General description and key features of TC100-F controllers
Specp. 1
Electrical, thermal, and mechanical specifications of the controller
Orderingp. 1
Model codes and available configurations for ordering
p. 1table, 35 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-100...40 °C
Cooling power
at 0°C
0,16 kW
at -20°C
0,15 kW
at -30°C
0,14 kW
at -50°C
0,13 kW
at -60°C
0,12 kW
at -80°C
0,12 kW
at -90°C
0,07 kW
at -100°C
0,01 kW
Safety classification
I / NFL
Refrigeration machine
air-cooled, natural
refrigerant
Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)
Global Warming Potential (GWP)
0
Refrigerant quantity
0,114 kg
CO2 equivalent
0,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-1150 (A3, H220)
Global Warming Potential (GWP)
4
Refrigerant quantity 2nd stage
0,056 kg
CO2-Äquivalent 2.Stufe
0,0 t
Nominal diameter probe
13 mm
Length of probe
900 mm
Length flexible connection
1150 mm
Overall dimensions WxDxH **
295x500x570 mm
Net weight
57 kg
sound pressure level +/- 4 dB(A)
58 dB(A)
Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
7 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
Art. 4.3 PED
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S569390
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is rated for an operating temperature range of -100 to 40 °C, so it must be selected only for processes within that envelope.
Defines the process temperature window the immersion cooler can serve.
Source
“Operating temperature range
-100...40 °C”
Performance
Cooling capacity drops sharply near the cold end, from 0.12 kW at -80 °C to 0.07 kW at -90 °C and only 0.01 kW at -100 °C, so the lowest temperatures are practical only with very small heat loads.
Prevents oversizing expectations at the extreme low end of the range.
Source
“at -90°C
0,07 kW
at -100°C
0,01 kW”
Performance
Published cooling capacities are valid at 20 °C room temperature and may fall at higher ambient, so capacity margins must be checked against the actual site ambient.
Site ambient directly affects achievable cooling power.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit may only be operated with ambient temperature between 5 °C and 40 °C, so the installation location must stay within this range.
Ambient limits constrain where the air-cooled unit can be installed.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The power supply must be 220-240 V single or two-phase at 50/60 Hz with a maximum current of 7 A and a fuse between 10 A and 16 A.
Determines electrical supply and protection sizing at the installation site.
Source
“Power supply requirement
220-240V 1~/2~ 50/60Hz
max. current
7 A
min. Fuse
10A
max. Fuse
16A”
Installation
Voltage and frequency tolerances are coupled: ±5% voltage is allowed only with ±2% frequency, and combinations such as -5% voltage with +2% frequency are not allowed.
Prevents installation on a supply whose combined voltage/frequency deviation exceeds the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering, while three-phase devices below 63 A come with cable but without plug.
Affects ordering configuration and on-site electrical connection work.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Safety
The refrigerant is R-1270 (A3, H220) with GWP 0 and a charge of 0.114 kg, and the second stage uses R-1150 (A3, H220) with 0.056 kg, so both circuits are flammable A3 refrigerants requiring appropriate site precautions.
Flammable refrigerant classification drives ventilation, ignition-source and handling requirements.
Source
“Refrigerant (ASHRAE, GHS)
R-1270 (A3, H220)”
Compliance
The equipment is classified as pressure equipment category Art. 4.3 PED, which defines the applicable pressure-equipment compliance route.
Determines which pressure-equipment documentation and conformity requirements apply.
Source
“Pressure equipment category
Art. 4.3 PED”
Compliance
The unit is EMC-classified as Class A, Group 1 to EN55011, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment has installation restrictions regarding residential/light-industrial use.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Space requirements around the unit must be respected, as referenced in the operating conditions.
Air-cooled units need clearance for heat rejection and service access.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Selection
The probe has a nominal diameter of 13 mm and a length of 900 mm with a 1150 mm flexible connection, which must fit the vessel and routing at the point of use.
Probe and connection geometry determine mechanical fit in the application.
Source
“Nominal diameter probe
13 mm
Length of probe
900 mm
Length flexible connection
1150 mm”
Installation
The unit weighs 57 kg net and measures 295x500x570 mm, so handling and floor/bench loading must be planned accordingly.
Weight and footprint affect transport, lifting and installation space.
Source
“Overall dimensions WxDxH **
295x500x570 mm
Net weight
57 kg”
Operation
The refrigeration unit works continuously, which must be considered in operating strategy and energy/heat-load planning.
Continuous operation affects heat rejection, noise and wear expectations.