Engineering Digest Document Viewer Related Models

Ministat 230w Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of the Ministat 230w circulator
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits per DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Consumption at water 15°C, flow 0°C20,4 l/h
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
Bath volume3,5 l
min. filling capacity3 l
Bath capacity with displacement rack1,7 l
Width bath opening WxD / bath depth170 x 85 / 135 mm
Overall dimensions WxDxH **255x450x476 mm
Net weight36 kg
Power supply requirement208-240V 1~/2~ 50/60Hz
max. current11 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S393233

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Ministat 230w (Order-No.: 2015.0007.01)

CharacteristicValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Heating power at 208V1,6 kW
Cooling power at 100°C0,42 kW
Cooling power at 20°C0,42 kW
Cooling power at 0°C0,38 kW
Cooling power at -10°C0,33 kW
Cooling power at -20°C0,25 kW
Cooling power at -30°C0,14 kW
Cooling power at -40°C0,05 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,06 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Pressure pumpyes
max. delivery22 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)16 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure between 2 bar and 6 bar, so the facility water supply must be sized accordingly.

Insufficient or excessive cooling water pressure will prevent proper operation or damage the cooling circuit.

Source
“min. cooling water differential pressure 2 bar max. cooling water pressure 6 bar”
Installation

The cooling water intake temperature must be between 5°C and 20°C, and pipes must be insulated if the water can fall below the dew point.

Out-of-range cooling water temperature reduces cooling power and can cause condensation and corrosion.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Aggressive or incompatible cooling water will corrode or degrade these materials and cause leaks.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Voltage tolerance is ±5% only when frequency tolerance is simultaneously ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

The supply must be checked against this combined tolerance rule before connecting the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is classified EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments and may cause interference in residential areas.

Installation in a residential or sensitive EMC environment may require additional filtering or separation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be stated on the order to avoid a non-usable power connection.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Selection

Cooling power falls from 0.42 kW at 20°C to 0.05 kW at -40°C, so the cooling capacity at the lowest operating temperature is very limited.

The heat load at low temperature must be kept below this value or the set point cannot be reached.

Source
“at -40°C 0,05 kW”
Selection

The pump delivers up to 22 l/min at 0.7 bar pressure and suction up to 16 l/min at 0.4 bar, which limits the external circuit resistance that can be handled.

External loop pressure drop must be checked against the pump curve to ensure adequate flow.

Source
“max. delivery 22 l/min max. delivery pressure 0,7 bar”
Selection

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s, which restricts the use of highly viscous fluids at low temperature.

A fluid exceeding this viscosity will overload the pump and reduce circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Selection

The bath volume is 3.5 l with a minimum filling of 3 l, and only 1.7 l remains when a displacement rack is used.

The usable bath volume determines the size of samples and the amount of fluid to prepare.

Source
“Bath volume 3,5 l min. filling capacity 3 l Bath capacity with displacement rack 1,7 l”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, with a charge of only 0.06 kg.

Flammable refrigerant classification affects room ventilation and ignition-source restrictions at installation.

Source
“R-290 (A3, H220)”
Safety

The unit has safety classification III / FL according to DIN 12876 and adjustable overtemperature protection.

The safety class determines the permitted bath fluids and the required protective measures.

Source
“Safety classification III / FL”
Operation

The maximum ambient temperature is 40°C and the minimum is 5°C, so the installation room must stay within this range.

Operation outside this ambient range degrades cooling performance and may trip protection.

Source
“max. ambient temperature 40 °C min. ambient temperature 5 °C”
Compliance

The equipment complies with US-SNAP and applicable EU laws, with the US-SNAP end-use being industrial process refrigeration.

This confirms refrigerant compliance for industrial process use in the US and EU markets.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”