Product cover and model identification for KISS K20
Specp. 2
Technical data table according to DIN 12876
p. 2table, 7 rows
Technical data according to DIN 12876
Parameter
Value
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S393794
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a wide operating temperature range from -30 to 200 °C, making it suitable for both sub-ambient and high-temperature bath applications.
Defines the process window the engineer can cover with a single unit, avoiding a separate chiller for low temperatures.
Source
“Operating temperature range -30...200 °C”
Performance
Cooling capacity drops sharply at low temperatures, from 0.4 kW at 20 °C to only 0.065 kW at -30 °C, so heat loads must be sized against the actual operating temperature.
Prevents oversizing assumptions based on nominal cooling power and ensures the bath can hold setpoint under load at the coldest point.
Source
“at -30°C 0,065 kW”
Safety
The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects room ventilation, ignition-source control and refrigerant handling procedures.
The refrigerating unit is air-cooled, so adequate air flow and clearance around the unit are required; space requirements must be checked before siting.
Air-cooled condensers lose capacity and can trip on high pressure if installed in tight or poorly ventilated spaces.
Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Installation
The unit requires a 220-240 V single- or two-phase 50/60 Hz supply and draws up to 10 A, so the circuit and plug must be rated accordingly.
Determines the electrical connection, breaker sizing and whether a country-specific plug must be specified at order.
Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation
Voltage and frequency tolerances are limited to ±5 % and ±2 % respectively, and the combination of -5 % voltage with +2 % frequency is explicitly not allowed.
Prevents installation on unstable or generator-fed supplies where simultaneous voltage and frequency deviations could damage the compressor.
Source
“-5% voltage and +2% frequency -> not allowed!”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic pump ordered.
Prevents chemical attack and pump failure when selecting the heat transfer fluid for the application.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection the manufacturer recommends water mixed with glysantine or ethylene glycol rather than pure water.
Guides fluid selection to avoid freezing and damage when operating below ambient or near 0 °C.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The unit is rated IP20, so it must be installed in a dry, protected location and not exposed to water spray or condensation.
IP20 limits the environment of installation and rules out washdown or wet locations.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must stay between 5 and 40 °C, and cooling capacity may drop if ambient rises above 20 °C.
Room conditions directly affect achievable cooling and must be considered when sizing and siting the unit.
Source
“If the ambient temperature rises, the cooling capacity may drop.”
Compliance
The equipment is EMC-classified as Class A, Group 1 to EN55011, which is intended for industrial environments and may cause interference in residential areas.
Class A equipment has implications for installation in non-industrial or residential environments and for EMC coordination.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The unit provides adjustable overtemperature protection according to DIN 12876 and is classified Safety class III / FL.
Confirms the safety concept required for unattended operation and for protection of temperature-sensitive samples.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
An autostart function restarts the unit after a power failure, which may be undesirable for some processes and should be evaluated.
Autostart behavior after mains interruption must be considered for process safety and unattended operation.
Source
“Autostart function for power failure”
Selection
A Pt100 sensor connection is available only factory-fitted as an option and is used to display, not control, an external process temperature.
Clarifies that external temperature monitoring is display-only and must be ordered with the unit, not retrofitted.
Source
“Option: Pt100 sensor connection #10688 to display (not control)”