cc series: siblings CC-405w, CC-410wl, CC-415wl and CC-505wl share the series architecture but the CC-905 is the variant specified down to -90 °C.
- Cascade refrigeration uses R-452A (1.5 kg, GWP 2141) in the first stage and R-508B (0.42 kg, GWP 13396) in the second; the unit is A1, H280 (pressurised gas) and falls under PED Art. 4.3.
- Heating capacity is 3 kW; temperature stability is stated as ±0.02 K.
Behaviour across the working range
- Cooling capacity is strongly temperature-dependent: 2 kW at 20 °C falls to 0.13 kW at -90 °C, so low-temperature heat loads must be checked against the curve rather than the nominal figure.
- Published output data are valid at 20 °C room temperature; higher ambient erodes cooling capacity.
- Circulation is by a variable-speed pressure-and-suction pump: pressure side up to 25 l/min at 0.7 bar, suction side up to 18.5 l/min at 0.4 bar. This sets the allowable external circuit pressure drop.
- Maximum permissible kinematic viscosity is 50 mm²/s, which constrains heat-transfer fluid choice at the cold end.
Installation, control and operation
- Power: 400 V 3~N 50 Hz, 14 A maximum, fused 3x16 A. Three-phase devices below 63 A are delivered with cable but without a plug, so the site plug must be provided.
- Voltage and frequency tolerances are ±5 % and ±2 % and must not be combined unfavourably (-5 % voltage with +2 % frequency is not allowed).
- Control is via the Pilot ONE controller with Ethernet, USB host/device and RS232; an external Pt100 sensor can be connected.
- Pump ports are M16x1 male. Protection rating is IP20; ambient range 5–40 °C. EMC classification is Class A, Group 1 to EN55011 (industrial emission class).
- Footprint 605 x 706 x 1136 mm, 171 kg; noise 64 dB(A).
Integration & Operating Notes
The unit draws up to 14 A per the datasheet and requires a 400 V three-phase neutral supply at 50 Hz with 3x16 A fusing. Cable sizing, breaker selection and phase balance must be planned before delivery; the unit arrives with cable but no plug, so the mating plug must be sourced on site.
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the unfavourable combination (-5 % voltage together with +2 % frequency) is explicitly not allowed. Verify site power quality against this rule before commissioning the compressor-driven unit.
Cooling capacity drops from 2 kW at 20 °C to 0.13 kW at -90 °C. Any process heat load at the cold end must be compared with the curve at the intended setpoint, otherwise pull-down time or hold temperature will not be met.
The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar. Hoses, external application and fittings must be sized so the circuit pressure drop stays within the available head at the required flow.
The maximum permissible kinematic viscosity is 50 mm²/s. Fluid selection must account for viscosity rise at low temperature, and pump connections are M16x1 male, so hoses and adapters must match this thread.
The unit is IP20 and rated for ambient temperatures of 5 to 40 °C. Cooling capacity data are valid at 20 °C room temperature and may drop as ambient rises, so a hot or cramped location erodes published duty and may trip the machine.
R-452A (1.5 kg, GWP 2141, 3.2 t CO2e) and R-508B (0.42 kg, GWP 13396, 5.6 t CO2e) are A1, H280 pressurised gases under PED Art. 4.3. Leak-check intervals, labelling and refrigerant handling obligations follow from these charges.
The Pilot ONE controller provides Ethernet, USB host and device, and RS232 interfaces, and accepts an external Pt100 sensor. These interfaces allow remote monitoring and external-temperature control of the application.
Selection & Integration Q&A
Choose the CC-905 when the process needs setpoints down to -90 °C or a 26 l bath; its cascade refrigeration is what reaches the deep-cold end. Siblings such as CC-505wl cover less extreme ranges and smaller duties, so pick them when -90 °C is not required.
The loop must fit within 0.7 bar pressure-side head (25 l/min max) and 0.4 bar suction-side (18.5 l/min), with M16x1 male connections. Fluid kinematic viscosity must stay at or below 50 mm²/s at operating temperature.
Install at 5–40 °C ambient, IP20, with 400 V 3~N 50 Hz supply, 14 A max and 3x16 A fusing. Voltage/frequency tolerances are ±5 %/±2 % but must not combine unfavourably (-5 % V with +2 % Hz is not allowed).
Three-phase units below 63 A ship with cable but without a plug, so the site plug must be provided. Heat-transfer fluid, external Pt100 sensor and hoses are not part of the base unit and must be specified separately.