Peter Huber Kältemaschinenbau Ministat 240w Kompakter Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau Ministat 240w Kompakter Kälte-Badumwälzthermostat — Thermostatic baths.
The Ministat 240w is a water-cooled circulating bath thermostat with a 5.5 L bath and a variable-speed pressure/suction pump, covering -45 to 200 °C with ±0.02 K stability. It is the water-cooled variant of the Ministat series, distinguished by a natural R-290 refrigerant and facility-water condenser. It serves as a precise temperature source for external circuits or bath immersion in laboratory and pilot processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Ministat Series — Model Comparison & Selection
| Model | Max delivery flow rate, liter/minute | Power supply | Temperature range, °C | Weight, kg | Ambient temperature limit, °C | Controller |
|---|---|---|---|---|---|---|
| Ministat 240w Kompakter Kälte-Badumwälzthermostat Current | 22 | 220-240V 1~/2~ 50/60Hz | -45–200 | 41 | 5–40 | Pilot ONE |
| Ministat 125w Kompakter Kälte-Badumwälzthermostat | 22 | 110-120V 1~ 60Hz | -25–150 | 26 | 5–40 | Pilot ONE |
| Ministat 230w Kompakter Kälte-Badumwälzthermostat | 16 | 208-240V 1~/2~ 50/60Hz | -40–200 | 36 | 5–40 | Pilot ONE |
Full specifications
| Maximum delivery flow rate | 22 l/min (pressure pump; 0, 7 bar) suction pump; 0, 4 bar: 16 l/min |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | Yes |
| Pump type | Pressure and suction pump, variable speed |
| Refrigerator | water-cooled, natural refrigerant |
| External sensor connection | Pt100 |
| Temperature range | -45...200 °C |
| Temperature stability mfr | ±0, 02 K |
| Dimensions (W x D x H) mfr | 300 x 465 x 516 mm |
| Heating capacity mfr | 1, 8 - 2, 1 kW |
| Bath opening WxD/BD mfr | 205 x 85 / 157 mm |
| Bath volume mfr | 5, 5 l |
| Filling volume min. mfr | 3, 5 l (min) |
| Weight | 41 kg |
| Fuse rating | 16 A |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 11, 5 A (max) |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Maximum viscosity | 50 mm²/s |
| Refrigerant quantity 1st stage mfr | 0, 06 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 34 l/h (flow 15°C) |
| Cooling water differential pressure | 2 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Cooling capacity mfr | 100:0, 6; 20:0, 6; 0:0, 55; -10:0, 42; -20:0, 35 |
| Safety class | III / FL |
Description
Construction and series positioning
- The Ministat 240w is a compact circulating bath thermostat (300 x 465 x 516 mm, 41 kg) with a 5.5 L bath, 205 x 85 mm opening and 157 mm depth; minimum filling is 3.5 L, reduced to 2.8 L with a displacement rack.
- Unlike the air-cooled Ministat 230 sibling, this 'w' modification rejects heat to facility cooling water through a water-cooled condenser using natural refrigerant R-290 (A3, H220; 0.06 kg charge).
- The pressure/suction pump is variable-speed, rated up to 22 l/min at 0.7 bar delivery and 16 l/min at 0.4 bar suction, with M16x1 male connections.
Thermal behavior and media
- Working range is -45 to 200 °C with ±0.02 K stability at -10 °C; heating capacity is 1.8–2.1 kW.
- Cooling capacity falls steeply at low temperature: 0.6 kW at 100 °C and 20 °C, 0.55 kW at 0 °C, 0.42 kW at -10 °C, 0.35 kW at -20 °C, 0.125 kW at -30 °C and only 0.05 kW at -40 °C, so low-temperature loads must be sized against this curve.
- Bath fluid must not exceed 50 mm²/s kinematic viscosity; the cooling-water circuit wetted materials are copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.
Control, installation and operation
- Control is via the Pilot ONE controller with Ethernet, USB host/device and RS232 interfaces, plus a Pt100 external sensor connection.
- Electrical supply is 220–240 V, 1~ or 2~, 50/60 Hz, 16 A fuse, max 11.5 A; voltage and frequency tolerances of ±5% and ±2% must not be combined in opposite directions.
- Cooling water must be supplied at 5–20 °C inlet with 2–6 bar differential pressure and 34 l/h consumption at 15 °C; pipes must be insulated where condensation can occur. The unit is IP20 and rated for 5–40 °C ambient.
Integration & Operating Notes
Below 2 bar the water-cooled condenser cannot reject rated heat, and above 6 bar the condenser and internal plumbing are at risk. Verify the supply pressure at the unit's G1/2 male connection under flowing conditions, not static, and provide 34 l/h at 15 °C inlet.
Warmer inlet water reduces cooling capacity and colder water raises condensation risk on uninsulated pipes. Insulate cooling-water lines wherever surface temperature can fall below ambient dew point, and confirm the facility can hold the inlet within this window year-round.
At low temperatures many fluids thicken beyond this limit, overloading the pump and reducing circulation. Select a heat-transfer fluid whose kinematic viscosity at the lowest operating temperature remains under 50 mm²/s, and account for the 3.5 L minimum filling (2.8 L with a displacement rack).
The unit has no water protection, so it must not be placed where splashes or condensation drip onto it. Ambient must remain between 5 and 40 °C; because the refrigerant is flammable R-290 (A3), assess the installation area for ignition sources and ventilation per local flammable-refrigerant rules.
Available cooling is 0.125 kW at -30 °C and 0.05 kW at -40 °C, versus 0.6 kW at 20 °C. Processes with meaningful heat load near the lower end of the -45 to 200 °C range must be checked against this curve, or a larger chiller selected.
Selection & Integration Q&A
Pick the 240w when facility cooling water is available and you need the 5.5 L bath with water-cooled heat rejection; the 230 is an air-cooled sibling, and the 125w is a smaller bath. Choose this one when low-temperature load is modest and water cooling avoids room heat rejection.
Cooling water at 5–20 °C inlet, 2–6 bar differential pressure, 34 l/h at 15 °C, connected via G1/2 male. The external loop connects to M16x1 male pump ports; the variable-speed pressure/suction pump delivers up to 22 l/min at 0.7 bar or 16 l/min suction at 0.4 bar.
Operate between 5 and 40 °C ambient in a dry location, as the IP20 enclosure offers no water protection. The R-290 (A3) refrigerant requires a flammable-refrigerant assessment of the area. Electrical supply is 220–240 V 1~/2~ 50/60 Hz with a 16 A fuse and 11.5 A max current.
The datasheet does not list a maintenance schedule; plan for cooling-water circuit inspection given copper/brass/EPDM wetted parts, and periodic bath-fluid checks against the 50 mm²/s viscosity limit. External Pt100 sensor, bath fluid, hoses and fittings are not included and must be specified separately.