Installation
The unit requires cooling water with a differential pressure between 2 and 6 bar, so the facility water supply must be sized accordingly.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the water-cooled condenser.
Source
“min. cooling water differential pressure
2 bar
max. cooling water pressure
6 bar”
Installation
Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Out-of-range cooling water temperature reduces cooling capacity and risks condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops significantly at low temperatures, from 0.6 kW at 20 °C to only 0.05 kW at -40 °C.
The user must verify that available cooling capacity at the target low temperature meets the process heat load.
Source
“at -40°C
0,05 kW”
Installation
The electrical supply must be 220-240 V single or two-phase at 50/60 Hz with a 16 A fuse and max current of 11.5 A.
The site electrical installation must provide the correct voltage, phase configuration and protection rating.
Source
“Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
11,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.
Unstable mains supply outside these combined limits can cause malfunction or damage to the compressor and electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-290 (propane, A3 flammable, H220), so the installation area must be assessed for flammable refrigerant handling.
A3 flammable refrigerant imposes ventilation, ignition-source and handling requirements on the installation site.
Source
“Refrigerant (ASHRAE, GHS)
R-290 (A3, H220)”
Operation
The pump delivers up to 22 l/min at 0.7 bar pressure and 16 l/min suction at 0.4 bar, with M16x1 male connections.
External circuit pressure drop and hose sizing must be checked against these pump limits.
Source
“max. delivery
22 l/min
max. delivery pressure
0,7 bar”
Operation
The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s.
Selecting a fluid too viscous at low temperature will overload the pump and reduce circulation.
Source
“max. permissible kin. viscosity
50 mm²/s”
Installation
The bath requires a minimum filling capacity of 3.5 l, and only 2.8 l remains when a displacement rack is used.
Underfilling risks pump cavitation and poor temperature control; the displacement rack reduces usable volume.
Source
“min. filling capacity
3,5 l
Bath capacity with displacement rack
2,8 l”
Installation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.
IP20 offers no water protection, and ambient limits affect cooling performance and electronics reliability.
Source
“Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C”
Compliance
The unit is classified as pressure equipment category Art. 4.3 PED and safety class III / FL.
This determines the applicable pressure-equipment compliance and documentation obligations.
Source
“Pressure equipment category
Art. 4.3 PED”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.
Aggressive or scaling cooling water will corrode these materials and impair heat transfer.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Operation
The unit consumes 34 l/h of cooling water at 15 °C inlet and 15 °C flow.
The facility must supply adequate cooling water flow rate to sustain rated cooling capacity.
Source
“Consumption at water 15°C, flow 15°C
34 l/h”
Selection
The operating temperature range is -45 to 200 °C with a temperature stability of 0.02 K at -10 °C.
These define whether the unit can meet the process temperature window and stability requirement.
Source
“Operating temperature range
-45...200 °C
Temperature stability at -10°C
0,02 K”